Index: AlarmMapping.csv =================================================================== diff -u -rdaad774fa7aece882a72d6e579a7cb87988fd6a5 -raa8f2c87c14c68d1fda6da2540d47144990a596c --- AlarmMapping.csv (.../AlarmMapping.csv) (revision daad774fa7aece882a72d6e579a7cb87988fd6a5) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision aa8f2c87c14c68d1fda6da2540d47144990a596c) @@ -123,7 +123,7 @@ 122,"DG small primary heater voltage out of range." 123,"DG trimmer heater voltage out of range." 124,"HD end of treatment alarm (high priority - user still not acting to end treatment)." - 125,"Blood sitting too long after treatment stopped by user alarm (>10 min)." + 125,"Blood sitting too long after treatment stopped by user alarm (>5 min)." 126,"HD blood leak detected alarm." 127,"Venous pressure too low during treatment." 128,"HD venous air bubble detected alarm." @@ -283,4 +283,7 @@ 282,"Bicarbonate concentration bottle low volume alarm." 283,"DG load cells weight out of range." 284,"DG load cells primary/back up drift out of range." - 285,"Total number of alarms." + 285,"HD treatment recirculate timeout warning." + 286,"HD treatment rinseback complete timeout warning." + 287,"HD processor clock speed checks against FPGA clock failure." + 288,"Total number of alarms."