This is a list of all comments for HD-DEN-16563-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/PresOccl.c Revision Comment by Sean Nash on 25 January 2024, 10:15 https://devapps.diality.us/cru/HD-DEN-16563-1#c19358 MIN_TIME_BETWEEN_AIR_TRAP_FILL_LOW_PRESSURE_ALARM_EXEMPTIONS_MS (as a suggestion). Also be more descriptive in comments. Reply by Sean Nash on 02 February 2024, 10:09 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 02 February 2024, 10:07 https://devapps.diality.us/cru/HD-DEN-16563-1#c19390 This effectively changes exemption start to time of valve open instead of close. We don't want to do that. Reply by Vinayakam Mani on 02 February 2024, 16:52 > Done Reply by Sean Nash on 05 February 2024, 13:20 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 05 February 2024, 13:24 https://devapps.diality.us/cru/HD-DEN-16563-1#c19406 I think reset flag should not be set here. Should be set by new function when valve is closed I think. Reply by Vinayakam Mani on 05 February 2024, 15:18 > lowVenousPressureExemptCheck - flag initialized to TRUE in > the Init routine. Whenever valve opens (fill), it resets the > flag to "FALSE" ,the Exempt counter is not incremented. When > valve closes, flag is set back to TRUE and Exempt counter > being incremented. Reply by Sean Nash on 06 February 2024, 09:23 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 05 February 2024, 13:20 https://devapps.diality.us/cru/HD-DEN-16563-1#c19403 Shouldn't this be set to TRUE? Reply by Vinayakam Mani on 05 February 2024, 15:23 > lowVenousPressureExemptCheck - flag initialized to TRUE in > the Init routine. Whenever valve opens (fill), it resets the > flag to "FALSE" ,the Exempt counter is not incremented. When > valve closes, flag is set back to TRUE and Exempt counter > being incremented. Reply by Sean Nash on 06 February 2024, 09:23 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 05 February 2024, 13:18 https://devapps.diality.us/cru/HD-DEN-16563-1#c19402 Add space between if and (. Add to comment that low venous pressure must be exempt to increment counter. Reply by Vinayakam Mani on 05 February 2024, 15:53 > Done. Reply by Sean Nash on 06 February 2024, 09:12 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 05 February 2024, 13:22 https://devapps.diality.us/cru/HD-DEN-16563-1#c19405 Should also set new flag to FALSE here I think. Reply by Vinayakam Mani on 05 February 2024, 15:55 > lowVenousPressureExemptCheck - flag initialized to TRUE in > the Init routine. Whenever valve opens (fill), it resets the > flag to "FALSE" ,the Exempt counter is not incremented. When > valve closes, flag is set back to TRUE and Exempt counter > being incremented. Reply by Sean Nash on 06 February 2024, 09:23 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 05 February 2024, 13:24 https://devapps.diality.us/cru/HD-DEN-16563-1#c19407 I think this function should set the old reset flag to FALSE here. Function doesn't need a parameter because it is only called when valve closes which should always set reset flag to FALSE. Reply by Vinayakam Mani on 05 February 2024, 15:54 > Done. Removed the parameter in function. Reply by Sean Nash on 06 February 2024, 09:22 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-16563-1 https://devapps.diality.us/cru/HD-DEN-16563-1 Title: HD-DEN-16563_Venous Pressure Readings Were Below 20Mmhg For 1.5 Minutes Before Low Venous Statement of Objectives: State: Closed Summary: Author: Vinayakam Mani Moderator: Vinayakam Mani Reviewers: (3 active, 2 completed*) Sean Nash (*) Michael Garthwaite (*) jpaguio Dara Navaei Darren Cox