[0] Title = ListTitle = Message = Indication for when no alarms is a possible situation [1] Title = HD POST: Stuck Button ListTitle = Stop or Off button are pressed for more than 1 sec during POST. Message = Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up [2] Title = HD POST: FPGA Compatibility ListTitle = Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. Message = HD FPGA POST failure [3] Title = DG POST: FPGA Compatibility ListTitle = Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header Message = DG FPGA POST failure SRSDG 602 SRSDG 814 SRSDG 1061 = [4] Title = HD POST: WD Safestate ListTitle = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA Message = HD Watchdog POST failure [5] Title = DG POST: WD Safestate ListTitle = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V Message = DG Watchdog POST failure SRSDG 53 = [6] Title = HD POST: Proc- UI Comm ListTitle = If the UI failed to checkin via the CAN bus within 40 seconds Message = UI communication POST failure. UI failed to communicate within a reasonable time after power up [7] Title = HD Alarm (Low): RTC Bat Low ListTitle = If the RTC is not functional as indicated by the bits in the first 3 control registers. Message = Real-time clock configuration error [8] Title = HD POST: Acc Vector ListTitle = If the calculated vector length of the accelerometer x, y, z is not the nominal length. Message = HD accelerometer failed POST [9] Title = DG POST: Acc Vector ListTitle = If the calculated vector length of the accelerometer x, y, z is not the nominal length. Message = DG accelerometer failed POST SRSDG 323 = [10] Title = HD POST: BLD Config Emb Mode ListTitle = If the blood leak sensor embedded mode is unresponsive. Message = HD blood leak sensor embedded mode communication failed [11] Title = Dial. Gen: Dialysate Temp Low ListTitle = If the dialysate temperature is below the target temperature for a certain period of time. Message = Dialysate temperature is below the target temperature [12] Title = HD POST: HD Image ListTitle = If the calculated CRC of the HD image does not match the expected CRC. Message = HD firmware image integrity POST test failed [13] Title = DG POST: DG Image ListTitle = If the calculated CRC of the DG firmware image does not matched the expected CRC Message = DG firmware image integrity POST test failed SRSDG 813 = [14] Title = DG POST: Usage Rec NVRAM CRC ListTitle = If the DG usage information record's CRC does not match the calculated CRC in POST. Message = DG non-volatile usage information group invalid record CRC SRSDG 957 = [15] Title = Dial. Gen: Dialysate Temp High ListTitle = If the dialysate temperature is > 46 C for a certain period of time. Message = Dialysate temperature is out of high safety range [16] Title = HD POST: Primary Alarm Current ListTitle = If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output Message = HD alarm audio failed POST [17] Title = HD POST: UI POST Not Complete ListTitle = If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. Message = HD UI POST failed [18] Title = Loss of DG Comm (DG ACK) ListTitle = HD message to DG that requires acknowledge was not acknowledged within a certain time. Message = HD to DG comm error. [19] Title = Dial. Gen: Dialysate Temp High ListTitle = If the dialysate temperature is above the target temperature for a certain period of time. Message = Dialysate temperature is above the target temperature [20] Title = DG POST: Conductivity Cal ListTitle = If the calibration data that was received from NVDataMgmt is NOT legitimate Message = DG conductivity sensors invalid calibration record SRSDG 16 = [21] Title = DG POST: Drain Line Vol Cal ListTitle = If the calibration data that was received from NVDataMgmt is NOT legitimate Message = DG drain line volume invalid calibration record SRSDG 936 = [22] Title = DG POST: Reservoir Vol Cal ListTitle = If the calibration data that was received from NVDataMgmt is NOT legitimate Message = DG reservoirs invalid calibration record SRSDG 935 = [23] Title = DG POST: Acid Conc Cal ListTitle = If the non-volatile data is valid and if not raise the alarm Message = DG acid concentrate calibration record SRSDG 934 = [24] Title = DG POST: Bicarb Conc Cal ListTitle = If the non-volatile data is valid and if not raise the alarm Message = DG bicarb concentrate calibration record SRSDG 933 = [25] Title = DG POST: Acc Cal ListTitle = If the calibration data that was received from NVDataMgmt is NOT legitimate Message = DG accelerometer invalid calibration record SRSDG 817 = [26] Title = HD POST: Acc Cal ListTitle = If the accel calibration data that was received from NVDataMgmt is NOT legitimate. Message = HD accelerometer invalid calibration record [27] Title = Process: Pre- Tx Wet Self- Test ListTitle = BLD fails to complete zero sequence. Message = HD BLD sensor failed zero sequence [28] Title = DG Fault: Conductivity RTD Comm ListTitle = If the FPGA does not report fresh data for a certain period of time Message = DG two wire sensors FPGA fault SRSDG 864 = [29] Title = HD POST: SP DAC Cal ListTitle = If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. Message = HD heparin force sensor invalid calibration record [30] Title = HD Fault: SW Invalid State ListTitle = If the specific software faults listed in alarm management SW faults has been detected Message = HD Software fault. Software found itself in an unexpected state [31] Title = HD Fault: BP Current ListTitle = If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. Message = Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running [32] Title = HD Fault: BP Speed ListTitle = If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM Message = Blood pump failed motor off check. Measured speed while commanded off [33] Title = HD Fault: BP Direction ListTitle = If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. Message = Blood pump failed motor direction check. Measured vs. commanded [34] Title = HD Fault: BP Rotor ListTitle = If the delta of the measured motor speed and the commanded motor speed is not within tolerance. Message = Blood pump failed rotor speed check. Mismatch with rotor and motor speeds [35] Title = HD Fault: DPi Current ListTitle = If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. Message = Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running [36] Title = HD Fault: DPi Speed ListTitle = If during command off state, the DPi pump motor speed is measured > 100.0 RMP. Message = Dialysis inlet pump failed motor off check. Measured speed while commanded off [37] Title = HD Fault: DPi Direction ListTitle = If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. Message = Dialysis inlet pump failed motor direction check. Measured vs. commanded [38] Title = HD Fault: DPi Rotor ListTitle = If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. Message = Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds [39] Title = HD Fault: DPo Current ListTitle = If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. Message = Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running [40] Title = HD Fault: DPo Speed ListTitle = If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. Message = Dialysis outlet pump failed motor off check. Measured speed while commanded off [41] Title = HD Fault: DPo Direction ListTitle = If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. Message = Dialysis outlet pump failed motor direction check. Measured vs. commanded [42] Title = HD Fault: DPo Rotor ListTitle = If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed Message = Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds [43] Title = DG Fault: Iso ADC Comm ListTitle = If the FPGA does not report fresh data for a certain period of time Message = Main primary heater FPGA fault SRSDG 954 = [44] Title = HD Fault: UI Comm ListTitle = If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. Message = User interface communication timeout [45] Title = HD Fault: CAN CRC Com ListTitle = If too many bad messages CRCs are received within a set period of time Message = Too many bad CRCs detected on received system messages [46] Title = HD Fault: CAN ACK UI Com ListTitle = If no retries left in the pending acknowlegement list. Message = System message that required acknowledgment was not acknowledged [47] Title = ListTitle = Message = [48] Title = HD Fault: UF Rate Tare Err ListTitle = If the UF volume (refUFvolume - measureVolume) > 100.0 mL. Message = Ultrafiltration volume accuracy error during treatment [49] Title = HD Fault: FPGA Com ListTitle = If there are more than 3 comm errors in a 1 min period. Message = HD FPGA communication down for too long [50] Title = DG Fault: Valve Control ListTitle = If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. Message = DG FPGA not accepting commanded valve states SRSDG 26 = [51] Title = HD Fault: BP Speed ListTitle = If measured motor speed versus commanded rotor speed exceeded limits. Message = Blood pump failed motor speed check. Measured vs. commanded [52] Title = HD Fault: DPi Speed ListTitle = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. Message = Dialysate inlet pump failed motor speed check. Measured vs. commanded [53] Title = HD Fault: DPo Speed ListTitle = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. Message = Dialysate outlet pump failed motor speed check. Measured vs. commanded [54] Title = HD Fault: Critical Data Err ListTitle = If the critical data failed range check. Message = HD critical data integrity check failed [55] Title = DG Fault: Critical Data Err ListTitle = If the critical data failed range check. Message = DG critical data integrity check failed SRSDG 921 = [56] Title = HD Fault: Acc Com ListTitle = If no fresh sample of accelerometer data has been detected for more than 100 msec. Message = HD accelerometer error (no readings or FPGA reports error) [57] Title = DG Fault: Acc Com ListTitle = If no fresh sample of accelerometer data has been detected for more than 100 msec. Message = DG accelerometer error (no readings or FPGA reports error) SRSDG 869 SRSDG 870 = [58] Title = HD Fault: Valve Home ListTitle = If the number of failed homing > = 3 counts. Message = HD valve homing failed [59] Title = HD Fault: Valve Transition Err ListTitle = if the valve's transition time has timed out (3000msec) Message = HD valve transition time out [60] Title = HD Fault: Valve Control ListTitle = If the valve bit masks are not as expected in different valve control mode. Message = HD valve not functional [61] Title = HD Fault: Valve Current ListTitle = If the valve current is over the threshold of 1.0 A for 1000 msec. Message = HD valve current out of range [62] Title = HD Fault: Valve Position ListTitle = If the current position has deviated from the expected position for more than 1 second. Message = HD valve position out of target [63] Title = ListTitle = Message = [64] Title = DG Fault: Baro Range ListTitle = If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. Message = Barometric pressure is out of range SRSDG 874 = [65] Title = Loss of DG Comm (Reject) ListTitle = If the latest DG command response indicated invalid parameter Message = HD requests DG command with invalid parameter fault [66] Title = HD POST: BLD Config Set Point ListTitle = If the set point of the blood leak sensor does not match the set point in the calibration record. Message = HD blood leak sensor set point set failure [67] Title = Process: Pre- Tx Remove Cartridge ListTitle = If the blood pump pressure > 5000 Message = HD blood pump occlusion self-test failure alarm [68] Title = Dial. Gen: Dialysate not ready ListTitle = If the recirculation level has exceeded the max allowed Message = HD active reservoir recirculation out of range [69] Title = HD POST: BLD Cal ListTitle = If the BLD data that was received from NVDataMgmt is NOT legitimate. Message = HD blood leak sensor invalid calibration record [70] Title = Process: Pre- Tx Remove Cartridge ListTitle = If arterial pressure < -300 or > 100 mmHg Message = HD arterial pressure self-test failure alarm [71] Title = Process: Pre- Tx Remove Cartridge ListTitle = If the venous pressure < -100 or > 600 mmHg Message = HD venous pressure self-test failure alarm [72] Title = Loss of DG Comm (LC) ListTitle = No new load cell data for 2 seconds. Message = HD No load cell data message receive at least once every 2 seconds [73] Title = Loss of DG Comm (TDI/TRO) ListTitle = No new dialysate temperature data for 2 seconds. Message = HD No dialysate temperature data message received at least once every 2 seconds [74] Title = DG Fault: UVi On No Flow ListTitle = If the inlet UV reactor is on while VPi is closed for a certain period of time. Message = DG inlet UV reactor is on while there is no flow SRSDG 600 = [75] Title = HD Fault: Syringe DAC ListTitle = If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. Message = HD syringe pump self-test failure alarm [76] Title = HD Fault: Voltage Monitor ListTitle = If the HD monitored voltages are out of range for a certain period of time. Message = HD monitored voltage is out of range [77] Title = DG