This is a list of all comments for HD-DEN-16518-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/DialOutFlow.c Revision Comment by Michael Garthwaite on 15 January 2024, 10:25 https://devapps.diality.us/cru/HD-DEN-16518-1#c19345 This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit? Reply by Sean Nash on 15 January 2024, 10:42 > Do we want to have the 11% minimum here? I thought it was > just for closed loop control. Reply by Vinayakam Mani on 15 January 2024, 11:01 > Updated code changes. Open loop still maintains the minimum > PWM duty cycle as 10%. Reply by Vinayakam Mani on 15 January 2024, 11:04 > Updated code changes. Open loop still maintains the minimum > PWM duty cycle as 10%. Reply by Michael Garthwaite on 15 January 2024, 11:10 > RESOLVED IN CODE WALKTHROUGH. --- ID: HD-DEN-16518-1 https://devapps.diality.us/cru/HD-DEN-16518-1 Title: HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment Statement of Objectives: State: Closed Summary: Author: Vinayakam Mani Moderator: Vinayakam Mani Reviewers: (3 active, 2 completed*) Sean Nash (*) Michael Garthwaite (*) jpaguio Dara Navaei Darren Cox