This is a list of all comments for DG-DEN-16680-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/ModeHeatDisinfect.c Revision Comment by Vinayakam Mani on 28 June 2024, 16:48 https://devapps.diality.us/cru/DG-DEN-16680-1#c19790 Will there be any worst-case scenario that the temperature never come down (say below 60) and get into a cyclic loop between cancel path and cool idle state? Reply by Dara Navaei on 01 July 2024, 09:21 > If it does not come down below 60 and in the cool idle cycle > if any issue occurs, the DG Software transitions to the > cancel basic path. Reply by Vinayakam Mani on 01 July 2024, 09:47 > RESOLVED IN CODEWALKTHROUGH. Revision Comment by Sean Nash on 01 July 2024, 17:00 https://devapps.diality.us/cru/DG-DEN-16680-1#c19801 This flag is really indicating drain water is not too hot OR we are not doing a nocturnal heat disinfect. Flag name could be better. Reply by Dara Navaei on 02 July 2024, 08:55 > Done Reply by Sean Nash on 03 July 2024, 12:13 > RESOLVED in CODE WALKTHROUGH ---------------------------------------- File: firmware/App/Services/SystemCommMessages.c Revision Comment by Sean Nash on 01 July 2024, 17:05 https://devapps.diality.us/cru/DG-DEN-16680-1#c19802 Is this necessary in else? If so, can move above or below the if so you just do it once. Reply by Dara Navaei on 02 July 2024, 08:51 > Removed the one in else. Reply by Sean Nash on 03 July 2024, 17:23 > RESOLVED in CODE WALKTHROUGH. ---------------------------------------- File: AlarmDefs.h Revision Comment by Vinayakam Mani on 28 June 2024, 16:19 https://devapps.diality.us/cru/DG-DEN-16680-1#c19789 Looks ALARM_PRIORITY_LOW ranks are in 700 to 800 series range. Medium priority rank ranging 400 to 600. Do you want to consider this as low priority or medium priority? if medium priority rank, can we get rank in to 400 to 600 series? Reply by Dara Navaei on 01 July 2024, 09:27 > Thanks I meant medium priority. Changed the rank to 500. Reply by Vinayakam Mani on 01 July 2024, 09:47 > RESOLVED IN CODEWALKTHROUGH. Revision Comment by Vinayakam Mani on 28 June 2024, 16:13 https://devapps.diality.us/cru/DG-DEN-16680-1#c19788 Do we need "\n" after the " Turn of Inlet water valves."? Reply by Dara Navaei on 01 July 2024, 09:20 > Removed \n Reply by Vinayakam Mani on 01 July 2024, 09:48 > RESOLVED IN CODEWALKTHROUGH. ---------------------------------------- File: DGDefs.h Revision Comment by Sean Nash on 01 July 2024, 16:52 https://devapps.diality.us/cru/DG-DEN-16680-1#c19800 This could just be an event at start. Reply by Dara Navaei on 02 July 2024, 09:02 > Added an event. Reply by Sean Nash on 03 July 2024, 09:50 > RESOLVED in CODE WALKTHROUGH. --- ID: DG-DEN-16680-1 https://devapps.diality.us/cru/DG-DEN-16680-1 Title: DG-DEN-16680_P1B Inlet Water Shutdown During Heat Disinfect Staging Statement of Objectives: State: Closed Summary: Author: Dara Navaei Moderator: Dara Navaei Reviewers: (2 active, 2 completed*) Sean Nash (*) Vinayakam Mani (*) jpaguio Michael Garthwaite