Fault: Voltage Monitor ListTitle = If the DG monitored voltages are out of range for a certain period of time. Message = DG monitored voltage is out of range SRSDG 46 SRSDG 866 SRSDG 1088 = [78] Title = HD Fault: Syringe Pump Enc Dir ListTitle = If the measured encoder direction of the syringe pump does not agree with the expected direction Message = HD syringe pump direction (from encoder) error [79] Title = HD Fault: Syringe Pump Drv Dir ListTitle = If the measured direction of the syringe pump motor controller does not agree with the expected direction Message = HD syringe pump direction (from controller) error [80] Title = HD Fault: Syringe Pump Driver ListTitle = If the FPGA reported the syringe pump fault Message = HD syringe pump fault reported by FPGA [81] Title = HD Fault: Syringe Pump OT ListTitle = If the syringe pump position is deemed to have overtraveled Message = HD syringe pump over travel error [82] Title = HD Fault: Syringe Pump DAC Err ListTitle = If the syringe pump DAC write error is reported by the FPGA Message = HD syringe pump DAC write failure [83] Title = HD Fault: Syringe Pump On Err ListTitle = If the syringe pump is on while the blood pump is off Message = HD syringe pump is running while the BP is off [84] Title = DG POST: DG RTC Time ListTitle = If DG RTC set year is < 2021. Message = DG RTC set year is invalid SRSDG 815 = [85] Title = Blood Stop: Cartridge Lever Open ListTitle = If the pump track lever is detected as unlatched when it should be. Message = HD pump track lever is not fully latched [86] Title = HD POST: HD RTC Time ListTitle = If HD RTC set year is < 2021. Message = HD RTC set year is invalid [87] Title = DG POST: Heat Cal ListTitle = If the heating paramters' non-volatile record's CRC does not match the calculated CRC. Message = DG heating parameters invalid calibration record SRSDG 929 = [88] Title = DG Fault: CP1 CP2 Speed Range ListTitle = Hall sensor pulse to pulse timing is < 1ms (speed too fast). Message = DG concentrate pump hall sensor out of range. [89] Title = DG Fault: UVo On No Flow ListTitle = If the outlet UV reactor is on while the RO flow is less than 200 mL/min. Message = DG outlet UV reactor is on while there is no flow SRSDG 599 = [90] Title = DG Fault: LC A1/B1 Comm ListTitle = If the FPGA does not report fresh data for a certain period of time. Message = DG load cells A1/B1 FPGA fault SRSDG 604 = [91] Title = Loss of DG Comm (FMD) ListTitle = If not receiving new dialysate flow readings in expected interval (3 seconds) Message = HD No dialysate flow data received in the last 3 seconds [92] Title = ListTitle = Message = [93] Title = DG Fault: RO Permeate Cond High ListTitle = If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. Message = DG outlet primary conductivity sensor out of range SRSDG 430 = [94] Title = DG Fault: Pressure Sensor Range ListTitle = If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. Message = DG pressure sensor out of range SRSDG 629 = [95] Title = DG Fault: WD Error ListTitle = If the watchdog signal is low while not in POST. Message = DG watchdog expired SRSDG 907 = [96] Title = Dial. Gen: Inlet Water Temp High ListTitle = If the inlet water temperature is > 37 C. Message = DG inlet water temperature out of range SRSDG 410 = [97] Title = Dial. Gen: Dialysate Cond. ListTitle = If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. Message = DG fill conductivity out of range SRSDG 397 = [98] Title = HD Alarm (Low): Battery Com ListTitle = If battery I2C communication fault persistence count > 5. Message = HD battery or battery charger communication fault [99] Title = HD Fault: Syringe Pump Stall ListTitle = If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. Message = HD syringe pump stall alarm [100] Title = ListTitle = Message = [101] Title = HD POST: UI- Proc Comm ListTitle = If the HD is not communicating with the UI during POST. Message = UI POST FW Communication failure [102] Title = Cleaning: Tpo THD Temp Delta ListTitle = If temperature gradient between Tpo and Thd exceeds maximum 15°C. Message = DG heat disinfection temperature gradient out of range SRSDG 856 SRSDG 930 = [103] Title = HD POST: Cal NVRAM CRC ListTitle = If the computed CRC of the calibration group record is not equal to the record CRC. Message = HD non-volatile calibration group invalid record CRC [104] Title = HD Fault: Air Trap Sensor ListTitle = If the air trap level readings are not valid for more than 2 seconds. Message = Air trap level sensors reporting illegal combination of air/fluid [105] Title = DG POST: Cal NVRAM CRC ListTitle = If the DG NVDATAMGMT CAL group calculated CRC does not matched the expected CRC. Message = DG non-volatile system group invalid record CRC SRSDG 608 = [106] Title = DG POST: Flow Sensor Cal ListTitle = If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. Message = DG flow sensors calibration error SRSDG 820 SRSDG 821 SRSDG 1087 = [107] Title = DG Fault: DG Restarted ListTitle = If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. Message = HD reports DG restarted fault [108] Title = HD Fault: Syringe Pump ADC Com ListTitle = If the ADC read is not fresh after POST init and the ADC read counter has not changed. Message = HD syringe pump ADC error [109] Title = HD Fault: Syringe Pump Vol ListTitle = If the volume delivered is off by more than 5% or 0.1mL. Message = HD syringe pump volume check error [110] Title = HD Fault: Syringe Pump Speed ListTitle = If the syringe measured rate is off by more than 5% or 0.1mL. Message = HD syringe pump speed check error [111] Title = HD Fault: Syringe Pump Speed ListTitle = If the syringe pump position is not equal to its last position. Message = HD syringe pump not stopped in off state error [112] Title = DG Fault: LC A2/B2 Comm ListTitle = If the FPGA does not report fresh data for a certain period of time. Message = DG load cells A2/B2 FPGA fault SRSDG 953 = [113] Title = DG Fault: DRP Current ListTitle = If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. Message = DG drain pump current out of range SRSDG 890 = [114] Title = Process: Pre- Tx Wet Self- Test ListTitle = If the venous air bubble self test did not detect bubble for more than 500 msec. Message = HD venous air bubble detector self-test failure [115] Title = DG Fault: Temp Sensor Range ListTitle = If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. Message = DG temperature sensor out of range SRSDG 224 = [116] Title = HD Alarm (Low): SD- Card Error ListTitle = SD card is non-operational Message = SD card is non-operational [117] Title = Loss of HD Comm (HD ACK) ListTitle = If no retries left in the pending acknowlegement list. Message = System message that required acknowledgment was not acknowledged SRSDG 909 = [118] Title = DG Fault: RTC Config ListTitle = If the RTC configuration registers report an error. Message = DG RTC configuration error SRSDG 43 = [119] Title = DG Alarm (Low): RTC Bat Low ListTitle = If the RTC battery bit report a low alarm. Message = DG RTC battery low SRSDG 908 = [120] Title = Process: Pre- Tx Wet Self- Test ListTitle = If the HD pre-treatment wet self test failed. Message = HD pre-treatment wet prime test failure [121] Title = DG Fault: Prim Heater Lg Control ListTitle = If the main primary heater's voltage is out of range. Message = DG main primary heater voltage out of range SRSDG 611 = [122] Title = DG Fault: Prim Heater Sm Control ListTitle = If the small primary heater's voltage is out of range. Message = DG small primary heater voltage out of range SRSDG 612 = [123] Title = DG Fault: Trimmer Heater Control ListTitle = If the trimmer voltage is out of range. Message = DG trimmer heater voltage out of range SRSDG 613 = [124] Title = Blood Stop: Blood Clot Risk ListTitle = In treatment end paused state for > 5 minutes. Message = HD end of treatment alarm (high priority - user still not acting to end treatment) [125] Title = Blood Stop: Blood Clot Risk ListTitle = If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. Message = Blood sitting too long after treatment stopped by user alarm (> 5 min) [126] Title = Blood Stop: Dialyzer Blood Leak ListTitle = If the blood leak is detected based on the blood leak status reported by the sensor. Message = HD blood leak detected alarm [127] Title = Blood Stop: Venous Pres. Low ListTitle = If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. Message = Venous pressure too low during treatment [128] Title = Blood Stop: Venous Air Bubble ListTitle = If bubbles are detected during the treatment. Message = HD venous air bubble detected alarm [129] Title = Blood Stop: Dialyzer Blood Leak Bypass ListTitle = If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. Message = HD blood leak detected and it is in the process of attempting to remove the bubbles/foam [130] Title = Blood Stop: Venous Pres. High ListTitle = If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. Message = Venous pressure too high during treatment [131] Title = Blood Stop: Arterial Pres. Low ListTitle = If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. Message = Arterial pressure too low during treatment [132] Title = Blood Stop: Arterial Pres. High ListTitle = If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. Message = Arterial pressure too high during treatment [133] Title = DG Fault: Fluid Leak in DG ListTitle = If the fluid leak is detected as reported by the fluid leak detector. Message = DG fluid leak detected alarm SRSDG 328 = [134] Title = Blood Stop: Fluid Leak in HD ListTitle = If the fluid leak is detected as reported by the fluid leak detector. Message = HD fluid leak detected alarm [135] Title = Blood Stop: HD Shock Detected ListTitle = If the system experienced excessive shock > 2.5 g. Message = HD experienced large acceleration [136] Title = Dial. Gen: Shock- Vibration ListTitle = If the system experienced excessive shock > 2.5 g. Message = DG experienced large acceleration SRSDG 327 = [137] Title = Blood Stop: HD Tilted ListTitle = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. Message = HD tilt exceeds maximum [138] Title = Dial. Gen: Tilt ListTitle = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. Message = DG tilt exceeds maximum SRSDG 326 = [139] Title = Loss of AC Power ListTitle = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. Message = HD AC power lost alarm [140] Title = Loss of DG Comm (Timeout) ListTitle = If DG communication timeout is detected for more than 2 seconds. Message = Dialysate generator communication timeout [141] Title = Blood Stop: Air Trap Fill ListTitle = If the air trap fill timeout during treatment. Message = Air trap fill timeout during treatment [142] Title = Blood Stop: Blood Pump Occlusion ListTitle = If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall Message = Blood pump occlusion detected [143] Title = DG Fault: TRo TDi Delta ListTitle = If TDi and TRo have drifted for more 1 C for a certain period of time. Message = DG dialysate temperature sensors drift timeout. SRSDG 912 = [144] Title = Cleaning: Inlet Water Cond High ListTitle = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode Message = DG cleaning mode inlet water conductivity out of high range SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] Title = Dial. Gen: Acid Conc. Cond. ListTitle = If the conductivity value for the acid concentrate is out of range for more than 45 secs. Message = DG Concentrate conductivity after adding acid out of range alarm SRSDG 846 = [146] Title = DG POST: DG RTC Timer ListTitle = If RTC and timer are not aligned on 1 second elapsed time. Message = DG RTC vs. Timer accuracy error SRSDG 312 = [147] Title = Dial. Gen: Dialysate not ready ListTitle = If dialysate conductivity is not with 5% of the TCD Message = DG dialysate generation conductivity fault alarm SRSDG 398 = [148] Title = Cleaning: Inlet Water Temp High ListTitle = Inlet temperature > 37 C. Message = Cleaning Mode Inlet water temperature too high SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] Title = HD Fault: WD Error ListTitle = If the watchdog signal is low while not in POST. Message = HD watchdog expired [150] Title = Dial. Gen: Inlet Water Cond High ListTitle = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). Message = Inlet water conductivity too high SRSDG 391 = [151] Title = Dial. Gen: Inlet Water Cond Low ListTitle = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). Message = Inlet water conductivity too low SRSDG 826 = [152] Title = Dial. Gen: Inlet Water Pres Low ListTitle = Inlet water pressure < 25 PSI. Message = Inlet water pressure too low SRSDG 417 = [153] Title = Process: Pre- Tx Terminated ListTitle = In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) Message = HD prime completed high priority alarm [154] Title = Cleaning: Inlet Water Cond Low ListTitle = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode Message = Cleaning mode inlet water conductivity too low SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] Title = Loss of DG Comm (Reservoir) ListTitle = No new reservoir data for 2 seconds. Message = HD No reservoir data message