This is a list of all comments for LEAHI-DD-FIRMWARE-LDT-4024-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/ModeHeatDisinfect.c Revision Comment by Vinayakam Mani on 19 August 2026, 11:02 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4024-1#c31484 D35, D40, D34,D91,D95 : Open M12, P39 : Closed ( keep a TODO comment to remove them later) Revision Comment by Vinayakam Mani on 19 August 2026, 11:08 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4024-1#c31485 Remove D95 from closed state and add D88 to closed state. Revision Comment by Vinayakam Mani on 19 August 2026, 11:11 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4024-1#c31487 Add D92 pump On at 50ml/min. Revision Comment by Vinayakam Mani on 19 August 2026, 10:55 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4024-1#c31482 D80 : open, D65: Close? Revision Comment by Vinayakam Mani on 19 August 2026, 10:59 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4024-1#c31483 close D80, D64, D14? Revision Comment by Vinayakam Mani on 19 August 2026, 12:10 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4024-1#c31490 Can we add current DD recirculation Dwell Time, WCID recirculation Dwell time variables as well and get the required dialin/automated system testing changes? --- ID: LEAHI-DD-FIRMWARE-LDT-4024-1 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4024-1 Title: LEAHI-DD-FIRMWARE-LDT-4024_Disinfection (DD heat, heat+descale) - FW Implementation - 3/4: Statement of Objectives: State: Review Author: Vendor - TEL - Jashwant Gantyada Moderator: Vendor - TEL - Jashwant Gantyada Reviewers: (4 active, 0 completed*) Vendor - TEL - Arpita Srivastava Sean Nash Vinayakam Mani Dara Navaei