Index: AlarmMapping.csv =================================================================== diff -u -r1f2e7dbd08b00f8c56eedf29f083733570aa642e -r99e5b87f73c3b849ba33a212ec84cf4bf6ef38f1 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 1f2e7dbd08b00f8c56eedf29f083733570aa642e) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 99e5b87f73c3b849ba33a212ec84cf4bf6ef38f1) @@ -1,323 +1,324 @@ - 0,"Indication for when no alarms is a possible situation." - 1,"Stuck button POST failure.\nStop or Off button detected to be pressed for at least 1 second during test shortly after power up." - 2,"HD FPGA POST failure." - 3,"DG FPGA POST failure." - 4,"HD Watchdog POST failure." - 5,"DG Watchdog POST failure." - 6,"UI communication POST failure.\nUI failed to communicate within a reasonable time after power up." - 7,"HD RTC battery low." - 8,"HD accelerometer failed POST." - 9,"DG accelerometer failed POST." - 10,"HD blood leak sensor setting embedded mode failure." - 11,"HD dialysate temperature below target temperature." - 12,"HD firmware image integrity POST test failed." - 13,"DG firmware image integrity POST test failed." - 14,"DG invalid usage record CRC." - 15,"HD invalid usage record CRC." - 16,"HD alarm audio failed POST." - 17,"HD UI POST failed." - 18,"DG load cells invalid calibration record." - 19,"HD dialysate temperature above target temperature." - 20,"DG conductivity sensors invalid calibration record." - 21,"DG drain line volume invalid calibration record." - 22,"DG reservoirs invalid calibration record." - 23,"DG acid concentrate calibration record." - 24,"DG bicarb concentrate calibration record." - 25,"DG accelerometer invalid calibration record." - 26,"HD accelerometer invalid calibration record." - 27,"HD blood leak sensor zero and self test sequence failed." - 28,"DG two wire sensors FPGA fault." - 29,"HD heparin force sensor invalid calibration record." - 30,"HD Software fault.\nSoftware found itself in an unexpected state." - 31,"Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." - 32,"Blood pump failed motor off check.\nMeasured speed while commanded off." - 33,"Blood pump failed motor direction check.\nMeasured vs commanded." - 34,"Blood pump failed rotor speed check.\nMismatch with rotor and motor speeds." - 35,"Dialysis inlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." - 36,"Dialysis inlet pump failed motor off check.\nMeasured speed while commanded off." - 37,"Dialysis inlet pump failed motor direction check.\nMeasured vs commanded." - 38,"Dialysis inlet pump failed rotor speed check.\nMismatch with rotor and motor speeds." - 39,"Dialysis outlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." - 40,"Dialysis outlet pump failed motor off check.\nMeasured speed while commanded off." - 41,"Dialysis outlet pump failed motor direction check.\nMeasured vs commanded." - 42,"Dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds." - 43,"DG main primary heater FPGA fault." - 44,"User interface communication timeout." - 45,"HD too many bad communications CRC." - 46,"HD System message that required acknowledgment was not acknowledged." - 47,"Ultrafiltration rate is too high error during treatment." - 48,"Ultrafiltration volume accuracy error during treatment." - 49,"HD FPGA communication down for too long." - 50,"DG FPGA not accepting commanded valve states." - 51,"Blood pump failed motor speed check.\nMeasured vs commanded." - 52,"Dialysate inlet pump failed motor speed check.\nMeasured vs commanded." - 53,"Dialysate outlet pump failed motor speed check.