receive at least once every 2 seconds [156] Title = Loss of DG Comm (Op Mode) ListTitle = No DG op mode for 2 seconds. Message = HD No DG op mode message receive at least once every 2 seconds [157] Title = Cleaning: Chem Disinf PAA Prime ListTitle = If the disinfectant acid line is not primed within a certain period of time. Message = Chemical disinfect priming the acid line timeout SRSDG 988 = [158] Title = Dial. Gen: Inlet Water Temp Low ListTitle = Inlet temperature < 24 °C for 5 seconds. Message = Inlet water temperature low SRSDG 408 = [159] Title = Cleaning: Chem Disinf Connect Cap ListTitle = Start of chemical disinfect flush. Message = Close concentrate line be able to flush it SRSDG 1044 = [160] Title = HD Fault: BP Rotor Max Speed ListTitle = If blood pump motor speed exceeded 100.0 RPM. Message = Blood pump rotor speed too high [161] Title = DG Fault: HD- DG Comm ListTitle = If >= 5 bad CRCs detected on received CAN messages within 10 minutes. Message = Too many bad CRCs detected on received system messages SRSDG 122 = [162] Title = DG Fault: Proc- FPGA Clock ListTitle = If the FPGA clock speed is relatively close to the processor clock for a certain period of time. Message = FPGA clock speed failure SRSDG 867 = [163] Title = Dial. Gen: UF Rate Check ListTitle = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). Message = HD sees primary load cell for reservoir 1 change too much too fast [164] Title = Dial. Gen: UF Rate Check ListTitle = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). Message = HD sees primary load cell for reservoir 2 change too much too fast [165] Title = Blood Stop: Blood Clot Risk ListTitle = If the maximum time inrinseback operation has been exceeded. Message = HD in treatment rinseback operation timeout [166] Title = ListTitle = Message = [167] Title = Blood Stop: Door Open ListTitle = If the cartridge door is opened as reported by the FPGA. Message = HD cartridge door opened alarm [168] Title = ListTitle = Message = [169] Title = DG Fault: Dialysate Flow Sens Rg ListTitle = If the dialysate flow rate > 2.00 L/min. Message = Dialysate flow rate out of range SRSDG 643 = [170] Title = Heparin Syringe Empty ListTitle = If the syringe pump position is near empty position. Message = HD syringe empty alarm [171] Title = Process: Pre- Tx Hep. Pump Occ. ListTitle = If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. Message = HD syringe pump occlusion alarm [172] Title = Process: Pre- Tx Not Enough Hep. ListTitle = If the estimated syringe volume needed for treatment is not sufficient Message = HD syringe pump not enough Heparin alarm [173] Title = HD Fault: RTC Config ListTitle = If the RTC configuration registers report an error. Message = HD RTC configuration error [174] Title = HD POST: HD RTC Timer ListTitle = If RTC and timer are not aligned on 1 second elapsed time. Message = HD RTC vs. Timer accuracy error [175] Title = HD Fault: Pump Commutation Err ListTitle = If the set pump direction does not agree with the hall sensor direction. Message = HD pump commutation error [176] Title = ListTitle = Message = [177] Title = DG Fault: Software Invalid State ListTitle = If the specific software faults listed in alarm management SW faults has been detected Message = DG software fault. Software found itself in an unexpected state SRSDG 919 = [178] Title = HD Fault: UI- Proc Comm ListTitle = If the HD stops communicating with the UI. Message = HD communication timeout = SRSUI 1066 [179] Title = DG Fault: FPGA Com ListTitle = If retries for commands exceeds limit or FPGA reports comm error. Message = DG FPGA communication down for too long SRSDG 605 = [180] Title = DG Fault: RO Flow Sensor Range ListTitle = If the RO flow rate > 2.00 L/min. Message = RO flow rate out of range SRSDG 883 = [181] Title = DG Fault: Tare Out of Range ListTitle = If the load cell weight is out of range for tare. Message = DG load cells weight out of range for tare SRSDG 849 SRSDG 838 = [182] Title = DG POST: LC Cal ListTitle = If the CRC of load cells calibration data does not match the calculated CRC. Message = DG load cells invalid calibration SRSDG 162 = [183] Title = Cleaning: Invalid Resv Transfer ListTitle = If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. Message = DG invalid load cell value SRSDG 851 = [184] Title = DG Fault: UVi Error ListTitle = If the inlet reactor are not healthy. Message = Inlet reactor not healthy SRSDG 559 = [185] Title = DG Alarm (Low): Fan Speed ListTitle = If the fan RPM is out of Min/Max range. Message = DG fan RPM out of range SRSDG 628 = [186] Title = DG Fault: CP1/CP2 Fault ListTitle = If the concentrate pumps fault is reported by the FPGA. Message = DG concentrate pump fault SRSDG 891 = [187] Title = DG Fault: CP1 Speed Fault ListTitle = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. Message = DG concentrate pump CP1 speed control error SRSDG 905 = [188] Title = DG Fault: CP2 Speed Fault ListTitle = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. Message = DG concentrate pump CP2 speed control error SRSDG 906 = [189] Title = DG Fault: DRP Speed Fault ListTitle = If the drain pump RPM > 100.0 of the target RPM. Message = DG drain pump RPM out of range SRSDG 448 = [190] Title = DG Fault: DRP Speed Not Off ListTitle = If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. Message = DG drain pump off fault [191] Title = DG Fault: RO Flow High ListTitle = If the RO flow rate > 10% from the target flow rate. Message = DG flow rate out of upper range SRSDG 885 = [192] Title = DG Fault: RO Flow Low ListTitle = If the RO flow rate < 10% from the target flow rate. Message = DG flow rate out of lower range SRSDG 945 = [193] Title = ListTitle = Message = [194] Title = DG Fault: RO Driver Feedback ListTitle = If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. Message = DG RO pump duty cycle out of range SRSDG 881 = [195] Title = DG Fault: RO Pump High Pressure ListTitle = If the DG RO pump pressure > 135 psi for a certain period of time. Message = DG RO pump pressure out of range SRSDG 925 = [196] Title = DG Fault: CPi/CPo Comm ListTitle = If the FPGA reports error for a certain period of time. Message = DG CPi/CPo conductivity sensors FPGA fault SRSDG 955 = [197] Title = DG Fault: CD1/CD2 Comm ListTitle = If the FPGA reports error for a certain period of time. Message = DG CD1/CD2 conductivity sensors FPGA fault SRSDG 956 = [198] Title = DG Fault: HP On With No Flow ListTitle = If the RO flow is < 0.2 L/min for a certain period of time. Message = DG primary heater while RO flow is