Index: AlarmMapping.csv =================================================================== diff -u -raa8f2c87c14c68d1fda6da2540d47144990a596c -rc9dafe29d33091483dde64e999bf0d6fd841ba23 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision aa8f2c87c14c68d1fda6da2540d47144990a596c) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision c9dafe29d33091483dde64e999bf0d6fd841ba23) @@ -166,8 +166,8 @@ 165,"HD in treatment rinseback operation timeout." 166,"HD in treatment re-circ sub-mode for too long." 167,"HD cartridge door opened alarm." - 168,"HD dialysate temperature too high alarm." - 169,"HD dialysate temperature too low alarm." + 168,"This alarm ID is available for use." + 169,"This alarm ID is available for use." 170,"HD syringe empty alarm." 171,"HD syringe pump occlusion alarm." 172,"HD syringe pump not enough Heparin alarm." @@ -206,7 +206,7 @@ 205,"Conductivity sensor fault." 206,"DG dialysate fill runs out of time." 207,"DG flow meter check failure alarm." - 208,"DG heaters on with no flow time out." + 208,"DG primary heater on with no flow time out." 209,"DG drain circulation line timeout." 210,"HD battery pack detected an error." 211,"Blood sitting too long warning (>4 min)." @@ -286,4 +286,5 @@ 285,"HD treatment recirculate timeout warning." 286,"HD treatment rinseback complete timeout warning." 287,"HD processor clock speed checks against FPGA clock failure." - 288,"Total number of alarms." + 288,"DG trimmer heater on with flow timeout." + 289,"Total number of alarms."