Index: AlarmMapping.csv =================================================================== diff -u -reaf21ffe52c818b4c8abdb2084582ada9dc78ceb -r75219a6e89b20f405aad0c12c5aed5937118b2ad --- AlarmMapping.csv (.../AlarmMapping.csv) (revision eaf21ffe52c818b4c8abdb2084582ada9dc78ceb) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision 75219a6e89b20f405aad0c12c5aed5937118b2ad) @@ -90,7 +90,7 @@ 89,"HD BP occlusion sensor error." 90,"This alarm ID is available for use." 91,"HD No dialysate flow data receive in the last 3 seconds." - 92,"This alarm ID is available for use." + 92,"DG conductivity sensor bad status." 93,"This alarm ID is available for use." 94,"This alarm ID is available for use." 95,"This alarm ID is available for use." @@ -155,7 +155,7 @@ 154,"EEPROM operations (read, write, erase) failure." 155,"DG software configuration record invalid CRC." 156,"HW usage data (treatment time in HD and total consumed water in DG) failure." - 157,"This alarm ID is available for use." + 157,"DG chemical disinfect prime acid line timeout." 158,"This alarm ID is available for use." 159,"Dialysate inlet pump failed flow vs motor speed check." 160,"Blood pump rotor speed too high."