low SRSDG 872 = [199] Title = DG Fault: HR On With No Flow ListTitle = If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. Message = DG trimmer heater while dialysate flow is low SRSDG 871 = [200] Title = DG Fault: Thermistor Range ListTitle = If thermistor > 80.0 C or < 0.0 C. Message = DG thermistor or temperature sensor out of range SRSDG 178 = [201] Title = Process: Pre- Tx Wet Self- Test ListTitle = If the pre-treatment wet flow test has failed. Message = HD pre-treatment wet flow test failure [202] Title = Process: Pre- Tx Dry Self- Test ListTitle = If the pre-treatment dry pressure normal test failed. Message = HD pre-treatment dry pressure normal test failure [203] Title = ListTitle = Message = [204] Title = DG Fault: RO Rejection ListTitle = If RO rejection ratio > 0.1 Message = RO rejection ratio out of range SRSDG 429 = [205] Title = DG Fault: Cond Sensors Error ListTitle = If conductivity sensor reports 0x11 for more than 2 consecutive reads. Message = Conductivity sensor fault SRSDG 913 = [206] Title = DG Fault: Reservoir Fill Timeout ListTitle = If DG dialysate fill > 5 minutes. Message = DG dialysate fill runs out of time SRSDG 437 = [207] Title = DG Fault: FMP Flow Rate Error ListTitle = If integrated Volume To Load Cell Reading Percent > 0.1. Message = DG flow meter check failure alarm SRSDG 240 = [208] Title = Blood Stop: Venous Pres. High ListTitle = If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. Message = HD venous line occlusion fault [209] Title = Cleaning: Flush Circ Error ListTitle = If the integrated volume < 0.1 for more than 60 secs. Message = DG drain circulation line timeout SRSDG 980 = [210] Title = HD Alarm (Low): Battery Fault ListTitle = If the battery pack status error is detected from the slave device I2C address 0x0B. Message = HD battery pack detected an error [211] Title = Blood Stopped: Warning ListTitle = During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. Message = Blood sitting too long warning (> 4 min) [212] Title = Blood Stopped: Warning ListTitle = During treatment end paused state, if the blood is in the line for more than 4 minutes. Message = Blood sitting too long warning (> 4 min) [213] Title = Process: Pre- Tx Prime Complete ListTitle = In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) Message = HD prime completed medium priority alarm [214] Title = ListTitle = Message = [215] Title = ListTitle = Message = [216] Title = Process: End of Dialysis Tx ListTitle = If the treatment ended successfully. Message = HD end treatment sub-mode timeout alarm [217] Title = ListTitle = Message = [218] Title = Process: Remove Syringe ListTitle = If the syringe detection switch > 2.0 volts. Message = HD syringe detected alarm [219] Title = Blood Stop: Syringe Not Detected ListTitle = If the syringe detection switch < 2.0 volts. Message = HD syringe pump syringe removed alarm [220] Title = DG Fault: THD RTD Comm ListTitle = If the FPGA does not report fresh data for a certain period of time Message = DG THd sensor FPGA fault SRSDG 947 = [221] Title = Blood Stop: Saline Bag Empty ListTitle = If the measured arterial pressure < -300.0 mmHg for more than 250 msec. Message = HD Saline bag is empty [222] Title = HD Fault: BP Occ Com ListTitle = If the FPGA does not report fresh data or indicates an error for a certain period of time. Message = HD occlusion sensor FPGA fault [223] Title = HD Fault: Art. Pres. Com ListTitle = If the FPGA does not report fresh data for a certain period of time Message = HD arterial sensor FPGA fault [224] Title = Blood Stopped: Treatment Pause ListTitle = If the user has pressed the stop button. Message = Treatment stopped by user action - pressed stop button [225] Title = Process: End of Treatment ListTitle = If the treatment duration has exceeded the prescribed treatment time entered by the user. Message = HD end of treatment warning [226] Title = Process: Pre- Tx Priming Complete ListTitle = Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection Message = HD prime completed low priority alarm [227] Title = ListTitle = Message = [228] Title = Process: Pre- Tx Prime Timeout ListTitle = During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. Message = HD time out on prime sub-mode purge air state. [229] Title = Process: Pre- Tx Prime Timeout ListTitle = If the priming of the dialyzer dialysate flow path has exceeded the time limit. Message = HD prime dialysate dialyzer time out alarm [230] Title = Process: Pre- Tx Prime Timeout ListTitle = If the priming of the dialyzer bypass flow path has exceeded the time limit. Message = HD prime dialysate bypass time out alarm [231] Title = Process: Pre- Tx Dry Self- Test ListTitle = If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState Message = HD pre-treatment mode dry self-tests failure [232] Title = Process: Pre- Tx Wet Self- Test ListTitle = If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState Message = HD pre-treatment mode wet self-tests failure [233] Title = DG Fault: Baro Cal CRC ListTitle = If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. Message = DG barometric pressure sensor bad coefficients CRC SRSDG 964 SRSDG 949 SRSDG 1089 = [234] Title = Dial. Gen: Inlet Water Pres High ListTitle = If the inlet water pressure is > 80 psig. Message = DG inlet water pressure is too high SRSDG 1037 = [235] Title = Process: Rinseback Complete ListTitle = If there is no active alarm and rinseback has completed. Message = HD in treatment stopped sub-mode after rinseback completed (no escalation) [236] Title = Process: Pre- Tx Dry Self- Test ListTitle = If the air trap sensor reported AIR and bubble sensor detected bubbles. Message = HD needs new cartridge to be installed [237] Title = ListTitle = Message = [238] Title = Process: Pre- Tx Dry Self- Test ListTitle = If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. Message = HD no cartridge loaded or installed improperly alarm [239] Title = Process: Post- Tx Remove Cartridge ListTitle = In the post treatment state, if the cartridge has not been removed. Message = HD fail to remove cartridge alarm [240] Title = Dial. Gen: Bicarb Conc. Cond. ListTitle = If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. Message = Bicarb conductivity out of range during bicarb pump check alarm SRSDG 482 = [241] Title = Cleaning: Resv Drain Timeout ListTitle = If the DG reservoir drain is not complete in 120 secs. Message = DG reservoir drain time out SRSDG 848 SRSDG 90 SRSDG 561 SRSDG 577 SRSDG 590 SRSDG 591 SRSDG 336 SRSDG 879 SRSDG 340 SRSDG 496 SRSDG 501 SRSDG 