This is a list of all comments for LEAHI-DD-FIRMWARE-LDT-4183-2. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/DryBiCart.c Revision Comment by Vinayakam Mani on 09 June 2026, 14:14 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29683 As suggested earlier in other code review, split this function into few init functions and call it here. Revision Comment by Vinayakam Mani on 09 June 2026, 14:23 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29688 function adjusts the volume of acid and bicarb based on the target conductivity of bicarb and dialysate solution? Revision Comment by Vinayakam Mani on 09 June 2026, 14:20 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29686 Add alarm message. Revision Comment by Vinayakam Mani on 09 June 2026, 14:16 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29684 Remove the commented code and add the comment for below function Revision Comment by Vinayakam Mani on 09 June 2026, 14:52 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29695 Do we need to call updateDryBicartOperation() here to determine the appropriate state. Revision Comment by Vinayakam Mani on 09 June 2026, 15:09 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29696 Can we keep getBicarbMixControlInterval( removing drybicart) from the function name? Revision Comment by Vinayakam Mani on 09 June 2026, 15:11 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29697 Same as above. Revision Comment by Vinayakam Mani on 09 June 2026, 15:13 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29698 Can we split the functions into two files: 1) DryBicart.c/h - Manages drybicart fill, supply, drain & depressurizing states and 2) MixingControl: for acid/bicarb open/close loop functionality? Revision Comment by Vinayakam Mani on 09 June 2026, 15:19 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29700 declare a local variable to get these function values and use the variables here? Revision Comment by Vinayakam Mani on 09 June 2026, 15:18 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29699 remove blank space before = Revision Comment by Vinayakam Mani on 09 June 2026, 15:33 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29703 Update Inputs and outputs. Revision Comment by Vinayakam Mani on 09 June 2026, 15:29 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29702 fix the typo. Revision Comment by Vinayakam Mani on 09 June 2026, 15:48 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29706 Does this supposed to be 14 or 17 PSI (based on manifold) for B2.1 hardware? Revision Comment by Vinayakam Mani on 09 June 2026, 15:41 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29705 Is this code updated for B2.1 fill implementation? Revision Comment by Vinayakam Mani on 09 June 2026, 15:56 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29707 Check pressure limit. Revision Comment by Vinayakam Mani on 09 June 2026, 15:59 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29708 This function is not checking the level anymore and may be move the functionality at end of fill water end state and transition to next chamber fill state directly, Revision Comment by Vinayakam Mani on 09 June 2026, 16:05 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2#c29709 fix the typo in 'persistence' across the file. --- ID: LEAHI-DD-FIRMWARE-LDT-4183-2 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4183-2 Title: LEAHI-DD-FIRMWARE-LDT-4183_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 2/1: Statement of Objectives: State: Review Author: Vendor - TEL - Sameer Poyil Moderator: Vendor - TEL - Sameer Poyil Reviewers: (4 active, 0 completed*) Vendor - TEL - Arpita Srivastava Sean Nash Vinayakam Mani Dara Navaei