Index: sources/model/hd/alarm/MAlarmMapping.cpp =================================================================== diff -u -r1861a3618bbfee48166aad0390dfb59524d48459 -r200398eaf784dd2965023151086ba1939d2eb746 --- sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 1861a3618bbfee48166aad0390dfb59524d48459) +++ sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 200398eaf784dd2965023151086ba1939d2eb746) @@ -7,7 +7,7 @@ * * \file MAlarmMapping.cpp * \author (last) Behrouz NematiPour - * \date (last) 21-Dec-2021 + * \date (last) 07-Feb-2022 * \author (original) Behrouz NematiPour * \date (original) 03-May-2021 * @@ -99,10 +99,10 @@ /*0063*/case GuiAlarmID::ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT : { result = QObject::tr("FPGA reports failure reading arterial pressure sensor." ); break; } /* 63*/ /*0064*/case GuiAlarmID::ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT : { result = QObject::tr("FPGA reports failure reading venous pressure sensor." ); break; } /* 64*/ /*0065*/case GuiAlarmID::ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT : { result = QObject::tr("HD requests DG command with invalid parameter fault." ); break; } /* 65*/ -/*0066*/case GuiAlarmID::ALARM_ID____AVAILABLE_21 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 66*/ +/*0066*/case GuiAlarmID::ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE : { result = QObject::tr("HD blood leak sensor set point set failure." ); break; } /* 66*/ /*0067*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE : { result = QObject::tr("HD blood pump occlusion self-test failure alarm." ); break; } /* 67*/ /*0068*/case GuiAlarmID::ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE : { result = QObject::tr("HD active reservoir recirculation out of range." ); break; } /* 68*/ -/*0069*/case GuiAlarmID::ALARM_ID____AVAILABLE_6 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 69*/ +/*0069*/case GuiAlarmID::ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD : { result = QObject::tr("HD blood leak sensor invalid calibration record." ); break; } /* 69*/ /*0070*/case GuiAlarmID::ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE : { result = QObject::tr("HD arterial pressure self-test failure alarm." ); break; } /* 70*/ /*0071*/case GuiAlarmID::ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE : { result = QObject::tr("HD venous pressure self-test failure alarm." ); break; } /* 71*/ /*0072*/case GuiAlarmID::ALARM_ID_HD_BLOOD_FLOW_STATUS_SELF_TEST_FAILURE : { result = QObject::tr("HD blood flow meter status check self-test failure alarm." ); break; } /* 72*/ @@ -120,10 +120,10 @@ /*0084*/case GuiAlarmID::ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR : { result = QObject::tr("HD venous pressure sensor not being read." ); break; } /* 84*/ /*0085*/case GuiAlarmID::ALARM_ID_HD_VENOUS_PRESSURE_SENSOR_TEMP_OUT_OF_RANGE : { result = QObject::tr("HD venous pressure sensor temperature out of range error." ); break; } /* 85*/ /*0086*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR : { result = QObject::tr("HD BP occlusion sensor not being read." ); break; } /* 86*/ -/*0087*/case GuiAlarmID::ALARM_ID____AVAILABLE_10 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 87*/ -/*0088*/case GuiAlarmID::ALARM_ID____AVAILABLE_11 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 88*/ +/*0087*/case GuiAlarmID::ALARM_ID_DG_HEATING_INVALID_CAL_RECORD : { result = QObject::tr("DG heating invalid calibration record." ); break; } /* 87*/ +/*0088*/case GuiAlarmID::ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE : { result = QObject::tr("DG concentrate pumps hall sensor out of range." ); break; } /* 88*/ /*0089*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR : { result = QObject::tr("HD BP occlusion sensor error." ); break; } /* 89*/ -/*0090*/case GuiAlarmID::ALARM_ID____AVAILABLE_12 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 