\nMeasured vs commanded." - 54,"HD critical data integrity check failed." - 55,"DG critical data integrity check failed." - 56,"HD accelerometer error (no readings or FPGA reports error)." - 57,"DG accelerometer error (no readings or FPGA reports error)." - 58,"HD valve homing failed." - 59,"HD valve transition time out." - 60,"HD valve not functional." - 61,"HD valve current out of range." - 62,"HD valve position out of target." - 63,"HD blood leak FPGA fault." - 64,"DG barometric pressure sensor out of range." - 65,"HD requests DG command with invalid parameter fault." - 66,"HD blood leak sensor set point set failure." - 67,"HD blood pump occlusion self-test failure alarm." - 68,"HD active reservoir recirculation out of range." - 69,"HD blood leak sensor invalid calibration record." - 70,"HD arterial pressure self-test failure alarm." - 71,"HD venous pressure self-test failure alarm." - 72,"HD No load cell data message receive at least once every 2 seconds." - 73,"HD No dialysate temperature data message receive at least once every 2 seconds." - 74,"DG inlet UV reactor is on with no flow." - 75,"HD syringe pump self-test failure alarm." - 76,"HD monitored voltage is out of range." - 77,"DG monitored voltage is out of range." - 78,"HD syringe pump direction (from encoder) error." - 79,"HD syringe pump direction (from controller) error." - 80,"HD syringe pump fault reported by FPGA." - 81,"HD syringe pump over travel error." - 82,"HD syringe pump DAC write failure." - 83,"HD syringe pump is running while the BP is off." - 84,"DG set RTC year is invalid." - 85,"HD pump track latch opened alarm." - 86,"HD set RTC year is invalid." - 87,"DG heating invalid calibration record." - 88,"DG concentrate pumps hall sensor out of range." - 89,"DG outlet UV reactor on with no flow." - 90,"DG load cells A1/B1 FPGA fault." - 91,"HD No dialysate flow data receive in the last 3 seconds." - 92,"DG conductivity sensor bad status." - 93,"DG outlet primary conductivity out of range." - 94,"DG pressure out of range." - 95,"DG watchdog expired." - 96,"Inlet water temperature in high range." - 97,"Fill conductivity out of range." - 98,"HD battery communication fault." - 99,"HD syringe pump stall alarm." - 100,"HD no cartridge self-test timeout." - 101,"HD dry self-test timeout." - 102,"DG heat disinfect temperature gradient out of range." - 103,"HD invalid calibration CRC." - 104,"Air trap level sensors reporting illegal combination of air/fluid." - 105,"DG invalid calibration CRC." - 106,"DG dialysate flow sensor invalid calibration record." - 107,"HD reports DG restarted fault." - 108,"HD syringe pump ADC FPGA fault." - 109,"HD syringe pump volume check error." - 110,"HD syringe pump speed check error." - 111,"HD syringe pump not stopped in off state error." - 112,"DG load cells A1/B1 FPGA fault." - 113,"DG Drain pump current out of range." - 114,"HD venous air bubble detector self-test failure." - 115,"DG temperature sensor out of range." - 116,"DG temperature sensor ADC fault." - 117,"DG CAN message not acked." - 118,"DG RTC config error." - 119,"DG RTC battery low error." - 120,"HD pre-treatment mode wet self-test prime check failure." - 121,"DG main primary heater voltage out of range." - 122,"DG small primary heater voltage out of range." - 123,"DG trimmer heater voltage out of range." - 124,"HD end of treatment alarm (high priority)." - 125,"Blood sitting too long after treatment stopped by user alarm (>5 min)." - 126,"HD blood leak detected alarm." - 127,"Venous pressure too low during treatment." - 128,"HD venous air bubble detected alarm." - 129,"HD venous air bubble detected rinseback alarm." - 130,"Venous pressure too high during treatment." - 131,"Arterial pressure too low during treatment." - 132,"Arterial pressure too high during treatment." - 133,"DG fluid leak detected alarm." - 134,"HD fluid leak detected alarm." - 135,"HD experienced large acceleration." - 136,"DG experienced large acceleration." - 137,"HD tilt exceeds maximum." - 138,"DG tilt exceeds maximum." - 139,"HD AC power lost alarm." - 140,"Dialysate generator communication timeout." - 141,"Air trap fill timeout during treatment." - 142,"Blood pump occlusion detected." - 143,"DG dialysate temperature sensors out of range." - 144,"DG cleaning mode inlet water conductivity out of high range." - 145,"Concentrate conductivity after adding acid out of range alarm." - 146,"DG RTC or timer accuracy failure." - 147,"Wait for the DG to produce dialysate." - 148,"DG Cleaning