This is a list of all comments for HD-DEN-15839-2. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/Dialysis.c Revision Comment by Dara Navaei on 27 July 2023, 08:32 https://devapps.diality.us/cru/HD-DEN-15839-2#c18283 These changes are in staging right now. Do we have to stop dialysis in this state? Reply by jtaylor on 27 July 2023, 09:22 > This is a second, separate branch addressing this ticket, and > these changes should only be in develop (for testing) at this > point. Eventually, this branch should replace the branch you > reverted out of staging. > > stopDialysis() almost exactly mirrors the four lines I > commented out, except the pump "hardstop" routines also > manage the control variables that start and stop the PI > controllers. This should still make zero difference to the > case Nick is investigating since these changes are local to a > saline bolus, which is not run at all while active ultra > filtration is being run, and only affect the initial bolus > state (0), while stopping the pumps. State 1, puts the pumps > back into closed loop control, once the valves separate the > patient, the blood and the dialysate from the dialyzer. --- ID: HD-DEN-15839-2 https://devapps.diality.us/cru/HD-DEN-15839-2 Title: HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus Statement of Objectives: State: Closed Summary: Author: jtaylor Moderator: jtaylor Reviewers: (6 active, 0 completed*) Sean Nash Michael Garthwaite wbracken Dara Navaei Darren Cox Steve Jarpe