90*/ +/*0090*/case GuiAlarmID::ALARM_ID____AVAILABLE_25 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 90*/ /*0091*/case GuiAlarmID::ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE : { result = QObject::tr("HD No dialysate flow data receive in the last 3 seconds." ); break; } /* 91*/ /*0092*/case GuiAlarmID::ALARM_ID____AVAILABLE_14 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 92*/ /*0093*/case GuiAlarmID::ALARM_ID____AVAILABLE_15 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 93*/ @@ -176,8 +176,8 @@ /*0140*/case GuiAlarmID::ALARM_ID_DG_COMM_TIMEOUT : { result = QObject::tr("Dialysate generator communication timeout." ); break; } /* 140*/ /*0141*/case GuiAlarmID::ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT : { result = QObject::tr("Air trap fill timeout during treatment." ); break; } /* 141*/ /*0142*/case GuiAlarmID::ALARM_ID_OCCLUSION_BLOOD_PUMP : { result = QObject::tr("Blood pump occlusion detected." ); break; } /* 142*/ -/*0143*/case GuiAlarmID::ALARM_ID____AVAILABLE_8 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 143*/ -/*0144*/case GuiAlarmID::ALARM_ID____AVAILABLE_9 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 144*/ +/*0143*/case GuiAlarmID::ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_DRFIT_TIMEOUT : { result = QObject::tr("DG dialysate temperature sensors drift timeout." ); break; } /* 143*/ +/*0144*/case GuiAlarmID::ALARM_ID_DG_SW_CONFIG_RECORD_INVALID_CRC : { result = QObject::tr("DG software configuration record invalid CRC." ); break; } /* 144*/ /*0145*/case GuiAlarmID::ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE : { result = QObject::tr("Concentrate conductivity after adding acid out of range alarm." ); break; } /* 145*/ /*0146*/case GuiAlarmID::ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE : { result = QObject::tr("Dialysate conductivity out of range." ); break; } /* 146*/ /*0147*/case GuiAlarmID::ALARM_ID_DIALYSATE_CONDUCTIVITY_FAULT : { result = QObject::tr("DG dialysate generation conductivity fault alarm." ); break; } /* 147*/ @@ -188,10 +188,10 @@ /*0152*/case GuiAlarmID::ALARM_ID_INLET_WATER_LOW_PRESSURE : { result = QObject::tr("Inlet water pressure too low." ); break; } /* 152*/ /*0153*/case GuiAlarmID::ALARM_ID_PRIME_COMPLETED_HIGH : { result = QObject::tr("HD prime completed high priority alarm." ); break; } /* 153*/ /*0154*/case GuiAlarmID::ALARM_ID_NVDATA_EEPROM_OPS_FAILURE : { result = QObject::tr("EEPROM operations (read, write, erase) failure." ); break; } /* 154*/ -/*0155*/case GuiAlarmID::ALARM_ID____AVAILABLE_22 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 155*/ +/*0155*/case GuiAlarmID::ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC : { result = QObject::tr("DG software configuration record invalid CRC." ); break; } /* 155*/ /*0156*/case GuiAlarmID::ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR : { result = QObject::tr("HW usage data (treatment time in HD and total consumed water in DG) failure." ); break; } /* 156*/ /*0157*/case GuiAlarmID::ALARM_ID____AVAILABLE_23 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 157*/ -/*0158*/case GuiAlarmID::ALARM_ID____AVAILABLE_24 : { result = QObject::tr("Blood pump failed flow vs motor speed check." ); break; } /* 158*/ +/*0158*/case GuiAlarmID::ALARM_ID____AVAILABLE_24 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 158*/ /*0159*/case GuiAlarmID::ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK : { result = QObject::tr("Dialysate inlet pump failed flow vs motor speed check." ); break; } /* 159*/ /*0160*/case GuiAlarmID::ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH : { result = QObject::tr("Blood pump rotor speed too high." ); break; } /* 160*/ /*0161*/case GuiAlarmID::ALARM_ID_BLOOD_FLOW_SIGNAL_STRENGTH_TOO_LOW : { result = QObject::tr("Blood flow sensor signal strength too low." ); break; } /* 161*/