505 SRSDG 532 SRSDG 535 SRSDG 556 SRSDG 557 SRSDG 1048 SRSDG 1049 SRSDG 651 SRSDG 655 SRSDG 744 SRSDG 745 SRSDG 1004 SRSDG 1008 SRSDG 1019 SRSDG 1023 SRSDG 1033 SRSDG 1034 = [242] Title = Cleaning: Resv Fill Timeout ListTitle = If the DG reservoir fill is not complete in 120 secs. Message = DG reservoir fill time out SRSDG 850 SRSDG 573 SRSDG 574 SRSDG 579 SRSDG 580 SRSDG 492 SRSDG 493 SRSDG 498 SRSDG 499 SRSDG 508 SRSDG 509 SRSDG 519 SRSDG 690 SRSDG 691 SRSDG 703 SRSDG 711 SRSDG 1020 SRSDG 1024 = [243] Title = Cleaning: Fluid Loss Detected ListTitle = If the reservoir level is drifted by 100 mL for more than 30 seconds. Message = DG reservoir leak time out SRSDG 853 = [244] Title = Cleaning: Temp Sensor Check ListTitle = If any of the temperature sensors are out of the average of the sensors by more than 3 C. Message = DG temperature sensors difference out of range SRSDG 852 = [245] Title = Cleaning: Heat Disinfect Temp TO ListTitle = If heating up to minimum temperature for heat disinfect failed after 4 hours. Message = DG heat disinfect target temperature time out (could not reach to temperature) SRSDG 854 = [246] Title = Cleaning: RO Water Cond Error ListTitle = If CPo and CD2 conductivity values are above 100 uS/cm. Message = DG chemical disinfect inlet conducitivty out of range [247] Title = ListTitle = Message = [248] Title = Cleaning: Chem Disinfect Proc ListTitle = If TPo > 50 C after 1 second. Message = DG chemical disinfect out of target temperature [249] Title = Cleaning: Chem Disinf PAA Cond ListTitle = If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. Message = DG chemical disinfect out of target conductivity SRSDG 992 = [250] Title = Cleaning: Chem Disinf Connect PAA ListTitle = Start of chemical disinfect Message = Insert the acid to start chemical disinfect SRSDG 1046 = [251] Title = HD POST: Sys Rec NVRAM CRC ListTitle = If the calculated CRC of the HD system record image does not match the expected CRC. Message = HD invalid system record [252] Title = HD POST: Serv Rec NVRAM CRC ListTitle = If the calculated CRC of the HD service record image does not match the expected CRC. Message = HD invalid service record [253] Title = DG POST: Sys Rec NVRAM CRC ListTitle = If the calculated CRC of the DG system record image does not match the expected CRC. Message = DG invalid system record SRSDG 609 = [254] Title = DG POST: Serv Rec NVRAM CRC ListTitle = If the calculated CRC of the DG service record image does not match the expected CRC. Message = DG invalid service record SRSDG 610 = [255] Title = HD POST: UI Compatibility ListTitle = If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. Message = HD and UI software builds are not compatible [256] Title = Cleaning: Chem Flush In Progress ListTitle = If DG flush successfully started from the UI. Message = DG flush started = SRSUI 497 [257] Title = ListTitle = Message = [258] Title = ListTitle = Message = [259] Title = HD Fault: Temp Sensors Error ListTitle = If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. Message = HD temperatures out of range [260] Title = HD POST: UI Image ListTitle = If the UI POST reports a shasum faillure. Message = UI POST Application Integrity (Sha256Sum) failure = SRSUI 1067 [261] Title = HD POST: UI CAN Driver ListTitle = If the UI POST reports a can bus failure. Message = UI POST CANBus failure = SRSUI 1068 [262] Title = ListTitle = If the UI POST reports a display failure. Message = UI POST Display failure [263] Title = HD POST: UI Touch Driver ListTitle = If the UI POST reports a touch failure. Message = UI POST Touch failure = SRSUI 1070 [264] Title = HD POST: UI SD- Card ListTitle = If the UI POST reports SD card failure. Message = UI POST SD-Card failure = SRSUI 1071 [265] Title = HD POST: UI RTC ListTitle = If the UI POST reports RTC failure. Message = UI POST RTC failure = SRSUI 1072 [266] Title = HD POST: UI WiFi ListTitle = If the UI POST reports WiFi failure. Message = UI POST WiFi failure = SRSUI 1073 [267] Title = HD POST: UI Bluetooth ListTitle = If the UI POST reports bluetooth failure. Message = UI POST Bluetooth failure = SRSUI 1074 [268] Title = HD POST: UI Ethernet ListTitle = If the UI POST reports ethernet failure. Message = UI POST Ethernet failure = SRSUI 1075 [269] Title = HD POST: UI Sound ListTitle = If the UI POST reports sound failure. Message = UI POST Sound failure = SRSUI 1076 [270] Title = HD POST: Proc Safestate ListTitle = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. Message = HD POST Safety Shutdown failure [271] Title = DG POST: Proc Safestate ListTitle = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. Message = DG POST Safety Shutdown failure SRSDG 822 = [272] Title = HD Alarm (Low): Fan Speed ListTitle = If the fan RPM is out of Min/Max range. Message = HD Fan RPM out of range [273] Title = ListTitle = Message = [274] Title = ListTitle = Message = [275] Title = HD Fault: Arterial Press Range ListTitle = If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. Message = HD arterial pressure sensor is reading out of range [276] Title = HD Fault: Venous Press Range ListTitle = If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. Message = HD venous pressure sensor is reading out of range [277] Title = HD Fault: Occlusion Sens Range ListTitle = If the blood pump occlusion count is < 2000. Message = HD BP occlusion sensor is reading out of range [278] Title = DG Fault: Reservoir Overfill ListTitle = If the active reservoir weight is > 1900.0 g. Message = HD active reservoir weight out of range. [279] Title = DG Fault: Reservoir Drain ListTitle = If the DG dialysate drain exceeded 120 seconds. Message = DG dialysate drain time out SRSDG 752 = [280] Title = DG Fault: DRP Direction Fault ListTitle = If the drain pump's direction is not 1 when draining. Message = DG drain pump invalid direction SRSDG 888 = [281] Title = Dial. Gen: Acid Conc. Low ListTitle = If the acid concentration bottle volume > 3240 (3340 - 100). Message = DG Acid concentration bottle low volume alarm SRSDG 836 = [282] Title = Dial. Gen: Bicarb Conc. Low ListTitle = If the bicarbonate concentration bottle volume > 3680 (3780 - 100). Message = DG Bicarbonate concentration bottle low volume alarm SRSDG 837 = [283] Title = DG Fault: LC Range ListTitle = If the weight of the DG load cells > 4500.0 g or < 0.0 g. Message = DG load cells weight out of range SRSDG 615 = [284] Title = DG Fault: Redundant LC Check ListTitle = If redundant load cells deviate by more than 60g after tare for 5 seconds. Message = DG load cells primary/back up drift