Index: AlarmMapping.csv =================================================================== diff -u -r465a935949a85d3d1bebd11979737ff38ef96122 -r35db01154f90ebfdb6174033a66f5c20af82e14d --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 465a935949a85d3d1bebd11979737ff38ef96122) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 35db01154f90ebfdb6174033a66f5c20af82e14d) @@ -12,8 +12,8 @@ 11,"HD dialysate temperature below target temperature." 12,"HD firmware image integrity POST test failed." 13,"DG firmware image integrity POST test failed." - 14,"Not used." - 15,"Not used." + 14,"DG invalid usage record CRC." + 15,"HD invalid usage record CRC." 16,"HD alarm audio failed POST." 17,"HD UI POST failed." 18,"DG load cells invalid calibration record." @@ -101,9 +101,9 @@ 100,"HD no cartridge self-test timeout." 101,"HD dry self-test timeout." 102,"DG heat disinfect temperature gradient out of range." - 103,"HD non-volatile calibration group invalid record CRC." + 103,"HD invalid calibration CRC." 104,"Air trap level sensors reporting illegal combination of air/fluid." - 105,"DG non-volatile system group invalid record CRC." + 105,"DG invalid calibration CRC." 106,"DG dialysate flow sensor invalid calibration record." 107,"HD reports DG restarted fault." 108,"HD syringe pump ADC error."