mode inlet water temperature too high." - 149,"HD watchdog expired." - 150,"Inlet water conductivity in high range." - 151,"Inlet water conductivity in low range." - 152,"Inlet water pressure in low range." - 153,"HD prime completed high priority alarm." - 154,"DG cleaning mode inlet water conductivity out of low range." - 155,"HD No reservoirs data message receive at least once every 2 seconds." - 156,"HD No DG operation mode message receive at least once every 2 seconds." - 157,"DG chemical disinfect prime acid line timeout." - 158,"Inlet water temperature is in the low range." - 159,"DG chemical disinfect flush remove acid and close the concentrate cap." - 160,"Blood pump rotor speed too high." - 161,"DG comm too many bad CRCs." - 162,"DG FPGA clock speed check failure." - 163,"HD sees primary load cell for reservoir 1 change too much too fast." - 164,"HD sees primary load cell for reservoir 2 change too much too fast." - 165,"HD in treatment rinseback operation timeout." - 166,"HD in treatment re-circ sub-mode for too long." - 167,"HD cartridge door opened alarm." - 168,"HD active reservoir depletion timeout." - 169,"DG Dialysate flow rate out of maximum range." - 170,"HD syringe empty alarm." - 171,"HD syringe pump occlusion alarm." - 172,"HD syringe pump not enough Heparin alarm." - 173,"HD RTC configuration error." - 174,"HD RTC or timer accuracy failure." - 175,"HD pump direction status error." - 176,"HD dialysate temperature out of high safety range." - 177,"DG software fault.\nSoftware found itself in an unexpected state." - 178,"HD communication timeout." - 179,"DG FPGA communication down for too long." - 180,"DG RO flow out of maximum range." - 181,"DG load cells weight out of range for tare." - 182,"DG load cells invalid calibration." - 183,"DG invalid load cell value." - 184,"DG inlet UV reactor not healthy." - 185,"DG fan RPM out of range." - 186,"DG concentrate pump fault." - 187,"DG concentrate pump CP1 speed control error." - 188,"DG concentrate pump CP2 speed control error." - 189,"DG drain pump RPM out of range." - 190,"DG drain pump off fault." - 191,"DG flow rate out of upper range." - 192,"DG flow rate out of lower range." - 193,"HD dialysate temperature out of low safety range." - 194,"DG RO pump duty cycle out of range." - 195,"DG RO pump pressure out of range." - 196,"DG CPi/CPo sensors FPGA fault." - 197,"DG CD1/CD2 sensors FPGA fault." - 198,"RO flow too low while primary heater is on." - 199,"Dialysate flow too low while trimmer heater is on." - 200,"DG thermistors/sensors temperature out of range." - 201,"HD pre-treatment mode wet self-test lc vs FMD failure." - 202,"HD pre-treatment mode dry pressure normal self-test failure." - 203,"Pressure sensor fault." - 204,"RO rejection ratio out of range." - 205,"Conductivity sensor fault." - 206,"DG dialysate fill runs out of time." - 207,"DG flow meter check failure alarm." - 208,"HD venous pressure occlusion alarm." - 209,"DG drain circulation line timeout." - 210,"HD battery pack detected an error." - 211,"Blood sitting too long warning (>4 min)." - 212,"HD end of treatment alarm (user not acting to end treatment)." - 213,"HD prime completed medium priority alarm." - 214,"HD Saline bolus volume check failure." - 215,"HD rinseback volume check failure." - 216,"HD end treatment sub-mode timeout alarm." - 217,"HD blood prime volume check failure." - 218,"HD syringe detected alarm." - 219,"HD syringe pump syringe removed alarm." - 220,"DG THd sensors FPGA fault." - 221,"HD Saline bag is empty." - 222,"HD occlusion sensor FPGA fault." - 223,"HD arterial sensor FPGA fault." - 224,"Treatment stopped by user action." - 225,"HD end of treatment warning." - 226,"HD prime completed low priority alarm." - 227,"HD prime out of time alarm." - 228,"HD time out on prime saline purge air state." - 229,"HD prime dialysate dialyzer time out alarm." - 230,"HD prime dialysate bypass time out alarm." - 231,"HD pre-treatment mode dry pressure self-test failure." - 232,"HD pre-treatment mode wet self-test