This is a list of all comments for HD-DEN-7802-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/ModeTreatment.c Revision Comment by Sean Nash on 27 August 2021, 08:57 https://devapps.diality.us/cru/HD-DEN-7802-1#c10825 I think we do want to check dialysate temperature during treatment, but maybe not in every sub-mode. For example, I think in treatment end sub-mode we have the dialyzer bypassed (no longer performing dialysis) so there is no need to worry about the dialysate temp anymore. That is the only exception I can think of at the moment. Reply by qnguyen on 27 August 2021, 10:27 > Removed dialysate temp check for treatment end sub-mode. Reply by Sean Nash on 31 August 2021, 16:22 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DEN-7802-1 https://devapps.diality.us/cru/HD-DEN-7802-1 Title: HD-DEN-7802_DG-DEV Dialysate Generation Statement of Objectives: State: Closed Summary: Author: qnguyen Reviewers: (0 active, 2 completed*) Sean Nash (*) Dara Navaei (*)