Index: AlarmMapping.csv =================================================================== diff -u -rebf0b149223da8884c9c55380365f36160a0f9e9 -rf7ca683570f12ffe06ba685d0ad2cb79b1ecba36 --- AlarmMapping.csv (.../AlarmMapping.csv) (revision ebf0b149223da8884c9c55380365f36160a0f9e9) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision f7ca683570f12ffe06ba685d0ad2cb79b1ecba36) @@ -8,7 +8,7 @@ 7,"HD RTC battery low." 8,"HD accelerometer failed POST." 9,"DG accelerometer failed POST." - 10,"Not used." + 10,"HD blood leak sensor setting embedded mode failure." 11,"HD dialysate temperature below target temperature." 12,"HD firmware image integrity POST test failed." 13,"DG firmware image integrity POST test failed." @@ -25,8 +25,8 @@ 24,"DG bicarb concentrate calibration record." 25,"DG accelerometer invalid calibration record." 26,"HD accelerometer invalid calibration record." - 27,"Not used." - 28,"Not used." + 27,"HD blood leak sensor zero and self test sequence failed." + 28,"DG two wire sensors FPGA fault." 29,"HD heparin force sensor invalid calibration record." 30,"HD Software fault.\nSoftware found itself in an unexpected state." 31,"Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." @@ -41,14 +41,14 @@ 40,"Dialysis outlet pump failed motor off check.\nMeasured speed while commanded off." 41,"Dialysis outlet pump failed motor direction check.\nMeasured vs commanded." 42,"Dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds." - 43,"HD blood leak self test fault." + 43,"DG main primary heater FPGA fault." 44,"User interface communication timeout." 45,"HD too many bad communications CRC." - 46,"System message that required acknowledgment was not acknowledged." + 46,"HD System message that required acknowledgment was not acknowledged." 47,"Ultrafiltration rate is too high error during treatment." 48,"Ultrafiltration volume accuracy error during treatment." 49,"HD FPGA communication down for too long." - 50,"FPGA not accepting commanded valve states." + 50,"DG FPGA not accepting commanded valve states." 51,"Blood pump failed motor speed check.\nMeasured vs commanded." 52,"Dialysate inlet pump failed motor speed check.\nMeasured vs commanded." 53,"Dialysate outlet pump failed motor speed check.\nMeasured vs commanded." @@ -61,8 +61,8 @@ 60,"HD valve not functional." 61,"HD valve current out of range." 62,"HD valve position out of target." - 63,"FPGA reports failure reading arterial pressure sensor." - 64,"FPGA reports failure reading venous pressure sensor." + 63,"HD blood leak FPGA fault." + 64,"DG barometric pressure sensor out of range." 65,"HD requests DG command with invalid parameter fault." 66,"HD blood leak sensor set point set failure." 67,"HD blood pump occlusion self-test failure alarm." @@ -72,7 +72,7 @@ 71,"HD venous pressure self-test failure alarm." 72,"HD No load cell data message receive at least once every 2 seconds." 73,"HD No dialysate temperature data message receive at least once every 2 seconds." - 74,"HD blood leak detector self-test transmit timeout." + 74,"DG inlet UV reactor is on with no flow." 75,"HD syringe pump self-test failure alarm." 76,"HD monitored voltage is out of range." 77,"DG monitored voltage is out of range." @@ -82,13 +82,13 @@ 81,"HD syringe pump over travel error." 82,"HD syringe pump DAC write failure." 83,"HD syringe pump is running while the BP is off." - 84,"HD venous pressure sensor not being read." + 84,"DG set RTC year is invalid." 85,"HD pump track latch opened alarm." - 86,"HD BP occlusion sensor not being read." + 86,"HD set RTC year is invalid." 87,"DG heating invalid calibration record." 88,"DG concentrate pumps hall sensor out of range." - 89,"HD BP occlusion sensor error." - 90,"DG load cells FPGA read error." + 89,"DG outlet UV reactor on with no flow." + 90,"DG load cells A1/B1 FPGA fault." 91,"HD No dialysate flow data receive in the last 3 seconds." 92,"DG conductivity sensor bad status." 93,"DG outlet primary conductivity out of range." @@ -106,12 +106,12 @@ 105,"DG invalid calibration CRC." 106,"DG dialysate flow sensor invalid calibration record." 107,"HD reports DG restarted fault." - 108,"HD syringe pump ADC error." + 108,"HD syringe pump ADC FPGA fault." 109,"HD syringe pump volume check error." 110,"HD syringe pump speed check error." 