LC vs LC failure." - 233,"DG barometric sensor coefficients bad CRC." - 234,"DG inlet water pressure in high range." - 235,"HD in treatment stopped sub-mode after rinseback completed (no escalation)." - 236,"HD needs new cartridge to be installed." - 237,"HD prime saline dialyzer time out alarm." - 238,"HD no cartridge loaded or installed improperly alarm." - 239,"HD fail to remove cartridge alarm." - 240,"Bicarb conductivity out of range during bicarb pump check alarm." - 241,"DG reservoir drain time out." - 242,"DG reservoir fill time out." - 243,"DG reservoir leak time out." - 244,"DG cleaning mode temperature sensors difference out of range." - 245,"DG heat disinfect target temperature time out (could not reach to temperature)." - 246,"DG cleaning mode conductivity sensors out of range." - 247,"DG heat disinfect active cool time tour." - 248,"DG chemical disinfect could not reach to target temperature out of range." - 249,"DG chemical disinfect target conductivity out of range." - 250,"DG chemical disinfect insert acid and remove the concentrate cap." - 251,"HD invalid system record." - 252,"HD invalid service record." - 253,"DG invalid system record." - 254,"DG invalid service record." - 255,"HD and UI software builds are not compatible." - 256,"Alarm ID available for use." - 257,"DG FPGA power out timeout." - 258,"HD FPGA power out timeout." - 259,"HD temperatures out of range." - 260,"UI POST Application Integrity (Sha256Sum) failure." - 261,"UI POST CANBus failure." - 262,"UI POST Display failure." - 263,"UI POST Touch failure." - 264,"UI POST SD-Card failure." - 265,"UI POST RTC failure." - 266,"UI POST WiFi failure." - 267,"UI POST Bluetooth failure." - 268,"UI POST Ethernet failure." - 269,"UI POST Sound failure." - 270,"HD POST Safety Shutdown failure." - 271,"DG POST Safety Shutdown failure." - 272,"HD Fan RPM out of range." - 273,"HD measured blood pump flow rate is out of range." - 274,"HD measured dialysate inlet pump flow rate is out of range." - 275,"HD arterial pressure sensor is reading out of range." - 276,"HD venous pressure sensor is reading out of range." - 277,"HD BP occlusion sensor is reading out of range." - 278,"HD active reservoir weight out of range." - 279,"DG dialysate drain time out." - 280,"DG drain pump direction invalid." - 281,"Acid concentration bottle low volume alarm." - 282,"Bicarbonate concentration bottle low volume alarm." - 283,"DG load cells weight out of range." - 284,"DG load cells primary/back up drift out of range." - 285,"HD treatment recirculate timeout warning." - 286,"HD treatment rinseback complete timeout warning." - 287,"HD processor clock speed checks against FPGA clock failure." - 288,"HD load cells primary/back up drift out of range." - 289,"DG dialysate or concentrate caps not closed." - 290,"HD venous pressure sensor FPGA fault." - 291,"HD processor is in Disinfect Flush mode." - 292,"HD processor is in Disinfect Heat mode." - 293,"HD processor is in Disinfect Chemical mode." - 294,"HD processor Disinfect Chemical mode instructions to user." - 295,"UI POST CloudSync failure." - 296,"DG TDi sensors FPGA fault." - 297,"DG TRo sensors FPGA fault." - 298,"DG baro sensor FPGA fault." - 299,"DG invalid serial number." - 300,"HD invalid serial number." - 301,"DG primary heater power voltage out of range." - 302,"DG processor RAM error." - 303,"DG drain pump direction FPGA fault." - 304,"HD invalid usage record." - 305,"HD processor RAM error." - 306,"HD air trap over-fill alarm." - 307,"HD AC power lost second alarm." - 308,"DG CPi conductivity sensor invalid character received." - 309,"DG CPo conductivity sensor invalid character received." - 310,"DG CD1 conductivity sensor invalid character received." - 311,"DG CD2 conductivity sensor invalid character received." - 312,"HD Dialysate In flow rate to too low." - 313,"DG cleaning mode inlet water temperature too low." - 314,"DG cleaning mode inlet water pressure too high." - 315,"DG cleaning mode inlet water pressure too low." - 316,"DG acid concentrate pump park fault." - 317,"DG bicarb concentrate pump part fault." - 318,"HD Air Pump timeout alarm." - 319,"DG chemical disinfect flush flush sample." - 320,"DG chemical disinfect flush sample timeout." - 321,"DG outlet UV reactor not healthy." - 322,"Total number of alarms." + 0,"Indication for when no alarms is a possible situation." + 1,"HD stuck button POST failure.