Index: AlarmMapping.csv =================================================================== diff -u -r3561f79af1a92356eea01d5d0c3297c69d2ecdf2 -rdf0329c47627ceb1fe460cda2ae3ff8c9c851106 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 3561f79af1a92356eea01d5d0c3297c69d2ecdf2) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision df0329c47627ceb1fe460cda2ae3ff8c9c851106) @@ -94,8 +94,8 @@ 93,"This alarm ID is available for use." 94,"This alarm ID is available for use." 95,"This alarm ID is available for use." - 96,"This alarm ID is available for use." - 97,"This alarm ID is available for use." + 96,"Inlet water temperature in high range." + 97,"Fill conductivity out of range." 98,"HD battery communication fault." 99,"HD syringe pump stall alarm." 100,"HD no cartridge self-test timeout." @@ -104,7 +104,7 @@ 103,"HD non-volatile calibration group invalid record CRC." 104,"Air trap level sensors reporting illegal combination of air/fluid." 105,"DG non-volatile system group invalid record CRC." - 106,"This alarm ID is available for use." + 106,"DG inlet water conductivity is greater than threshold." 107,"HD reports DG restarted fault." 108,"HD syringe pump ADC error." 109,"HD syringe pump volume check error." @@ -115,10 +115,10 @@ 114,"HD venous air bubble detector self-test failure." 115,"DG temperature sensor out of range." 116,"DG temperature sensor ADC out of range." - 117,"This alarm ID is available for use." - 118,"This alarm ID is available for use." - 119,"This alarm ID is available for use." - 120,"This alarm ID is available for use." + 117,"DG primary heater internal temperature out of range." + 118,"DG primary heater cold junction temperature out of range." + 119,"DG trimmer heater internal temperature out of range." + 120,"DG trimmer heater cold junction temperature out of range." 121,"DG main primary heater voltage out of range." 122,"DG small primary heater voltage out of range." 123,"DG trimmer heater voltage out of range." @@ -145,18 +145,18 @@ 144,"DG software configuration record invalid CRC." 145,"Concentrate conductivity after adding acid out of range alarm." 146,"Dialysate conductivity out of range." - 147,"DG dialysate generation conductivity fault alarm." + 147,"Wait for the DG to produce dialysate." 148,"Inlet water temperature too high." - 149,"Inlet water temperature too low." - 150,"Inlet water conductivity too high." - 151,"Inlet water conductivity too low." - 152,"Inlet water pressure too low." + 149,"Alarm ID available for use." + 150,"Inlet water conductivity in high range." + 151,"Inlet water conductivity in low range." + 152,"Inlet water pressure in low range." 153,"HD prime completed high priority alarm." 154,"EEPROM operations (read, write, erase) failure." 155,"DG software configuration record invalid CRC." 156,"HW usage data (treatment time in HD and total consumed water in DG) failure." 157,"DG chemical disinfect prime acid line timeout." - 158,"This alarm ID is available for use." + 158,"Inlet water temperature is in the low range." 159,"Dialysate inlet pump failed flow vs motor speed check." 160,"Blood pump rotor speed too high." 161,"This alarm ID is available for use."