This is a list of all comments for LEAHI-DD-FIRMWARE-LDT-4410-3. Review Summary: No summary ---------------------------------------- File: firmware/App/Controllers/MixingControl.c Revision Comment by Vinayakam Mani on 22 June 2026, 12:07 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4410-3#c30039 Add MULTIPLIER at the end. Revision Comment by Vinayakam Mani on 22 June 2026, 12:14 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4410-3#c30040 both acid and bicarb mix volume Revision Comment by Vinayakam Mani on 22 June 2026, 12:15 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4410-3#c30041 Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5 cycles of balancing chamber switching time. ---------------------------------------- File: firmware/App/Monitors/Conductivity.c Revision Comment by Vinayakam Mani on 22 June 2026, 11:39 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4410-3#c30037 Remove sample filter ms and task interval. rename as FILTER_SIZE_MULTIPLIER. Revision Comment by Vinayakam Mani on 22 June 2026, 11:47 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4410-3#c30038 Initialize at init function. Revision Comment by Vinayakam Mani on 22 June 2026, 11:37 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4410-3#c30036 Lets remove this. --- ID: LEAHI-DD-FIRMWARE-LDT-4410-3 https://devapps.diality.us/cru/LEAHI-DD-FIRMWARE-LDT-4410-3 Title: LEAHI-DD-FIRMWARE-LDT-4410_Dialysate Composition (Mixing) - FW Implementation - 3/1: Statement of Objectives: State: Review Author: Vendor - TEL - Jashwant Gantyada Moderator: Vendor - TEL - Jashwant Gantyada Reviewers: (4 active, 0 completed*) Vendor - TEL - Arpita Srivastava Sean Nash Vinayakam Mani Dara Navaei