\nStop or Off button detected to be pressed for at least 1 second during test shortly after power up." + 2,"HD FPGA POST failure." + 3,"DG FPGA POST failure." + 4,"HD Watchdog POST failure." + 5,"DG Watchdog POST failure." + 6,"HD UI communication POST failure.\nUI failed to communicate within a reasonable time after power up." + 7,"HD RTC battery low." + 8,"HD accelerometer failed POST." + 9,"DG accelerometer failed POST." + 10,"HD blood leak sensor setting embedded mode failure." + 11,"HD dialysate temperature below target temperature or below safety temp." + 12,"HD firmware image integrity POST test failed." + 13,"DG firmware image integrity POST test failed." + 14,"DG invalid usage record CRC." + 15,"HD dialysate temperature above high safety." + 16,"HD alarm audio failed POST." + 17,"HD UI POST failed." + 18,"HD didn't get ACK on message to DG that required acknowledgment." + 19,"HD dialysate temperature above target temperature." + 20,"DG conductivity sensors invalid calibration record." + 21,"DG drain line volume invalid calibration record." + 22,"DG reservoirs invalid calibration record." + 23,"DG acid concentrate calibration record." + 24,"DG bicarb concentrate calibration record." + 25,"DG accelerometer invalid calibration record." + 26,"HD accelerometer invalid calibration record." + 27,"HD blood leak sensor zero and self test sequence failed." + 28,"DG two wire sensors FPGA fault." + 29,"HD heparin force sensor invalid calibration record." + 30,"HD Software fault.\nSoftware found itself in an unexpected state." + 31,"HD blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." + 32,"HD blood pump failed motor off check.\nMeasured speed while commanded off." + 33,"HD blood pump failed motor direction check.\nMeasured vs commanded." + 34,"HD blood pump failed rotor speed check.\nMismatch with rotor and motor speeds." + 35,"HD dialysis inlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." + 36,"HD dialysis inlet pump failed motor off check.\nMeasured speed while commanded off." + 37,"HD dialysis inlet pump failed motor direction check.\nMeasured vs commanded." + 38,"HD dialysis inlet pump failed rotor speed check.\nMismatch with rotor and motor speeds." + 39,"HD dialysis outlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." + 40,"HD dialysis outlet pump failed motor off check.\nMeasured speed while commanded off." + 41,"HD dialysis outlet pump failed motor direction check.\nMeasured vs commanded." + 42,"HD dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds." + 43,"DG main primary heater FPGA fault." + 44,"HD UI communication timeout." + 45,"HD too many bad communications CRC." + 46,"HD didn't get ACK on message to UI that required acknowledgment." + 47,"Available for use." + 48,"HD ultrafiltration volume accuracy error during treatment." + 49,"HD FPGA communication down for too long." + 50,"DG FPGA not accepting commanded valve states." + 51,"HD blood pump failed motor speed check.\nMeasured vs commanded." + 52,"HD dialysate inlet pump failed motor speed check.\nMeasured vs commanded." + 53,"HD dialysate outlet pump failed motor speed check.