Index: AlarmMapping.csv =================================================================== diff -u -rdb92783a0a9dab0ddf2b68c51d15b1ae1cb3ddcc -r5b46bc8e4724e4ddb648f901d76bb7ec24f8b6c7 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision db92783a0a9dab0ddf2b68c51d15b1ae1cb3ddcc) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 5b46bc8e4724e4ddb648f901d76bb7ec24f8b6c7) @@ -5,28 +5,28 @@ 4,"HD Watchdog POST failure." 5,"DG Watchdog POST failure." 6,"UI communication POST failure.\nUI failed to communicate within a reasonable time after power up." - 7,"Real-time clock configuration error." + 7,"HD RTC battery low." 8,"HD accelerometer failed POST." 9,"DG accelerometer failed POST." - 10,"RTC or MCU timer inaccurate." - 11,"DG heaters failed POST." + 10,"HD blood leak sensor setting embedded mode failure." + 11,"HD dialysate temperature below target temperature." 12,"HD firmware image integrity POST test failed." 13,"DG firmware image integrity POST test failed." - 14,"HD blood flow invalid calibration." - 15,"HD dialysate flow invalid calibration." + 14,"DG invalid usage record CRC." + 15,"HD invalid usage record CRC." 16,"HD alarm audio failed POST." 17,"HD UI POST failed." 18,"DG load cells invalid calibration record." - 19,"DG flow sensors invalid calibration record." + 19,"HD dialysate temperature above target temperature." 20,"DG conductivity sensors invalid calibration record." 21,"DG drain line volume invalid calibration record." 22,"DG reservoirs invalid calibration record." 23,"DG acid concentrate calibration record." 24,"DG bicarb concentrate calibration record." 25,"DG accelerometer invalid calibration record." 26,"HD accelerometer invalid calibration record." - 27,"HD blood flow invalid calibration record." - 28,"HD dialyzer flow invalid calibration record." + 27,"HD blood leak sensor zero and self test sequence failed." + 28,"DG two wire sensors FPGA fault." 29,"HD heparin force sensor invalid calibration record." 30,"HD Software fault.\nSoftware found itself in an unexpected state." 31,"Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." @@ -41,14 +41,14 @@ 40,"Dialysis outlet pump failed motor off check.\nMeasured speed while commanded off." 41,"Dialysis outlet pump failed motor direction check.\nMeasured vs commanded." 42,"Dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds." - 43,"Watchdog expired error.\nWatchdog was not \"pet\" within the appropriate time." + 43,"DG main primary heater FPGA fault." 44,"User interface communication timeout." - 45,"Too many bad CRCs detected on received system messages." - 46,"System message that required acknowledgment was not acknowledged." + 45,"HD too many bad communications CRC." + 46,"HD System message that required acknowledgment was not acknowledged." 47,"Ultrafiltration rate is too high error during treatment." 48,"Ultrafiltration volume accuracy error during treatment." 49,"HD FPGA communication down for too long." - 50,"FPGA not accepting commanded valve states." + 50,"DG FPGA not accepting commanded valve states." 51,"Blood pump failed motor speed check.\nMeasured vs commanded." 52,"Dialysate inlet pump failed motor speed check.\nMeasured vs commanded." 53,"Dialysate outlet pump failed motor speed check.\nMeasured vs commanded." @@ -61,18 +61,18 @@ 60,"HD valve not functional." 61,"HD valve current out of range." 62,"HD valve position out of target." - 63,"FPGA reports failure reading arterial pressure sensor." - 64,"FPGA reports failure reading venous pressure sensor." + 63,"HD blood leak FPGA fault." + 64,"DG barometric pressure sensor out of range." 65,"HD requests DG command with invalid parameter fault." - 66,"This alarm ID is available for use." + 66,"HD blood leak sensor set point set failure." 67,"HD blood pump occlusion self-test failure alarm." 68,"HD active reservoir recirculation out of range." - 69,"This alarm ID is available for use." + 69,"HD blood leak sensor invalid calibration record." 70,"HD arterial pressure self-test failure alarm." 71,"HD venous pressure self-test failure alarm." - 72,"HD blood flow meter status check self-test failure alarm." - 73,"HD dialysate flow meter status check self-test failure alarm." - 74,"HD blood leak detector self-test failure alarm." + 72,"HD No load cell data message receive at least once every 2 seconds." + 73,"HD No dialysate temperature data message receive at least once every 2 seconds." + 74,"DG inlet UV reactor is on with no flow." 75,"HD syringe pump self-test failure alarm." 