This is a list of all comments for DG-DEN-16093-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Services/SystemCommMessages.c Revision Comment by Sean Nash on 21 August 2023, 13:48 https://devapps.diality.us/cru/DG-DEN-16093-1#c18659 Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting). Function would also create and send the cmd response message and return TRUE/FALSE which would set local result here for the ack response message below. We should reject cmd if pumps are busy/on and/or if current DG mode is fill or fault mode. Reply by Vinayakam Mani on 21 August 2023, 18:05 > required changes are done. Reply by Sean Nash on 22 August 2023, 10:07 > RESOLVED in CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Controllers/ConcentratePumps.c Revision Comment by Sean Nash on 22 August 2023, 09:57 https://devapps.diality.us/cru/DG-DEN-16093-1#c18667 #includes are sorted. Put OperationModes.h in its proper place. Reply by Sean Nash on 22 August 2023, 11:04 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 22 August 2023, 09:58 https://devapps.diality.us/cru/DG-DEN-16093-1#c18668 Add a blank line between declarations and code. Reply by Sean Nash on 22 August 2023, 10:56 > RESOLVED in CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Modes/ModeFill.c Revision Comment by Sean Nash on 25 August 2023, 14:24 https://devapps.diality.us/cru/DG-DEN-16093-1#c18742 100% now. Reply by Vinayakam Mani on 25 August 2023, 15:03 > Done. Reply by Sean Nash on 28 August 2023, 10:16 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 25 August 2023, 14:25 https://devapps.diality.us/cru/DG-DEN-16093-1#c18743 Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is reduced, this level should also be reduced accordingly." Reply by Vinayakam Mani on 25 August 2023, 15:03 > Done. Reply by Sean Nash on 28 August 2023, 10:16 > RESOLVED in CODE WALKTHROUGH. --- ID: DG-DEN-16093-1 https://devapps.diality.us/cru/DG-DEN-16093-1 Title: DG-DEN-16093_Alarm 97 Conductivity Out OF Range Triggered ON First Fill OF Treatment Statement of Objectives: State: Closed Summary: Author: Vinayakam Mani Moderator: Vinayakam Mani Reviewers: (4 active, 3 completed*) Sean Nash (*) wbracken (*) jtaylor (*) Michael Garthwaite Dara Navaei Darren Cox Steve Jarpe