This is a list of all comments for HD-DEN-15839-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/Dialysis.c Revision Comment by Darren Cox on 25 July 2023, 12:49 https://devapps.diality.us/cru/HD-DEN-15839-1#c18267 Where is the heater turned back on? Is the state saved? Reply by jtaylor on 25 July 2023, 13:00 > The trimmer heater is turned back on in the transition back > to dialysis: transitionToDialysis( ) > > This is definitely a place I would suggest refactoring, as > the trimmer heater should follow dialysate flow everywhere, > but is somewhat distributed. Reply by Darren Cox on 25 July 2023, 13:12 > RESOLVED IN CODE WALKTHROUGH --- ID: HD-DEN-15839-1 https://devapps.diality.us/cru/HD-DEN-15839-1 Title: HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus Statement of Objectives: State: Closed Summary: Author: jtaylor Moderator: jtaylor Reviewers: (4 active, 2 completed*) wbracken (*) Darren Cox (*) Sean Nash Michael Garthwaite Dara Navaei Steve Jarpe