This is a list of all comments for LEAHI-DD-FIRMWARE-LDT-3103-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/BalancingChamber.c Revision Comment by Sean Nash on 14 January 2026, 09:50 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26279 Add spaces before/after +=. Reply by Vendor - TEL - Jashwant Gantyada on 15 January 2026, 14:18 > done. Revision Comment by Vendor - TEL - Jashwant Gantyada on 13 January 2026, 13:46 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26268 Why are we checking only fresh side pressures and not spent side as well? Reply by Vendor - TEL - Jashwant Gantyada on 21 January 2026, 13:09 > We will have to look at it as well. But for now, we are just > monitoring the fresh side pressure variation. Revision Comment by Sean Nash on 14 January 2026, 10:53 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26290 Where is the closed loop control? Reply by Vendor - TEL - Jashwant Gantyada on 15 January 2026, 14:19 > We handle it in the checkSpentFillComplete() mentioned in > line 946. Revision Comment by Sean Nash on 14 January 2026, 10:55 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26291 I thought we were going to complete cycle when done (not based on time) and feed back error in cycle time to control D48 pump speed to reduce error going forward so we end up getting the target cycle time which will give us target flow rate. Reply by Vendor - TEL - Jashwant Gantyada on 15 January 2026, 14:18 > Fixed it. Reply by Sean Nash on 21 January 2026, 12:44 > Should we remove the condition instead of commenting it > out? Revision Comment by Sean Nash on 02 February 2026, 10:55 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26914 Add a blank line before return statement. Revision Comment by Sean Nash on 02 February 2026, 10:55 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26915 Remove blank line. Revision Comment by Vendor - TEL - Jashwant Gantyada on 14 January 2026, 15:05 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26313 Why are we adjusting/reducing the speed of D48 by a factor of 50%? Reply by Vendor - TEL - Jashwant Gantyada on 21 January 2026, 13:08 > This is a testable value. ---------------------------------------- File: firmware/App/Modes/ModeGenDialysate.c Revision Comment by Sean Nash on 02 February 2026, 10:51 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26912 d48PumpSpeed is an input. Revision Comment by Sean Nash on 14 January 2026, 10:51 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26285 Is this an initial guess for pump control? Reply by Vendor - TEL - Jashwant Gantyada on 15 January 2026, 14:20 > Yes. will tune them based on the test results. ---------------------------------------- File: firmware/App/Modes/ModeGenDialysate.h Revision Comment by Sean Nash on 14 January 2026, 09:46 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26278 Shouldn't this kind of function be in the pump controller unit? Assuming such functions already exist in the controller, so why do we need these? Reply by Vendor - TEL - Jashwant Gantyada on 15 January 2026, 14:21 > changed the name to differentiate between driver level and > controller level functions. ---------------------------------------- File: firmware/App/Controllers/Valves.c Revision Comment by Sean Nash on 02 February 2026, 10:52 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1#c26913 Why is this commented out? --- ID: LEAHI-DD-FIRMWARE-LDT-3103-1 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-3103-1 Title: LEAHI-DD-FIRMWARE-LDT-3103_implement closed loop control for d48 pump speed Statement of Objectives: State: Review Author: Vinayakam Mani Moderator: Vinayakam Mani Reviewers: (10 active, 0 completed*) Vendor - TEL - Arpita Srivastava Vendor - TEL - Jashwant Gantyada Sean Nash Vendor - TEL - Varshini Nagabooshanam Raghu Kallala Michael Garthwaite Dara Navaei Vendor - TEL - Sivvanarayana Kurapati Daniel Ho Vendor - TEL - Sameer Poyil