Index: AlarmMapping.csv =================================================================== diff -u -ra5fe0c9504558ddafc2af2ccd852b8f311a00616 -r6f2622be80fcaee8d71baf78404627381157970c --- AlarmMapping.csv (.../AlarmMapping.csv) (revision a5fe0c9504558ddafc2af2ccd852b8f311a00616) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 6f2622be80fcaee8d71baf78404627381157970c) @@ -196,8 +196,8 @@ 195,"DG RO pump pressure out of range." 196,"DG temperature sensor error bit fault." 197,"DG temperature sensors FPGA fault." - 198,"Alarm ID available for use." - 199,"Alarm ID available for use." + 198,"RO flow too low while primary heater is on." + 199,"Dialysate flow too low while trimmer heater is on." 200,"DG thermistors/sensors temperature out of range." 201,"HD pre-treatment mode wet self-test lc vs FMD failure." 202,"HD pre-treatment mode dry pressure normal self-test failure." @@ -206,7 +206,7 @@ 205,"Conductivity sensor fault." 206,"DG dialysate fill runs out of time." 207,"DG flow meter check failure alarm." - 208,"Alarm ID available for use." + 208,"HD venous pressure occlusion alarm." 209,"DG drain circulation line timeout." 210,"HD battery pack detected an error." 211,"Blood sitting too long warning (>4 min)."