Index: AlarmMapping.csv =================================================================== diff -u -rcf61c7736f11f4b75a4e2df803f29bdc7769db53 -rd826e189e0dff4b91f880c73236129fa3c25a978 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision cf61c7736f11f4b75a4e2df803f29bdc7769db53) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision d826e189e0dff4b91f880c73236129fa3c25a978) @@ -260,7 +260,7 @@ 259,"HD temperatures out of range." 260,"HD UI POST Application Integrity (Sha256Sum) failure." 261,"HD UI POST CANBus failure." - 262,"HD UI POST Display failure." + 262,"Alarm Id available." 263,"HD UI POST Touch failure." 264,"HD UI POST SD-Card failure." 265,"HD UI POST RTC failure." @@ -283,7 +283,7 @@ 282,"Bicarbonate concentration bottle low volume alarm." 283,"DG load cells weight out of range." 284,"DG load cells primary/back up drift out of range." - 285,"Available for use." + 285,"HD processor is in RO permeate sample mode." 286,"DG concentrate cap not in proper position." 287,"HD processor clock speed checks against FPGA clock failure." 288,"HD load cells primary/back up drift out of range." @@ -292,7 +292,7 @@ 291,"HD processor is in Disinfect Flush mode." 292,"HD processor is in Disinfect Heat mode." 293,"HD processor is in Disinfect Chemical mode." - 294,"Available for use." + 294,"HD processor is in Disinfect active cool mode." 295,"HD UI POST CloudSync failure." 296,"DG TDi sensors FPGA fault." 297,"DG TRo sensors FPGA fault."