This is a list of all comments for UI-DEN-16597-29. Review Summary: No summary --- ID: UI-DEN-16597-29 https://devapps.diality.us/cru/UI-DEN-16597-29 Title: UI Bugfix/DEN-16597: UI-BN-S114 Statement of Objectives: * DENBUG-136: Main treatment screen fails to prompt user with vital measurements entry after set interval - the issue cannot be reproduced. - added a count down timer for the vital entry to better understand how and when the timer works on each interval. * DENBUG-179, DENBUG-79: Export dialog files count issues. DENBUG-179: Export Logs text does not update after Factory Reset DENBUG-79: UI: SD-Card file count increments when scrolling up and down the log file window OPEN * DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient - the reset is added to the treatment parameters - the foundation of the pre-population has been added but still there is parts need to be added to store the sliders' On/Off switch to be restored which is needed for the slider to set properly. * DENBUG-182: Phantom Alarm Screen Appears After Clearing Last Alarm - the order of the QML PROPERTYs cannot exactly predicted therefore setting them if not needed has to be avoided. - set the alarmID to 0 to trigger the next change since we skip the property updated if the alarmID is 0. State: Closed Summary: Author: Behrouz NematiPour Moderator: Behrouz NematiPour Reviewers: (4 active, 3 completed*) Tiffany Mejia (*) msuleiman (*) Dara Navaei (*) Sean Nash vduong jpaguio Michael Garthwaite