\nMeasured vs commanded." + 54,"HD critical data integrity check failed." + 55,"DG critical data integrity check failed." + 56,"HD accelerometer error (no readings or FPGA reports error)." + 57,"DG accelerometer error (no readings or FPGA reports error)." + 58,"HD valve homing failed." + 59,"HD valve transition time out." + 60,"HD valve not functional." + 61,"HD valve current out of range." + 62,"HD valve position out of target." + 63,"HD blood leak FPGA fault." + 64,"DG barometric pressure sensor out of range." + 65,"HD DG requests DG command with invalid parameter fault." + 66,"HD blood leak sensor set point set failure." + 67,"HD blood pump occlusion self-test failure alarm." + 68,"HD active reservoir recirculation out of range." + 69,"HD blood leak sensor invalid calibration record." + 70,"HD arterial pressure self-test failure alarm." + 71,"HD venous pressure self-test failure alarm." + 72,"HD No load cell data message receive at least once every 2 seconds." + 73,"HD No dialysate temperature data message receive at least once every 2 seconds." + 74,"DG inlet UV reactor is on with no flow." + 75,"HD syringe pump self-test failure alarm." + 76,"HD monitored voltage is out of range." + 77,"DG monitored voltage is out of range." + 78,"HD syringe pump direction (from encoder) error." + 79,"HD syringe pump direction (from controller) error." + 80,"HD syringe pump fault reported by FPGA." + 81,"HD syringe pump over travel error." + 82,"HD syringe pump DAC write failure." + 83,"HD syringe pump is running while the BP is off." + 84,"DG set RTC year is invalid." + 85,"HD pump track latch opened alarm." + 86,"HD set RTC year is invalid." + 87,"DG heating invalid calibration record." + 88,"DG concentrate pumps hall sensor out of range." + 89,"DG outlet UV reactor on with no flow." + 90,"DG load cells A1/B1 FPGA fault." + 91,"HD No dialysate flow data receive in the last 3 seconds." + 92,"Available for use." + 93,"DG outlet primary conductivity out of range." + 94,"DG pressure out of range." + 95,"DG watchdog expired." + 96,"DG inlet water temperature in high range." + 97,"DG fill conductivity out of range." + 98,"HD battery communication fault." + 99,"HD syringe pump stall alarm." + 100,"HD no cartridge self-test timeout." + 101,"UI POST HD communication failure." + 102,"DG heat disinfect temperature gradient out of range." + 103,"HD invalid calibration CRC." + 104,"HD air trap level sensors reporting illegal combination of air/fluid." + 105,"DG invalid calibration CRC." + 106,"DG dialysate flow sensor invalid calibration record." + 107,"HD reports DG restarted fault." + 108,"HD syringe pump ADC FPGA fault." + 109,"HD syringe pump volume check error." + 110,"HD syringe pump speed check error." + 111,"HD syringe pump not stopped in off state error." + 112,"DG load cells A1/B1 FPGA fault." + 113,"DG Drain pump current out of range." + 114,"HD venous air bubble detector self-test failure." + 115,"DG temperature sensor out of range." + 116,"HD UI SD card failure." + 117,"DG CAN message not acked." + 118,"DG RTC config error." + 119,"DG RTC battery low error." + 120,"HD pre-treatment mode wet self-test prime check failure." + 121,"DG main primary heater voltage out of range." + 122,"DG small primary heater voltage out of range." + 123,"DG trimmer heater voltage out of range." + 124,"HD end of treatment alarm (high priority)." + 125,"HD blood sitting too long after treatment stopped by user alarm (>5 min)." + 126,"HD blood leak detected alarm." + 127,"HD venous pressure too low during treatment." + 128,"HD venous air bubble detected alarm." + 129,"Available for use." + 130,"HD venous pressure too high during treatment." + 131,"HD arterial pressure too low during treatment." + 132,"HD arterial pressure too high during treatment." + 133,"DG fluid leak detected alarm." + 134,"HD fluid leak detected alarm." + 135,"HD experienced large acceleration." + 136,"DG experienced large acceleration." + 137,"HD tilt exceeds maximum." + 138,"DG tilt exceeds maximum." + 139,"HD AC power lost alarm." + 140,"HD DG communication timeout." + 141,"HD air trap fill timeout during treatment." + 142,"HD blood pump occlusion detected." + 143,"DG dialysate temperature sensors out of range." + 144,"DG cleaning mode inlet water conductivity out of high range." + 145,"DG concentrate conductivity after adding acid out of range alarm." + 146,"DG RTC or timer accuracy failure." + 147,"DG wait for the DG to produce dialysate." + 148,"DG Cleaning mode inlet water temperature too high." + 149,"HD watchdog expired." + 