Index: sources/model/hd/alarm/MAlarmMapping.cpp =================================================================== diff -u -r1d3b5f0cf1ed5297d7cb66ff461da292c4d8766c -r3c8921f4ef60267a6d3c0c052255315031c2792d --- sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 1d3b5f0cf1ed5297d7cb66ff461da292c4d8766c) +++ sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 3c8921f4ef60267a6d3c0c052255315031c2792d) @@ -7,7 +7,7 @@ * * \file MAlarmMapping.cpp * \author (last) Behrouz NematiPour - * \date (last) 07-Oct-2022 + * \date (last) 10-Oct-2022 * \author (original) Behrouz NematiPour * \date (original) 03-May-2021 * @@ -107,7 +107,7 @@ /*0071*/case GuiAlarmID::ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE : { result = QObject::tr("HD venous pressure self-test failure alarm." ); break; } /* 71*/ /*0072*/case GuiAlarmID::ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE : { result = QObject::tr("HD No load cell data message receive at least once every 2 seconds." ); break; } /* 72*/ /*0073*/case GuiAlarmID::ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE: { result = QObject::tr("HD No dialysate temperature data message receive at least once every 2 seconds." ); break; } /* 73*/ -/*0074*/case GuiAlarmID::ALARM_ID__AVAILABLE_6 : { result = QObject::tr("Alarm ID available." ); break; } /* 74*/ +/*0074*/case GuiAlarmID::ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW : { result = QObject::tr("DG inlet UV reactor is on with no flow." ); break; } /* 74*/ /*0075*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE : { result = QObject::tr("HD syringe pump self-test failure alarm." ); break; } /* 75*/ /*0076*/case GuiAlarmID::ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE : { result = QObject::tr("HD monitored voltage is out of range." ); break; } /* 76*/ /*0077*/case GuiAlarmID::ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE : { result = QObject::tr("DG monitored voltage is out of range." ); break; } /* 77*/ @@ -122,7 +122,7 @@ /*0086*/case GuiAlarmID::ALARM_ID_HD_SET_RTC_YEAR_INVALID : { result = QObject::tr("HD set RTC year is invalid." ); break; } /* 86*/ /*0087*/case GuiAlarmID::ALARM_ID_DG_HEATING_INVALID_CAL_RECORD : { result = QObject::tr("DG heating invalid calibration record." ); break; } /* 87*/ /*0088*/case GuiAlarmID::ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE : { result = QObject::tr("DG concentrate pumps hall sensor out of range." ); break; } /* 88*/ -/*0089*/case GuiAlarmID::ALARM_ID__AVAILABLE_15 : { result = QObject::tr("Alarm ID available." ); break; } /* 89*/ +/*0089*/case GuiAlarmID::ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW : { result = QObject::tr("DG outlet UV reactor on with no flow." ); break; } /* 89*/ /*0090*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT : { result = QObject::tr("DG load cells A1/B1 FPGA fault." ); break; } /* 90*/ /*0091*/case GuiAlarmID::ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE : { result = QObject::tr("HD No dialysate flow data receive in the last 3 seconds." ); break; } /* 91*/ /*0092*/case GuiAlarmID::ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS : { result = QObject::tr("DG conductivity sensor bad status." ); break; } /* 92*/ @@ -146,7 +146,7 @@ /*0110*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR : { result = QObject::tr("HD syringe pump speed check error." ); break; } /* 110*/ /*0111*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR : { result = QObject::tr("HD syringe pump not stopped in off state error." ); break; } /* 111*/ /*0112*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT : { result = QObject::tr("DG load cells A1/B1 FPGA fault." ); break; } /* 112*/ -/*0113*/case GuiAlarmID::ALARM_ID__AVAILABLE_8 : { result = QObject::tr("Alarm ID available for use." ); break; } /* 113*/ +/*0113*/case GuiAlarmID::ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE : { result = QObject::tr("DG Drain pump current out of range." ); break; } /* 113*/ /*0114*/case GuiAlarmID::ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE : { result = QObject::tr("HD venous air bubble detector self-test failure." ); break; } /* 114*/ /*0115*/case GuiAlarmID::ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE : { result = QObject::tr("DG temperature sensor out of range." ); break; } /* 115*/ /*0116*/case GuiAlarmID::ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT : { result = QObject::tr("DG temperature sensor ADC fault." ); break; } /* 116*/ @@ -313,7 +313,7 @@ /*0277*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE : { result = QObject::tr("HD BP occlusion sensor is reading out of range." ); break; } /* 277*/ /*0278*/case GuiAlarmID::ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE : { result = QObject::tr("HD active reservoir weight out of range." ); break; } /* 278*/ /*0279*/case GuiAlarmID::ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT : { result = QObject::tr("DG dialysate drain time out." ); break; } /* 279*/ -/*0280*/case GuiAlarmID::ALARM_ID__AVAILABLE_10 : { result = QObject::tr("Alarm ID available for use." ); break; } /* 280*/ +/*0280*/case GuiAlarmID::ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID : { result = QObject::tr("DG drain pump direction invalid." ); break; } /* 280*/ /*0281*/case GuiAlarmID::ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME : { result = QObject::tr("Acid concentration bottle low volume alarm." ); break; } /* 281*/ /*0282*/case GuiAlarmID::ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME : { result = QObject::tr("Bicarbonate concentration bottle low volume alarm." ); break; } /* 282*/ /*0283*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE : { result = QObject::tr("DG load cells weight out of range." ); break; } /* 283*/