out of range SRSDG 383 = [285] Title = Cleaning: RO Permeate Sample In Progress ListTitle = If DG RO permeate sample successfully started from UI. Message = DG RO permeate sample started [286] Title = Cleaning: Conc Cap Removed ListTitle = If the concentrate cap is not in the right position in the cleaning modes. Message = DG cap in wrong state [287] Title = HD Fault: Proc- FPGA Clock ListTitle = If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. Message = HD processor clock speed checks against FPGA clock failure [288] Title = DG Fault: LC Drift Check ListTitle = If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. Message = HD load cell inconsistency [289] Title = Cleaning: Dialysate Cap Removed ListTitle = If the dialysate cap is not in the right position in the cleaning modes. Message = DG cap in wrong state SRSDG 928 = [290] Title = HD Fault: Ven. Pres. Com ListTitle = If the FPGA does not report fresh data for a certain period of time Message = HD venous sensor FPGA fault [291] Title = Cleaning: Flush In Progress ListTitle = If DG flush successfully started from the UI. Message = DG flush started [292] Title = Cleaning: Heat Disinf In Progress ListTitle = If DG heat disinfect successfully started from the UI. Message = DG heat disinfect started [293] Title = Cleaning: Chem Disinfect In Progress ListTitle = If DG flush successfully started from the UI. Message = DG chem disinfect started [294] Title = Cleaning: Active Cool In Progress ListTitle = If DG active cool successfully started. Message = DG active cool started [295] Title = HD POST: UI CloudSync ListTitle = If the UI POST reports cloud sync failure. Message = UI cloud sync POST failure = SRSUI 1238 [296] Title = DG Fault: TDi RTD Comm ListTitle = If the FPGA does not report fresh data for a certain period of time Message = DG TDi sensor FPGA fault SRSDG 951 = [297] Title = DG Fault: TRo RTD Comm ListTitle = If the FPGA does not report fresh data for a certain period of time Message = DG TRo sensor FPGA fault SRSDG 950 = [298] Title = DG Fault: Baro Pres Comm ListTitle = If the FPGA does not report fresh data for a certain period of time Message = DG barometric sensor FPGA fault SRSDG 948 = [299] Title = DG POST: Serial Number Check ListTitle = If the DG serial number is empty and does not have any characters. Message = DG invalid serial number SRSDG 952 = [300] Title = HD POST: Serial Number Check ListTitle = If the HD serial number is empty and does not have any characters. Message = HD invalid serial number [301] Title = Cleaning: RO Permeate Sample Connect Cap ListTitle = RO permeate sample collection state Message = Open dialysate cap to be able to collecte the dispensing fluid [302] Title = DG Fault: Processor RAM error ListTitle = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. Message = DG_CPU RAM error SRSDG 126 = [303] Title = DG Fault: DRP Hall Commutation ListTitle = If the FPGA indicates an error for a certain period of time. Message = DG drain pump direction FPGA fault SRSDG 958 = [304] Title = HD POST: Usage Rec NVRAM CRC ListTitle = If the CRC of the HD usage record does not match the calculated CRC. Message = HD invalid usage reocord [305] Title = HD Fault: Processor RAM error ListTitle = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. Message = HD_CPU RAM error [306] Title = DG Fault: CD1 CD2 Check ListTitle = CD1 and CD2 conductivity measurement during acid check is not within 5% of each other Message = [307] Title = Blood Stop: Loss of AC Power ListTitle = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. Message = HD AC power lost alarm [308] Title = DG Fault: CPi Comm ListTitle = If the CPi conductivity sensor transmits invalid characters for a certain period of time. Message = DG CPi conductivity sensor transmits invalid characters SRSDG 959 = [309] Title = DG Fault: CPo Comm ListTitle = If the CPo conductivity sensor transmits invalid characters for a certain period of time. Message = DG CPo conductivity sensor transmits invalid characters SRSDG 960 = [310] Title = DG Fault: CD1 Comm ListTitle = If the CD1 conductivity sensor transmits invalid characters for a certain period of time. Message = DG CD1 conductivity sensor transmits invalid characters SRSDG 962 = [311] Title = DG Fault: CD2 Comm ListTitle = If the CD2 conductivity sensor transmits invalid characters for a certain period of time. Message = DG CD2 conductivity sensor transmits invalid characters SRSDG 961 = [312] Title = Dial. Gen: Dialysate In Flow ListTitle = If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. Message = The dialysate inlet flow is too low. [313] Title = Cleaning: Inlet Water Temp Low ListTitle = If inlet water temperature is < 24 C. Message = DG cleaning mode inlet water temperature too low SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] Title = Cleaning: Inlet Water Press High ListTitle = If inlet water temperature is > 80 psig. Message = DG cleaning mode inlet water pressure too high SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] Title = Cleaning: Inlet Water Press Low ListTitle = If inlet water pressure is < 14 psig. Message = DG cleaning mode inlet water pressure too low SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] Title = DG Fault: Acid CP Pump Park ListTitle = If the park took more than 1 second. Message = Acid concentrate pump did not park successfully SRSDG 983 = [317] Title = DG Fault: Bicarb CP Pump Park ListTitle = If the park took more than 1 second. Message = Bicarb concentrate pump did not park successfully SRSDG 984 = [318] Title = HD Fault: Air Pump Timeout ListTitle = If air pump has been turned on for greater than 10s. Message = HD Air Pump operation timeout [319] Title = Cleaning: Chem Residual Sampling ListTitle = Chemical disinfect flush sample alarm is prompted. Message = Collect sample to check the chemical disinfectant acid SRSDG 1040 = [320] Title = Cleaning: Chem Residual Timeout ListTitle = Chemical disinfect flush sample was not collected form more than 30 minutes. Message = The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes SRSDG 1043 = [321] Title = DG Fault: UVo Error ListTitle = If the outlet reactor are not healthy. Message = Outlet reactor not healthy SRSDG 1069 = [322] Title = HD POST Alarm: UI Invalid Year ListTitle = If the UI POST reports a invalid year. Message = UI POST Invalid Year [323] Title = HD POST Alarm: UI Bad Settings ListTitle = If the UI POST reports a bad settings. (e.g. no settings read, empty settings file, invalid/empty parms, CRC failure) Message = UI POST Bad Settings