111,"HD syringe pump not stopped in off state error." - 112,"HD blood leak zero command transmit timeout." - 113,"HD blood leak zero command fault." + 112,"DG load cells A1/B1 FPGA fault." + 113,"DG Drain pump current out of range." 114,"HD venous air bubble detector self-test failure." 115,"DG temperature sensor out of range." 116,"DG temperature sensor ADC fault." @@ -142,7 +142,7 @@ 141,"Air trap fill timeout during treatment." 142,"Blood pump occlusion detected." 143,"DG dialysate temperature sensors out of range." - 144,"DG RO flow sensor invalid calibration record." + 144,"Alarm ID available for use." 145,"Concentrate conductivity after adding acid out of range alarm." 146,"DG RTC or timer accuracy failure." 147,"Wait for the DG to produce dialysate." @@ -167,7 +167,7 @@ 166,"HD in treatment re-circ sub-mode for too long." 167,"HD cartridge door opened alarm." 168,"HD active reservoir depletion timeout." - 169,"Dialysate flow rate out of range." + 169,"DG Dialysate flow rate out of maximum range." 170,"HD syringe empty alarm." 171,"HD syringe pump occlusion alarm." 172,"HD syringe pump not enough Heparin alarm." @@ -178,7 +178,7 @@ 177,"DG software fault.\nSoftware found itself in an unexpected state." 178,"HD communication timeout." 179,"DG FPGA communication down for too long." - 180,"DG load cell ADC error." + 180,"DG RO flow out of maximum range." 181,"DG load cells weight out of range for tare." 182,"DG load cells invalid calibration." 183,"DG invalid load cell value." @@ -194,8 +194,8 @@ 193,"HD dialysate temperature out of low safety range." 194,"DG RO pump duty cycle out of range." 195,"DG RO pump pressure out of range." - 196,"DG temperature sensor error bit fault." - 197,"DG temperature sensors FPGA fault." + 196,"DG CPi/CPo sensors FPGA fault." + 197,"DG CD1/CD2 sensors FPGA fault." 198,"RO flow too low while primary heater is on." 199,"Dialysate flow too low while trimmer heater is on." 200,"DG thermistors/sensors temperature out of range." @@ -218,10 +218,10 @@ 217,"HD blood prime volume check failure." 218,"HD syringe detected alarm." 219,"HD syringe pump syringe removed alarm." - 220,"HD battery pack SOC is too low." + 220,"DG THd sensors FPGA fault." 221,"HD Saline bag is empty." - 222,"HD dialysate temperature too high alarm." - 223,"HD dialysate temperature too low alarm." + 222,"HD occlusion sensor FPGA fault." + 223,"HD arterial sensor FPGA fault." 224,"Treatment stopped by user action." 225,"HD end of treatment warning." 226,"HD prime completed low priority alarm." @@ -278,7 +278,7 @@ 277,"HD BP occlusion sensor is reading out of range." 278,"HD active reservoir weight out of range." 279,"DG dialysate drain time out." - 280,"HD arterial pressure sensor read timeout error." + 280,"DG drain pump direction invalid." 281,"Acid concentration bottle low volume alarm." 282,"Bicarbonate concentration bottle low volume alarm." 283,"DG load cells weight out of range." @@ -288,10 +288,27 @@ 287,"HD processor clock speed checks against FPGA clock failure." 288,"HD load cells primary/back up drift out of range." 289,"DG dialysate or concentrate caps not closed." - 290,"DG flow too low while heater is on." + 290,"HD venous pressure sensor FPGA fault." 291,"HD processor is in Disinfect Flush mode." 292,"HD processor is in Disinfect Heat mode." 293,"HD processor is in Disinfect Chemical mode." 294,"HD processor Disinfect Chemical mode instructions to user." 295,"UI POST CloudSync failure." - 296,"Total number of alarms." + 296,"DG TDi sensors FPGA fault." + 297,"DG TRo sensors FPGA fault." + 298,"DG baro sensor FPGA fault." + 299,"DG invalid serial number." + 300,"HD invalid serial number." + 301,"DG primary heater power voltage out of range." + 302,"DG processor RAM error." + 303,"DG drain pump direction FPGA fault." + 304,"HD invalid usage record." + 305,"HD processor RAM error." + 306,"HD air trap over-fill alarm." + 307,"HD AC power lost second alarm." + 308,"DG CPi conductivity sensor invalid character received." + 309,"DG CPo conductivity sensor invalid character received." + 310,"DG CD1 conductivity sensor invalid character received." + 311,"DG CD2 conductivity sensor invalid character received." + 312,"HD Dialysate In flow rate to too low." + 313,"Total number of alarms."