Index: AlarmMapping.csv =================================================================== diff -u -red9ea343a2a22623796045d9e1229a2cbfa8c970 -rfd1b031dd0c66addaa751f5d69dfef46bdfc6b7a --- AlarmMapping.csv (.../AlarmMapping.csv) (revision ed9ea343a2a22623796045d9e1229a2cbfa8c970) +++ AlarmMapping.csv (.../AlarmMapping.csv) (revision fd1b031dd0c66addaa751f5d69dfef46bdfc6b7a) @@ -83,7 +83,7 @@ 82,"HD syringe pump DAC write failure." 83,"HD syringe pump is running while the BP is off." 84,"HD venous pressure sensor not being read." - 85,"This alarm ID is available for use." + 85,"HD pump track latch opened alarm." 86,"HD BP occlusion sensor not being read." 87,"DG heating invalid calibration record." 88,"DG concentrate pumps hall sensor out of range." @@ -148,14 +148,14 @@ 147,"DG dialysate generation conductivity fault alarm." 148,"Inlet water temperature too high." 149,"Inlet water temperature too low." - 150,"Inlet water conductivity is in warning range." + 150,"Inlet water conductivity too high." 151,"Inlet water conductivity too low." 152,"Inlet water pressure too low." 153,"HD prime completed high priority alarm." 154,"EEPROM operations (read, write, erase) failure." 155,"DG software configuration record invalid CRC." 156,"HW usage data (treatment time in HD and total consumed water in DG) failure." - 157,"Inlet water temperature is not within specified range." + 157,"This alarm ID is available for use." 158,"This alarm ID is available for use." 159,"Dialysate inlet pump failed flow vs motor speed check." 160,"Blood pump rotor speed too high."