76,"HD monitored voltage is out of range." 77,"DG monitored voltage is out of range." @@ -82,47 +82,47 @@ 81,"HD syringe pump over travel error." 82,"HD syringe pump DAC write failure." 83,"HD syringe pump is running while the BP is off." - 84,"HD venous pressure sensor not being read." - 85,"HD venous pressure sensor temperature out of range error." - 86,"HD BP occlusion sensor not being read." - 87,"This alarm ID is available for use." - 88,"This alarm ID is available for use." - 89,"HD BP occlusion sensor error." - 90,"This alarm ID is available for use." + 84,"DG set RTC year is invalid." + 85,"HD pump track latch opened alarm." + 86,"HD set RTC year is invalid." + 87,"DG heating invalid calibration record." + 88,"DG concentrate pumps hall sensor out of range." + 89,"DG outlet UV reactor on with no flow." + 90,"DG load cells A1/B1 FPGA fault." 91,"HD No dialysate flow data receive in the last 3 seconds." - 92,"This alarm ID is available for use." - 93,"This alarm ID is available for use." - 94,"This alarm ID is available for use." - 95,"This alarm ID is available for use." - 96,"This alarm ID is available for use." - 97,"This alarm ID is available for use." + 92,"DG conductivity sensor bad status." + 93,"DG outlet primary conductivity out of range." + 94,"DG pressure out of range." + 95,"DG watchdog expired." + 96,"Inlet water temperature in high range." + 97,"Fill conductivity out of range." 98,"HD battery communication fault." 99,"HD syringe pump stall alarm." 100,"HD no cartridge self-test timeout." 101,"HD dry self-test timeout." - 102,"Real-time clock communication error." - 103,"HD non-volatile calibration group invalid record CRC." + 102,"DG heat disinfect temperature gradient out of range." + 103,"HD invalid calibration CRC." 104,"Air trap level sensors reporting illegal combination of air/fluid." - 105,"DG non-volatile system group invalid record CRC." - 106,"This alarm ID is available for use." + 105,"DG invalid calibration CRC." + 106,"DG dialysate flow sensor invalid calibration record." 107,"HD reports DG restarted fault." - 108,"HD syringe pump ADC error." + 108,"HD syringe pump ADC FPGA fault." 109,"HD syringe pump volume check error." 110,"HD syringe pump speed check error." 111,"HD syringe pump not stopped in off state error." - 112,"HD blood leak detector fault." - 113,"HD arterial air bubble detector self-test failure." + 112,"DG load cells A1/B1 FPGA fault." + 113,"DG Drain pump current out of range." 114,"HD venous air bubble detector self-test failure." 115,"DG temperature sensor out of range." - 116,"DG temperature sensor ADC out of range." - 117,"DG primary heater internal temperature out of range." - 118,"DG primary heater cold junction temperature out of range." - 119,"DG trimmer heater internal temperature out of range." - 120,"DG trimmer heater cold junction temperature out of range." + 116,"DG temperature sensor ADC fault." + 117,"DG CAN message not acked." + 118,"DG RTC config error." + 119,"DG RTC battery low error." + 120,"HD pre-treatment mode wet self-test prime check failure." 121,"DG main primary heater voltage out of range." 122,"DG small primary heater voltage out of range." 123,"DG trimmer heater voltage out of range." - 124,"HD end of treatment alarm (high priority - user still not acting to end treatment)." + 124,"HD end of treatment alarm (high priority)." 125,"Blood sitting too long after treatment stopped by user alarm (>5 min)." 126,"HD blood leak detected alarm." 127,"Venous pressure too low during treatment." @@ -141,44 +141,44 @@ 140,"Dialysate generator communication timeout." 141,"Air trap fill timeout during treatment." 142,"Blood pump occlusion detected." - 143,"This alarm ID is available for use." - 144,"This alarm ID is available for use." + 143,"DG dialysate temperature sensors out of range." + 144,"DG cleaning mode inlet water conductivity out of high range." 145,"Concentrate conductivity after adding acid out of range alarm." - 146,"Dialysate conductivity out of range." - 147,"DG dialysate generation conductivity fault alarm." - 148,"Inlet water temperature too high." - 149,"Inlet water temperature too low." - 150,"Inlet water conductivity too high." - 151,"Inlet water conductivity too low." - 152,"Inlet water pressure too low." + 146,"DG RTC or timer accuracy failure." + 147,"Wait for the DG to produce dialysate." + 148,"DG Cleaning mode inlet water temperature too high." + 149,"HD watchdog expired." + 150,"Inlet water conductivity in high range." + 151,"Inlet water conductivity in low range." + 152,"Inlet water pressure in low range." 