Index: AlarmMapping.csv =================================================================== diff -u -re5577fbd781ded0678882680d5d31f08f56c9c68 -r3c10146198b48da18d2e8b242e0605461fdc92e8 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision e5577fbd781ded0678882680d5d31f08f56c9c68) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 3c10146198b48da18d2e8b242e0605461fdc92e8) @@ -8,7 +8,7 @@ 7,"HD RTC battery low." 8,"HD accelerometer failed POST." 9,"DG accelerometer failed POST." - 10,"Not used." + 10,"HD blood leak sensor setting embedded mode failure." 11,"HD dialysate temperature below target temperature." 12,"HD firmware image integrity POST test failed." 13,"DG firmware image integrity POST test failed." @@ -25,7 +25,7 @@ 24,"DG bicarb concentrate calibration record." 25,"DG accelerometer invalid calibration record." 26,"HD accelerometer invalid calibration record." - 27,"Not used." + 27,"HD blood leak sensor zero and self test sequence failed." 28,"Not used." 29,"HD heparin force sensor invalid calibration record." 30,"HD Software fault.\nSoftware found itself in an unexpected state." @@ -70,8 +70,8 @@ 69,"HD blood leak sensor invalid calibration record." 70,"HD arterial pressure self-test failure alarm." 71,"HD venous pressure self-test failure alarm." - 72,"Not used." - 73,"Not used." + 72,"HD No load cell data message receive at least once every 2 seconds." + 73,"HD No dialysate temperature data message receive at least once every 2 seconds." 74,"HD blood leak detector self-test transmit timeout." 75,"HD syringe pump self-test failure alarm." 76,"HD monitored voltage is out of range." @@ -153,8 +153,8 @@ 152,"Inlet water pressure in low range." 153,"HD prime completed high priority alarm." 154,"EEPROM operations (read, write, erase) failure." - 155,"Not used." - 156,"Not used." + 155,"HD No reservoirs data message receive at least once every 2 seconds." + 156,"HD No DG operation mode message receive at least once every 2 seconds." 157,"DG chemical disinfect prime acid line timeout." 158,"Inlet water temperature is in the low range." 159,"Dialysate inlet pump failed flow vs motor speed check." @@ -226,7 +226,7 @@ 225,"HD end of treatment warning." 226,"HD prime completed low priority alarm." 227,"HD prime out of time alarm." - 228,"HD time out on prime sub-mode purge air state." + 228,"HD time out on prime saline purge air state." 229,"HD prime dialysate dialyzer time out alarm." 230,"HD prime dialysate bypass time out alarm." 231,"HD pre-treatment mode dry pressure self-test failure."