Index: sources/gui/qml/globals/Variables.qml =================================================================== diff -u -r76f632ce5a9039ae662d9a52286079b772f8ae7f -r25e7d601d795c1622dd4c6e437b3dce91a676e83 --- sources/gui/qml/globals/Variables.qml (.../Variables.qml) (revision 76f632ce5a9039ae662d9a52286079b772f8ae7f) +++ sources/gui/qml/globals/Variables.qml (.../Variables.qml) (revision 25e7d601d795c1622dd4c6e437b3dce91a676e83) @@ -137,12 +137,10 @@ readonly property int arterialValueMax : +600; ///< PRS362 const // - Artrial - Adjust pressure limit range and manufacturing default - Low readonly property int arterialLimitLowMin : -300; ///< PRS354 const - // ref to Model // arterialLimitLowMin : -300; ///< PRS354 default \ref MAdjustPressuresLimitsReq - readonly property int arterialLimitLowMax : +200; ///< PRS354 const + readonly property int arterialLimitLowMax : 0; ///< PRS354 const // - Artrial - Adjust pressure limit range and manufacturing default - High readonly property int arterialLimitHighMin : -300; ///< PRS355 const - // ref to Model // arterialLimitHighMin : +100; ///< PRS355 default \ref MAdjustPressuresLimitsReq - readonly property int arterialLimitHighMax : +200; ///< PRS355 const + readonly property int arterialLimitHighMax : 0; ///< PRS355 const // - Artrial - Adjust pressure limit resolution readonly property int arterialLimitStep : +10; ///< PRS354,PRS355 // has to be the same since it's on the same line scale readonly property int arterialLimitGap : +30; @@ -152,12 +150,10 @@ readonly property int venousValueMin : -100; ///< PRS361 const readonly property int venousValueMax : +700; ///< PRS361 const // - Venous - Adjust pressure limit range and manufacturing default - Low - readonly property int venousLimitLowMin : -100; ///< PRS356 const - // ref to Model // venousLimitLowDef : -100; ///< PRS356 default \ref MAdjustPressuresLimitsReq + readonly property int venousLimitLowMin : +20; ///< PRS356 const readonly property int venousLimitLowMax : +600; ///< PRS356 const // - Venous - Adjust pressure limit range and manufacturing default - High - readonly property int venousLimitHighMin : +100; ///< PRS357 const - // ref to Model // venousLimitHighDef : +400; ///< PRS357 default \ref MAdjustPressuresLimitsReq + readonly property int venousLimitHighMin : +20; ///< PRS357 const readonly property int venousLimitHighMax : +600; ///< PRS357 const // - Venous - Adjust pressure limit resolution readonly property int venousLimitStep : +10; ///< PRS356,PRS357 // has to be the same since it's on the same line scale