This is a list of all comments for LEAHI-DIALIN-LDT-3160-1. Review Summary: No summary ---------------------------------------- File: leahi_dialin/common/test_config_defs.py Revision Comment by Raghu Kallala on 27 January 2026, 13:18 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26767 Here is the latest list from fw TEST_CONFIG_DD_FP_ENABLE_BETA_1_0_HW = 0, ///< Test configuration DD & FP enable Beta 1.0 hardware TEST_CONFIG_FIRST = TEST_CONFIG_DD_FP_ENABLE_BETA_1_0_HW, ///< Test configuration first configuration. TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS, ///< Test configuration DD disabling BC pressure alarms TEST_CONFIG_DD_ENABLE_DRY_BICARB, ///< Test configuration DD to use dry bicarb TEST_CONFIG_DD_ENABLE_4WIRE_RINSE_PUMP, ///< Test configuration DD enable 4-wire rinse pump. TEST_CONFIG_FP_SKIP_PRE_GEN_FLUSH, ///< Test configuration FP skip pre-gen flush TEST_CONFIG_DD_ENABLE_DIENER_1000_PUMP, ///< Test configuration to use diener 1000 pump for D48 TEST_CONFIG_DD_ENABLE_D79_PWM_CONTROL, ///< Test configuration to switch to PWM control for D79 TEST_CONFIG_DD_ENABLE_SPENT_CHAMBER_H_FILL, ///< Test configuration DD enable spent chamber H fill TEST_CONFIG_DD_DISABLE_CONDUCTIVITY_ALARMS, ///< Test configuration to disable DD conductivity alarms TEST_CONFIG_FP_DISABLE_CONDUCTIVITY_ALARMS, ///< Test configuration to disable FP conductivity alarms TEST_CONFIG_DD_ENABLE_DOSING_OPEN_LOOP_CONTROL, ///< Test configuration to switch to open loop control for concentrate dosing TEST_CONFIG_DD_ENABLE_UF_TEMP_COMPENSATION, ///< Test configuration for enabling UF temperature compensation NUM_OF_TEST_CONFIGS ///< Number of test configuration. Revision Comment by Vinayakam Mani on 27 January 2026, 11:32 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26752 TEST_CONFIG_DD_ENABLE_DIENER_1000_PUMP Reply by Zoltan Miskolci on 27 January 2026, 11:34 > renamed Revision Comment by Vinayakam Mani on 27 January 2026, 11:35 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26754 TEST_CONFIG_DD_ENABLE_D79_PWM_CONTROL. On/Off control is default support. Reply by Zoltan Miskolci on 27 January 2026, 11:40 > updated, thanks for the confirmation Revision Comment by Vinayakam Mani on 27 January 2026, 11:37 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26755 TEST_CONFIG_DD_ENABLE_SPENT_CHAMBER_H_FILL Reply by Zoltan Miskolci on 27 January 2026, 11:40 > updated Revision Comment by Vinayakam Mani on 27 January 2026, 11:38 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26756 ENABLE_DRY_BICART_FILL vs ENABLE_LIQUID_BICARB => we can use only one of these. We can keep test config 2 ( may rename as ENABLE_DRY_BICARB) and 8 can be deleted. Reply by Zoltan Miskolci on 27 January 2026, 13:32 > merged Revision Comment by Vinayakam Mani on 27 January 2026, 11:42 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26759 TEST_CONFIG_DD_ENABLE_DOSING_OPEN_LOOP_CONTROL Reply by Zoltan Miskolci on 27 January 2026, 13:32 > renamed Revision Comment by Vinayakam Mani on 27 January 2026, 11:44 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26760 TEST_CONFIG_DD_ENABLE_UF_TEMP_COMP. by default, temp compensation is disabled. Reply by Zoltan Miskolci on 27 January 2026, 13:32 > renamed Revision Comment by Vinayakam Mani on 27 January 2026, 11:48 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26761 Please sync all of these with Firmware order of definitions. Reply by Zoltan Miskolci on 27 January 2026, 13:32 > Updated based on FW ---------------------------------------- File: leahi_dialin/td/__init__.py Revision Comment by Raghu Kallala on 28 January 2026, 09:00 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1#c26780 why are we passing this exception? Reply by Zoltan Miskolci on 28 January 2026, 09:08 > This was made for the UI team, but it slowing down the init > process and always throwing that error, that is not an error > and not affecting the SW VV and SYS teams. So removing them > to fasten the init process and to not have confusing messages --- ID: LEAHI-DIALIN-LDT-3160-1 https://devapps.diality.us/cru/LEAHI-DIALIN-LDT-3160-1 Title: LEAHI-DIALIN-LDT-3160_Beta Dialin Test Configs Needed Statement of Objectives: State: Closed Summary: Author: Dara Navaei Moderator: Dara Navaei Reviewers: (8 active, 2 completed*) Vinayakam Mani (*) Raghu Kallala (*) Zoltan Miskolci Sean Nash Nicholas Ramirez Vendor - TEL - Varshini Nagabooshanam Tiffany Mejia Michael Garthwaite Behrouz NematiPour Stephen Quong