153,"HD prime completed high priority alarm." - 154,"EEPROM operations (read, write, erase) failure." - 155,"This alarm ID is available for use." - 156,"HW usage data (treatment time in HD and total consumed water in DG) failure." - 157,"This alarm ID is available for use." - 158,"Blood pump failed flow vs motor speed check." - 159,"Dialysate inlet pump failed flow vs motor speed check." + 154,"DG cleaning mode inlet water conductivity out of low range." + 155,"HD No reservoirs data message receive at least once every 2 seconds." + 156,"HD No DG operation mode message receive at least once every 2 seconds." + 157,"DG chemical disinfect prime acid line timeout." + 158,"Inlet water temperature is in the low range." + 159,"DG chemical disinfect flush remove acid and close the concentrate cap." 160,"Blood pump rotor speed too high." - 161,"Blood flow sensor signal strength too low." - 162,"Dialysate flow sensor signal strength too low." + 161,"DG comm too many bad CRCs." + 162,"DG FPGA clock speed check failure." 163,"HD sees primary load cell for reservoir 1 change too much too fast." 164,"HD sees primary load cell for reservoir 2 change too much too fast." 165,"HD in treatment rinseback operation timeout." 166,"HD in treatment re-circ sub-mode for too long." 167,"HD cartridge door opened alarm." 168,"HD active reservoir depletion timeout." - 169,"Dialysate flow rate out of range." + 169,"DG Dialysate flow rate out of maximum range." 170,"HD syringe empty alarm." 171,"HD syringe pump occlusion alarm." 172,"HD syringe pump not enough Heparin alarm." - 173,"HD arterial air bubble detected alarm." - 174,"HD arterial air bubble detected rinseback alarm." + 173,"HD RTC configuration error." + 174,"HD RTC or timer accuracy failure." 175,"HD pump direction status error." - 176,"HD active reservoir is full during treatment and DG not ready to switch reservoirs." + 176,"HD dialysate temperature out of high safety range." 177,"DG software fault.\nSoftware found itself in an unexpected state." 178,"HD communication timeout." 179,"DG FPGA communication down for too long." - 180,"DG load cell ADC error." + 180,"DG RO flow out of maximum range." 181,"DG load cells weight out of range for tare." 182,"DG load cells invalid calibration." 183,"DG invalid load cell value." @@ -191,22 +191,22 @@ 190,"DG drain pump off fault." 191,"DG flow rate out of upper range." 192,"DG flow rate out of lower range." - 193,"RO pump flow rate out of range." + 193,"HD dialysate temperature out of low safety range." 194,"DG RO pump duty cycle out of range." 195,"DG RO pump pressure out of range." - 196,"DG temperature sensor error flag fault." - 197,"DG temperature sensors ADC fault." - 198,"DG heaters cold junction temperature out of range." - 199,"DG heaters fault." + 196,"DG CPi/CPo sensors FPGA fault." + 197,"DG CD1/CD2 sensors FPGA fault." + 198,"RO flow too low while primary heater is on." + 199,"Dialysate flow too low while trimmer heater is on." 200,"DG thermistors/sensors temperature out of range." - 201,"Inlet water quality is bad (temperature, conductivity, pressure)." - 202,"Inlet water pressure fault." + 201,"HD pre-treatment mode wet self-test lc vs FMD failure." + 202,"HD pre-treatment mode dry pressure normal self-test failure." 203,"Pressure sensor fault." 204,"RO rejection ratio out of range." 205,"Conductivity sensor fault." 206,"DG dialysate fill runs out of time." 207,"DG flow meter check failure alarm." - 208,"DG primary heater on with no flow time out." + 208,"HD venous pressure occlusion alarm." 209,"DG drain circulation line timeout." 210,"HD battery pack detected an error." 211,"Blood sitting too long warning (>4 min)." @@ -218,21 +218,21 @@ 217,"HD blood prime volume check failure." 218,"HD syringe detected alarm." 219,"HD syringe pump syringe removed alarm." - 220,"HD battery pack SOC is too low." + 220,"DG THd sensors FPGA fault." 221,"HD Saline bag is empty." - 222,"HD dialysate temperature too high alarm." - 223,"HD dialysate temperature too low alarm." - 224,"Treatment stopped by user action - pressed stop button." + 222,"HD occlusion sensor FPGA fault." + 223,"HD arterial sensor FPGA fault." + 224,"Treatment stopped by user action." 