This is a list of all comments for LEAHI-TD-FIRMWARE-LDT-4329-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Services/TxParams.c Revision Comment by Sean Nash on 08 June 2026, 15:01 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-4329-1#c29641 Can you move this down with the other vitals functions? Revision Comment by Sean Nash on 08 June 2026, 15:03 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-4329-1#c29642 Remove blank line. Revision Comment by Sean Nash on 08 June 2026, 15:04 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-4329-1#c29643 It doesn't look like these memsets are necessary. Consider removing them. Revision Comment by Sean Nash on 08 June 2026, 15:07 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-4329-1#c29644 These should be moved down to an "else" in the if statement below (payload length check). Revision Comment by Sean Nash on 08 June 2026, 15:27 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-4329-1#c29645 Make a local BOOL variable and assign it TRUE or FALSE with a ternary operator and then return that local BOOl variable. Revision Comment by Sean Nash on 08 June 2026, 15:30 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-4329-1#c29646 60 should be coming from somewhere. Is there not a treatment parameter for vitals measurement interval? I think we do (TREATMENT_PARAM_BP_MEAS_INTERVAL). --- ID: LEAHI-TD-FIRMWARE-LDT-4329-1 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-4329-1 Title: LEAHI-TD-FIRMWARE-LDT-4329_Edit Treatment Parameters During Treatment - FW Implementation - 5/4: Statement of Objectives: State: Review Author: Vendor - TEL - Vijay Pamula Moderator: Vendor - TEL - Vijay Pamula Reviewers: (4 active, 0 completed*) Vendor - TEL - Arpita Srivastava Sean Nash Vinayakam Mani Dara Navaei