This is a list of all comments for HD-DEN-13154-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/SyringePump.c Revision Comment by Michael Garthwaite on 07 March 2023, 16:32 https://devapps.diality.us/cru/HD-DEN-13154-1#c16611 Should we check for stall in bolus & continuous states? Reply by Sean Nash on 08 March 2023, 09:35 > Concern was originally just for retract due to highest speed. > We ended up reducing retract speed to reduce likelihood of > stalls. Still a good idea to check for it though and since > retract, preload and seek all share same relatively high > speed, these are the ops that have the check. For bolus and > continuous, should not stall at these speeds and would catch > with rate alarm if it did. Reply by Michael Garthwaite on 08 March 2023, 09:47 > RESOLVED IN CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Controllers/BloodFlow.c Revision Comment by Michael Garthwaite on 01 March 2023, 16:04 https://devapps.diality.us/cru/HD-DEN-13154-1#c16557 This will hard fault the HD. Please change to getTreatmentParameterU32() Reply by Sean Nash on 06 March 2023, 11:37 > Fixed. Reply by Michael Garthwaite on 07 March 2023, 16:34 > RESOLVED IN CODE WALKTHROUGH. Revision Comment by Michael Garthwaite on 07 March 2023, 16:25 https://devapps.diality.us/cru/HD-DEN-13154-1#c16609 Shouldnt this be set to 0U instead of 0.0F? Reply by Sean Nash on 08 March 2023, 09:41 > Fixed. Reply by Michael Garthwaite on 08 March 2023, 09:47 > RESOLVED IN CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Controllers/BloodFlow.h Revision Comment by Michael Garthwaite on 07 March 2023, 16:24 https://devapps.diality.us/cru/HD-DEN-13154-1#c16608 This should be U32. Reply by Sean Nash on 08 March 2023, 09:33 > Fixed. Reply by Michael Garthwaite on 08 March 2023, 09:47 > RESOLVED IN CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Controllers/DialInFlow.c Revision Comment by Michael Garthwaite on 01 March 2023, 16:04 https://devapps.diality.us/cru/HD-DEN-13154-1#c16558 This will hard fault the HD. Please change to getTreatmentParameterU32() Reply by Sean Nash on 06 March 2023, 11:36 > Fixed. Reply by Michael Garthwaite on 07 March 2023, 16:33 > RESOLVED IN CODE WALKTHROUGH. Revision Comment by Michael Garthwaite on 07 March 2023, 16:34 https://devapps.diality.us/cru/HD-DEN-13154-1#c16614 Shouldnt this be set to 0U instead of 0.0F? Reply by Sean Nash on 08 March 2023, 09:30 > Fixed. Reply by Michael Garthwaite on 08 March 2023, 09:46 > RESOLVED IN CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Controllers/DialInFlow.h Revision Comment by Michael Garthwaite on 07 March 2023, 16:35 https://devapps.diality.us/cru/HD-DEN-13154-1#c16616 This should be U32. Reply by Sean Nash on 08 March 2023, 09:32 > Fixed. Reply by Michael Garthwaite on 08 March 2023, 09:47 > RESOLVED IN CODE WALKTHROUGH. ---------------------------------------- File: firmware/App/Controllers/DGInterface.c Revision Comment by Michael Garthwaite on 07 March 2023, 16:26 https://devapps.diality.us/cru/HD-DEN-13154-1#c16610 seems odd to create a variable just for it to be returned. Is this value supposed to be overridable? Reply by Sean Nash on 08 March 2023, 09:40 > Variable is also updated by receipts of DG flow broadcasts. > I believe it is overridable on DG side. Reply by Michael Garthwaite on 08 March 2023, 09:48 > RESOLVED IN CODE WALKTHROUGH. --- ID: HD-DEN-13154-1 https://devapps.diality.us/cru/HD-DEN-13154-1 Title: HD-DEN-13154_Heparin Syringe Empty Position Incorrect Statement of Objectives: State: Closed Summary: Author: Sean Nash Moderator: Sean Nash Reviewers: (4 active, 2 completed*) Michael Garthwaite (*) wbracken (*) Dara Navaei Darren Cox jtaylor Steve Jarpe