This is a list of all comments for DIALIN-DEN-15196-7. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/OperationModes.h Revision Comment by Sean Nash on 14 June 2023, 11:15 https://devapps.diality.us/cru/DIALIN-DEN-15196-7#c17884 Align comment. Reply by Michael Garthwaite on 14 June 2023, 11:21 > Fixed. Thanks! Reply by Sean Nash on 14 June 2023, 11:41 > RESOLVED in CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Modes/Dialysis.c Revision Comment by Sean Nash on 14 June 2023, 11:17 https://devapps.diality.us/cru/DIALIN-DEN-15196-7#c17885 Why does this situation require sending the status event here (instead of in op modes where this is usually done)? Reply by Michael Garthwaite on 14 June 2023, 11:18 > We need to display the last Saline event before we switch > back. > > Due to how we exit this state machine, we call > transitionToDialysis() if we go to > SALINE_BOLUS_STATE_MAX_DELIVERED which will update the > substate back to the current UF state. Reply by Sean Nash on 14 June 2023, 11:40 > RESOLVED in CODE WALKTHROUGH. --- ID: DIALIN-DEN-15196-7 https://devapps.diality.us/cru/DIALIN-DEN-15196-7 Title: DIAL-262-fixes Statement of Objectives: * DIAL-262: fixed operation status event to have more accurate transitioning in between dialysis sub modes. * DIAL-262: removed old 4th state data when hd pauses in priming * DIAL-262: added additional event calls for UF and saline transitions. * DIAL-262: fixes * Merge branch 'staging' into DEN-15196-DIAL-262-fixes * DIAL-264: fixed blood prime not showing sub state in operation status event State: Closed Summary: Author: Michael Garthwaite Moderator: Michael Garthwaite Reviewers: (4 active, 1 completed*) Sean Nash (*) wbracken Dara Navaei Darren Cox jtaylor