This is a list of all comments for DG-DEN-16280-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/ConductivitySensors.c Revision Comment by Sean Nash on 24 August 2023, 11:25 https://devapps.diality.us/cru/DG-DEN-16280-1#c18724 We don't trigger inlet water conductivity alarm in post-treatment mode? Any reason why we're not just saying "== MODE_POST"? What about inlet water pressure/temperature? FYI - getHDOperationMode() was not working - HD op mode handler was mishandling message - fixed in my DENBUG-59 branch, but not yet in staging. Reply by jtaylor on 24 August 2023, 13:58 > This was a specific change for this alarm (150), and the > associated alarms in this routine. It appeared there were > similar mode limits for temperature as well. Pressure limits > were more uniquely related to valve position, I thought. > > I had seen the related changes in 59, and included them in my > develop build. Reply by Sean Nash on 25 August 2023, 12:42 > As we discussed, I think alarm 150 and related inlet water > alarms are already set to block triggering during rinseback > and post-treatment modes. So these changes are likely not > needed. --- ID: DG-DEN-16280-1 https://devapps.diality.us/cru/DG-DEN-16280-1 Title: DG-DEN-16280_Stuck IN Post Treatment When Inlet Water Alarm Appears Statement of Objectives: State: Closed Summary: Author: jtaylor Moderator: jtaylor Reviewers: (6 active, 0 completed*) Sean Nash Michael Garthwaite wbracken Dara Navaei Darren Cox Steve Jarpe