150,"DG inlet water conductivity in high range." + 151,"DG inlet water conductivity in low range." + 152,"DG inlet water pressure in low range." + 153,"HD prime completed high priority alarm." + 154,"DG cleaning mode inlet water conductivity out of low range." + 155,"HD No reservoirs data message receive at least once every 2 seconds." + 156,"HD No DG operation mode message receive at least once every 2 seconds." + 157,"DG chemical disinfect prime acid line timeout." + 158,"DG inlet water temperature is in the low range." + 159,"DG chemical disinfect flush remove acid and close the concentrate cap." + 160,"HD blood pump rotor speed too high." + 161,"DG comm too many bad CRCs." + 162,"DG FPGA clock speed check failure." + 163,"HD sees primary load cell for reservoir 1 change too much too fast." + 164,"HD sees primary load cell for reservoir 2 change too much too fast." + 165,"HD in treatment rinseback operation timeout." + 166,"Available for use." + 167,"HD cartridge door opened alarm." + 168,"Available for use." + 169,"DG Dialysate flow rate out of maximum range." + 170,"HD syringe empty alarm." + 171,"HD syringe pump occlusion alarm." + 172,"HD syringe pump not enough Heparin alarm." + 173,"HD RTC configuration error." + 174,"HD RTC or timer accuracy failure." + 175,"HD pump direction status error." + 176,"HD dialysate temperature above high safety range." + 177,"DG software fault.\nSoftware found itself in an unexpected state." + 178,"HD communication timeout." + 179,"DG FPGA communication down for too long." + 180,"DG RO flow out of maximum range." + 181,"DG load cells weight out of range for tare." + 182,"DG load cells invalid calibration." + 183,"DG invalid load cell value." + 184,"DG inlet UV reactor not healthy." + 185,"DG fan RPM out of range." + 186,"DG concentrate pump fault." + 187,"DG concentrate pump CP1 speed control error." + 188,"DG concentrate pump CP2 speed control error." + 189,"DG drain pump RPM out of range." + 190,"DG drain pump off fault." + 191,"DG flow rate out of upper range." + 192,"DG flow rate out of lower range." + 193,"HD dialysate temperature out of low safety range." + 194,"DG RO pump duty cycle out of range." + 195,"DG RO pump pressure out of range." + 196,"DG CPi/CPo sensors FPGA fault." + 197,"DG CD1/CD2 sensors FPGA fault." + 198,"DG RO flow too low while primary heater is on." + 199,"DG Dialysate flow too low while trimmer heater is on." + 200,"DG thermistors/sensors temperature out of range." + 201,"HD pre-treatment mode wet self-test lc vs FMD failure." + 202,"HD pre-treatment mode dry pressure normal self-test failure." + 203,"Available for use." + 204,"DG RO rejection ratio out of range." + 205,"DG conductivity sensor fault." + 206,"DG dialysate fill runs out of time." + 207,"DG flow meter check failure alarm." + 208,"HD venous pressure occlusion alarm." + 209,"DG drain circulation line timeout." + 210,"HD battery pack detected an error." + 211,"HD blood sitting too long warning (>4 min)." + 212,"HD end of treatment alarm (user not acting to end treatment)." + 213,"HD prime completed medium priority alarm." + 214,"Available for use." + 215,"Available for use." + 216,"HD end treatment sub-mode timeout alarm." + 217,"Available for use." + 218,"HD syringe detected alarm." + 219,"HD syringe pump syringe removed alarm." + 220,"DG THd sensors FPGA fault." + 221,"HD Saline bag is empty." + 222,"HD occlusion sensor FPGA fault." + 223,"HD arterial sensor FPGA fault." + 224,"HD treatment stopped by user action." + 225,"HD end of treatment warning." + 226,"HD prime completed low priority alarm." + 227,"Available for use." + 228,"HD time out on prime saline purge air state." + 229,"HD prime dialysate dialyzer time out alarm." + 230,"HD prime dialysate bypass time out alarm." + 231,"HD pre-treatment mode dry pressure self-test failure." + 232,"HD pre-treatment mode wet self-test LC vs LC failure." + 233,"DG barometric sensor coefficients bad CRC." + 234,"DG inlet water pressure in high range." + 235,"HD in treatment stopped