This is a list of all comments for LEAHI-TD-FIRMWARE-LDT-5624-1. Review Summary: No summary ---------------------------------------- File: firmware/App/Modes/StateTxRinseback.c Revision Comment by Sean Nash on 21 August 2026, 17:36 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-5624-1#c31650 Should we check to see if blood sitting timer has not exceeded time limit before we resume (similar to what we do from rinseback stop state before we start an additional)? Reply by Vendor - TEL - Praneeth Bunne on 21 August 2026, 17:48 > Sure ---------------------------------------- File: firmware/App/Services/StateServices/FluidBolus.c Revision Comment by Sean Nash on 21 August 2026, 17:27 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-5624-1#c31649 So we will command BP every 50ms for saline (okay, but not the pull approach we talked about) and send DD Qsub commands every 50ms for sub fluid (probably not a good idea)? Reply by Vendor - TEL - Praneeth Bunne on 21 August 2026, 17:41 > I'll implement the pull approach later, this is temporary > fix. > We need to run Sub pump at Qb rate, if Qb is updated, how can > we increase Qsub? Reply by Sean Nash on 21 August 2026, 17:58 > I don't have a good option. If we're going to change it to > pull approach later, can't we just ignore rate changes when > doing fluid bolus with sub fluid until you implement the > pull approach? Are we even able to do sub fluid bolus yet? > Do we have HDF tube sets? --- ID: LEAHI-TD-FIRMWARE-LDT-5624-1 https://devapps.diality.us/cru/LEAHI-TD-FIRMWARE-LDT-5624-1 Title: LEAHI-TD-FIRMWARE-LDT-5624_(TD, SW) UI screen goes to main loading screen when pausing fluid bolus in rinseback Statement of Objectives: State: Review Author: Vendor - TEL - Praneeth Bunne Moderator: Vendor - TEL - Praneeth Bunne Reviewers: (4 active, 0 completed*) Vendor - TEL - Arpita Srivastava Sean Nash Vinayakam Mani Dara Navaei