Index: AlarmMapping.csv =================================================================== diff -u -r7ebc073150532073ae7f1a0e768272bcff5d6fb3 -r465a935949a85d3d1bebd11979737ff38ef96122 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision 7ebc073150532073ae7f1a0e768272bcff5d6fb3) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 465a935949a85d3d1bebd11979737ff38ef96122) @@ -8,25 +8,25 @@ 7,"HD RTC battery low." 8,"HD accelerometer failed POST." 9,"DG accelerometer failed POST." - 10,"Alarm ID available for use." - 11,"Alarm ID available for use." + 10,"Not used." + 11,"HD dialysate temperature below target temperature." 12,"HD firmware image integrity POST test failed." 13,"DG firmware image integrity POST test failed." - 14,"HD blood flow invalid calibration." - 15,"HD dialysate flow invalid calibration." + 14,"Not used." + 15,"Not used." 16,"HD alarm audio failed POST." 17,"HD UI POST failed." 18,"DG load cells invalid calibration record." - 19,"DG flow sensors invalid calibration record." + 19,"HD dialysate temperature above target temperature." 20,"DG conductivity sensors invalid calibration record." 21,"DG drain line volume invalid calibration record." 22,"DG reservoirs invalid calibration record." 23,"DG acid concentrate calibration record." 24,"DG bicarb concentrate calibration record." 25,"DG accelerometer invalid calibration record." 26,"HD accelerometer invalid calibration record." - 27,"HD blood flow invalid calibration record." - 28,"HD dialyzer flow invalid calibration record." + 27,"Not used." + 28,"Not used." 29,"HD heparin force sensor invalid calibration record." 30,"HD Software fault.\nSoftware found itself in an unexpected state." 31,"Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." @@ -70,8 +70,8 @@ 69,"HD blood leak sensor invalid calibration record." 70,"HD arterial pressure self-test failure alarm." 71,"HD venous pressure self-test failure alarm." - 72,"HD blood flow meter status check self-test failure alarm." - 73,"HD dialysate flow meter status check self-test failure alarm." + 72,"Not used." + 73,"Not used." 74,"HD blood leak detector self-test transmit timeout." 75,"HD syringe pump self-test failure alarm." 76,"HD monitored voltage is out of range." @@ -153,8 +153,8 @@ 152,"Inlet water pressure in low range." 153,"HD prime completed high priority alarm." 154,"EEPROM operations (read, write, erase) failure." - 155,"Alarm ID available for use." - 156,"Alarm ID available for use." + 155,"Not used." + 156,"Not used." 157,"DG chemical disinfect prime acid line timeout." 158,"Inlet water temperature is in the low range." 159,"Dialysate inlet pump failed flow vs motor speed check." @@ -174,7 +174,7 @@ 173,"HD RTC configuration error." 174,"HD RTC or timer accuracy failure." 175,"HD pump direction status error." - 176,"HD active reservoir is full during treatment and DG not ready to switch reservoirs." + 176,"HD dialysate temperature out of high safety range." 177,"DG software fault.\nSoftware found itself in an unexpected state." 178,"HD communication timeout." 179,"DG FPGA communication down for too long." @@ -191,7 +191,7 @@ 190,"DG drain pump off fault." 191,"DG flow rate out of upper range." 192,"DG flow rate out of lower range." - 193,"Alarm available for use." + 193,"HD dialysate temperature out of low safety range." 194,"DG RO pump duty cycle out of range." 195,"DG RO pump pressure out of range." 196,"DG temperature sensor error bit fault."