sub-mode after rinseback completed (no escalation)." + 236,"HD needs new cartridge to be installed." + 237,"HD prime saline dialyzer time out alarm." + 238,"HD no cartridge loaded or installed improperly alarm." + 239,"HD fail to remove cartridge alarm." + 240,"DG bicarb conductivity out of range during bicarb pump check alarm." + 241,"DG reservoir drain time out." + 242,"DG reservoir fill time out." + 243,"DG reservoir leak time out." + 244,"DG cleaning mode temperature sensors difference out of range." + 245,"DG heat disinfect target temperature time out (could not reach to temperature)." + 246,"DG cleaning mode conductivity sensors out of range." + 247,"Alarm Id available." + 248,"DG chemical disinfect could not reach to target temperature out of range." + 249,"DG chemical disinfect target conductivity out of range." + 250,"DG chemical disinfect insert acid and remove the concentrate cap." + 251,"HD invalid system record." + 252,"HD invalid service record." + 253,"DG invalid system record." + 254,"DG invalid service record." + 255,"HD and UI software builds are not compatible." + 256,"HD processor is in Disinfect Chemical Flush mode." + 257,"Alarm ID available for use." + 258,"Alarm Id available." + 259,"HD temperatures out of range." + 260,"HD UI POST Application Integrity (Sha256Sum) failure." + 261,"HD UI POST CANBus failure." + 262,"Alarm Id available." + 263,"HD UI POST Touch failure." + 264,"HD UI POST SD-Card failure." + 265,"HD UI POST RTC failure." + 266,"HD UI POST WiFi failure." + 267,"HD UI POST Bluetooth failure." + 268,"HD UI POST Ethernet failure." + 269,"HD UI POST Sound failure." + 270,"HD POST Safety Shutdown failure." + 271,"DG POST Safety Shutdown failure." + 272,"HD Fan RPM out of range." + 273,"Available for use." + 274,"Available for use." + 275,"HD arterial pressure sensor is reading out of range." + 276,"HD venous pressure sensor is reading out of range." + 277,"HD BP occlusion sensor is reading out of range." + 278,"HD active reservoir weight out of range." + 279,"DG dialysate drain time out." + 280,"DG drain pump direction invalid." + 281,"Acid concentration bottle low volume alarm." + 282,"Bicarbonate concentration bottle low volume alarm." + 283,"DG load cells weight out of range." + 284,"DG load cells primary/back up drift out of range." + 285,"Available for use." + 286,"DG concentrate cap not in proper position." + 287,"HD processor clock speed checks against FPGA clock failure." + 288,"HD load cells primary/back up drift out of range." + 289,"DG dialysate cap not closed." + 290,"HD venous pressure sensor FPGA fault." + 291,"HD processor is in Disinfect Flush mode." + 292,"HD processor is in Disinfect Heat mode." + 293,"HD processor is in Disinfect Chemical mode." + 294,"Available for use." + 295,"HD UI POST CloudSync failure." + 296,"DG TDi sensors FPGA fault." + 297,"DG TRo sensors FPGA fault." + 298,"DG baro sensor FPGA fault." + 299,"DG invalid serial number." + 300,"HD invalid serial number." + 301,"Alarm ID available." + 302,"DG processor RAM error." + 303,"DG drain pump direction FPGA fault." + 304,"HD invalid usage record." + 305,"HD processor RAM error." + 306,"DG CD1 CD2 check fault." + 307,"HD AC power lost second alarm." + 308,"DG CPi conductivity sensor invalid character received." + 309,"DG CPo conductivity sensor invalid character received." + 310,"DG CD1 conductivity sensor invalid character received." + 311,"DG CD2 conductivity sensor invalid character received." + 312,"HD Dialysate In flow rate to too low." + 313,"DG cleaning mode inlet water temperature too low." + 314,"DG cleaning mode inlet water pressure too high." + 315,"DG cleaning mode inlet water pressure too low." + 316,"DG acid concentrate pump park fault." + 317,"DG bicarb concentrate pump part fault." + 318,"HD Air Pump timeout alarm." + 319,"DG chemical disinfect flush flush sample." + 320,"DG chemical disinfect flush sample timeout." + 321,"DG outlet UV reactor not healthy." + 322,"HD UI POST Year check failure." + 323,"Total number of alarms."