This is a list of all comments for HD-DEN-16778-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/Prime.c Revision Comment by Sean Nash on 24 May 2024, 13:12 https://devapps.diality.us/cru/HD-DEN-16778-1#c19781 This looks like a threshold. So name should start with MIN or MAX. Revision Comment by Sean Nash on 24 May 2024, 13:17 https://devapps.diality.us/cru/HD-DEN-16778-1#c19782 Alarm above has stop property. So triggering alarm will take us out of this state and go to Prime Pause state. So this else will never happen I think. Reply by Vinayakam Mani on 24 May 2024, 13:42 > Once it goes into prime pause state -> say, user removes the > clamping and resume the priming -> again this state executes > to check the clamping condition. Now, venous pressure is in > range and raw flow rate more than minimum, we get into else > state with last set flag ( primeDialysateOcclusionDetected) > and change the priming state back to blood priming initial > states. Reply by Sean Nash on 24 May 2024, 13:58 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-16778-1 https://devapps.diality.us/cru/HD-DEN-16778-1 Title: HD-DEN-16778_System Does Not Check That Air Statement of Objectives: State: Closed Summary: Author: Vinayakam Mani Moderator: Vinayakam Mani Reviewers: (3 active, 1 completed*) Michael Garthwaite (*) Sean Nash jpaguio Dara Navaei