225,"HD end of treatment warning." 226,"HD prime completed low priority alarm." 227,"HD prime out of time alarm." - 228,"HD time out on prime sub-mode purge air state." + 228,"HD time out on prime saline purge air state." 229,"HD prime dialysate dialyzer time out alarm." 230,"HD prime dialysate bypass time out alarm." - 231,"HD pre-treatment mode dry self-tests failure." - 232,"HD pre-treatment mode wet self-tests failure." - 233,"RTC battery low." - 234,"RTC (i.e read time) or RAM operations failure (read or write)." + 231,"HD pre-treatment mode dry pressure self-test failure." + 232,"HD pre-treatment mode wet self-test LC vs LC failure." + 233,"DG barometric sensor coefficients bad CRC." + 234,"DG inlet water pressure in high range." 235,"HD in treatment stopped sub-mode after rinseback completed (no escalation)." 236,"HD needs new cartridge to be installed." 237,"HD prime saline dialyzer time out alarm." @@ -242,13 +242,13 @@ 241,"DG reservoir drain time out." 242,"DG reservoir fill time out." 243,"DG reservoir leak time out." - 244,"DG temperature sensors difference out of range." + 244,"DG cleaning mode temperature sensors difference out of range." 245,"DG heat disinfect target temperature time out (could not reach to temperature)." - 246,"DG heat disinfect inlet pressure and temperature sensors out of range." - 247,"DG heat disinfect inlet conductivity and temperature out of range." - 248,"DG chemical disinfect could not reach to target temperature." - 249,"DG chemical disinfect inlet pressure and temperature sensors out of range." - 250,"DG chemical disinfect inlet conductivity and temperature out of range." + 246,"DG cleaning mode conductivity sensors out of range." + 247,"DG heat disinfect active cool time tour." + 248,"DG chemical disinfect could not reach to target temperature out of range." + 249,"DG chemical disinfect target conductivity out of range." + 250,"DG chemical disinfect insert acid and remove the concentrate cap." 251,"HD invalid system record." 252,"HD invalid service record." 253,"DG invalid system record." @@ -278,13 +278,45 @@ 277,"HD BP occlusion sensor is reading out of range." 278,"HD active reservoir weight out of range." 279,"DG dialysate drain time out." - 280,"HD arterial pressure sensor read timeout error." + 280,"DG drain pump direction invalid." 281,"Acid concentration bottle low volume alarm." 282,"Bicarbonate concentration bottle low volume alarm." 283,"DG load cells weight out of range." 284,"DG load cells primary/back up drift out of range." 285,"HD treatment recirculate timeout warning." 286,"HD treatment rinseback complete timeout warning." 287,"HD processor clock speed checks against FPGA clock failure." - 288,"DG trimmer heater on with flow timeout." - 289,"Total number of alarms." + 288,"HD load cells primary/back up drift out of range." + 289,"DG dialysate or concentrate caps not closed." + 290,"HD venous pressure sensor FPGA fault." + 291,"HD processor is in Disinfect Flush mode." + 292,"HD processor is in Disinfect Heat mode." + 293,"HD processor is in Disinfect Chemical mode." + 294,"HD processor Disinfect Chemical mode instructions to user." + 295,"UI POST CloudSync failure." + 296,"DG TDi sensors FPGA fault." + 297,"DG TRo sensors FPGA fault." + 298,"DG baro sensor FPGA fault." + 299,"DG invalid serial number." + 300,"HD invalid serial number." + 301,"DG primary heater power voltage out of range." + 302,"DG processor RAM error." + 303,"DG drain pump direction FPGA fault." + 304,"HD invalid usage record." + 305,"HD processor RAM error." + 306,"HD air trap over-fill alarm." + 307,"HD AC power lost second alarm." + 308,"DG CPi conductivity sensor invalid character received." + 309,"DG CPo conductivity sensor invalid character received." + 310,"DG CD1 conductivity sensor invalid character received." + 311,"DG CD2 conductivity sensor invalid character received." + 312,"HD Dialysate In flow rate to too low." + 313,"DG cleaning mode inlet water temperature too low." + 314,"DG cleaning mode inlet water pressure too high." + 315,"DG cleaning mode inlet water pressure too low." + 316,"DG acid concentrate pump park fault." + 317,"DG bicarb concentrate pump part fault." + 318,"HD Air Pump timeout alarm." + 319,"DG chemical disinfect flush flush sample." + 320,"DG chemical disinfect flush sample timeout." + 321,"Total number of alarms."