Index: sources/model/hd/alarm/MAlarmMapping.cpp =================================================================== diff -u -r6ea9e90a8da8e207dada0d9a22b2d9523a95fe10 -rfe28eeae194ab09dfe2a9604206fd0b784f2c460 --- sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 6ea9e90a8da8e207dada0d9a22b2d9523a95fe10) +++ sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision fe28eeae194ab09dfe2a9604206fd0b784f2c460) @@ -183,14 +183,14 @@ /*0147*/case GuiAlarmID::ALARM_ID_DIALYSATE_CONDUCTIVITY_FAULT : { result = QObject::tr("DG dialysate generation conductivity fault alarm." ); break; } /* 147*/ /*0148*/case GuiAlarmID::ALARM_ID_INLET_WATER_HIGH_TEMPERATURE : { result = QObject::tr("Inlet water temperature too high." ); break; } /* 148*/ /*0149*/case GuiAlarmID::ALARM_ID_INLET_WATER_LOW_TEMPERATURE : { result = QObject::tr("Inlet water temperature too low." ); break; } /* 149*/ -/*0150*/case GuiAlarmID::ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY : { result = QObject::tr("Inlet water conductivity too high." ); break; } /* 150*/ +/*0150*/case GuiAlarmID::ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_WARNING_RANGE : { result = QObject::tr("Inlet water conductivity is in warning range." ); break; } /* 150*/ /*0151*/case GuiAlarmID::ALARM_ID_INLET_WATER_LOW_CONDUCTIVITY : { result = QObject::tr("Inlet water conductivity too low." ); break; } /* 151*/ /*0152*/case GuiAlarmID::ALARM_ID_INLET_WATER_LOW_PRESSURE : { result = QObject::tr("Inlet water pressure too low." ); break; } /* 152*/ /*0153*/case GuiAlarmID::ALARM_ID_PRIME_COMPLETED_HIGH : { result = QObject::tr("HD prime completed high priority alarm." ); break; } /* 153*/ /*0154*/case GuiAlarmID::ALARM_ID_NVDATA_EEPROM_OPS_FAILURE : { result = QObject::tr("EEPROM operations (read, write, erase) failure." ); break; } /* 154*/ /*0155*/case GuiAlarmID::ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC : { result = QObject::tr("DG software configuration record invalid CRC." ); break; } /* 155*/ /*0156*/case GuiAlarmID::ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR : { result = QObject::tr("HW usage data (treatment time in HD and total consumed water in DG) failure." ); break; } /* 156*/ -/*0157*/case GuiAlarmID::ALARM_ID____AVAILABLE_23 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 157*/ +/*0157*/case GuiAlarmID::ALARM_ID_INLET_WATER_TEMPERATURE_OUT_OF_RANGE : { result = QObject::tr("Inlet water temperature is not within specified range." ); break; } /* 157*/ /*0158*/case GuiAlarmID::ALARM_ID____AVAILABLE_24 : { result = QObject::tr("This alarm ID is available for use." ); break; } /* 158*/ /*0159*/case GuiAlarmID::ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK : { result = QObject::tr("Dialysate inlet pump failed flow vs motor speed check." ); break; } /* 159*/ /*0160*/case GuiAlarmID::ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH : { result = QObject::tr("Blood pump rotor speed too high." ); break; } /* 160*/