Index: AlarmMapping.csv =================================================================== diff -u -r10f946b2ecf6eed25d07d6a22216393dca34ffcc -rac49f0bd00d33f7326e82ae2d213e1b21e9083d4 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 10f946b2ecf6eed25d07d6a22216393dca34ffcc) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision ac49f0bd00d33f7326e82ae2d213e1b21e9083d4) @@ -90,15 +90,15 @@ 89,"DG outlet UV reactor on with no flow." 90,"DG load cells A1/B1 FPGA fault." 91,"HD No dialysate flow data receive in the last 3 seconds." - 92,"Available for use." + 92,"DG temperature sensors invalid calibration record." 93,"DG outlet primary conductivity out of range." 94,"DG pressure out of range." 95,"DG watchdog expired." 96,"DG inlet water temperature in high range." 97,"DG fill conductivity out of range." 98,"HD battery communication fault." 99,"HD syringe pump stall alarm." - 100,"Available for use." + 100,"DG conductivity sensors invalid temperature compensation calibration record." 101,"UI POST HD communication failure." 102,"DG heat disinfect temperature gradient out of range." 103,"HD invalid calibration CRC."