This is a list of all comments for HD-DENBUG-64-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/BloodLeak.c Revision Comment by Sean Nash on 18 July 2023, 10:37 https://devapps.diality.us/cru/HD-DENBUG-64-1#c18189 Should be the recovery alarm. Reply by Dara Navaei on 18 July 2023, 11:16 > Done Reply by Sean Nash on 18 July 2023, 12:46 > RESOLVED in CODE WALKTHROUGH. Revision Comment by wbracken on 18 July 2023, 10:47 https://devapps.diality.us/cru/HD-DENBUG-64-1#c18190 Is this being permanently removed. I believe there is a requirement for generating this alarm. Reply by Dara Navaei on 18 October 2023, 20:35 > RESOLVED in CODE WALKTHROUGH ---------------------------------------- File: firmware/App/Modes/TreatmentStop.c Revision Comment by Sean Nash on 18 July 2023, 10:54 https://devapps.diality.us/cru/HD-DENBUG-64-1#c18195 Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer(). Reply by Dara Navaei on 18 July 2023, 11:00 > Done Reply by Sean Nash on 18 July 2023, 13:53 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 18 July 2023, 10:51 https://devapps.diality.us/cru/HD-DENBUG-64-1#c18192 Move up with other setup functions. Reply by Dara Navaei on 18 July 2023, 10:52 > Done Reply by Sean Nash on 18 July 2023, 12:44 > RESOLVED in CODE WALKTHROUGH. Revision Comment by Sean Nash on 18 July 2023, 10:52 https://devapps.diality.us/cru/HD-DENBUG-64-1#c18193 We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater. Reply by Dara Navaei on 18 July 2023, 11:01 > Correct. Alarm 126 is not blood and dialysate recirculation. Reply by Sean Nash on 18 July 2023, 12:45 > See the "also" in my comment above. Reply by Dara Navaei on 18 July 2023, 13:06 > Done Reply by Sean Nash on 18 July 2023, 13:55 > You are stopping trimmer heater. Since DPi is running, > you need to start the trimmer heater. Reply by Dara Navaei on 19 July 2023, 13:27 > Done Reply by Sean Nash on 20 July 2023, 15:28 > RESOLVED in CODE WALKTHROUGH. --- ID: HD-DENBUG-64-1 https://devapps.diality.us/cru/HD-DENBUG-64-1 Title: HD-DENBUG-64_Sys Dialyzer Blood Leak Alarm Triggered During Treatment Statement of Objectives: State: Closed Summary: Author: Dara Navaei Moderator: Dara Navaei Reviewers: (4 active, 2 completed*) Sean Nash (*) wbracken (*) Michael Garthwaite Darren Cox jtaylor Steve Jarpe