Index: sources/model/hd/alarm/MAlarmMapping.cpp =================================================================== diff -u -rac49f0bd00d33f7326e82ae2d213e1b21e9083d4 -r9d4fb83207bd61a6030ebdd5a3e7002410fc1124 --- sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision ac49f0bd00d33f7326e82ae2d213e1b21e9083d4) +++ sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 9d4fb83207bd61a6030ebdd5a3e7002410fc1124) @@ -125,15 +125,15 @@ /*0089*/case GuiAlarmID::ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW : { result = QObject::tr("DG outlet UV reactor on with no flow." ); break; } /* 89*/ /*0090*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT : { result = QObject::tr("DG load cells A1/B1 FPGA fault." ); break; } /* 90*/ /*0091*/case GuiAlarmID::ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE : { result = QObject::tr("HD No dialysate flow data receive in the last 3 seconds." ); break; } /* 91*/ -/*0092*/case GuiAlarmID::ALARM_ID_DG_TEMPERATURE_SENSORS_INVALID_CAL_RECORD : { result = QObject::tr("DG temperature sensors invalid calibration record." ); break; } /* 92*/ +/*0092*/case GuiAlarmID::ALARM_ID_AVAILABLE_1 : { result = QObject::tr("Available for use." ); break; } /* 92*/ /*0093*/case GuiAlarmID::ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE : { result = QObject::tr("DG outlet primary conductivity out of range." ); break; } /* 93*/ /*0094*/case GuiAlarmID::ALARM_ID_DG_PRESSURE_OUT_OF_RANGE : { result = QObject::tr("DG pressure out of range." ); break; } /* 94*/ /*0095*/case GuiAlarmID::ALARM_ID_DG_WATCHDOG_EXPIRED : { result = QObject::tr("DG watchdog expired." ); break; } /* 95*/ /*0096*/case GuiAlarmID::ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE : { result = QObject::tr("DG inlet water temperature in high range." ); break; } /* 96*/ /*0097*/case GuiAlarmID::ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE : { result = QObject::tr("DG fill conductivity out of range." ); break; } /* 97*/ /*0098*/case GuiAlarmID::ALARM_ID_HD_BATTERY_COMM_FAULT : { result = QObject::tr("HD battery communication fault." ); break; } /* 98*/ /*0099*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_STALL : { result = QObject::tr("HD syringe pump stall alarm." ); break; } /* 99*/ -/*0100*/case GuiAlarmID::ALARM_ID_DG_COND_SENSORS_INVALID_TEMP_COMP_CAL_RECORD : { result = QObject::tr("DG conductivity sensors invalid temperature compensation calibration record." ); break; } /* 100*/ +/*0100*/case GuiAlarmID::ALARM_ID_AVAILABLE_12 : { result = QObject::tr("Available for use." ); break; } /* 100*/ /*0101*/case GuiAlarmID::ALARM_ID_UI_POST_HD_COMM : { result = QObject::tr("UI POST HD communication failure." ); break; } /* 101*/ /*0102*/case GuiAlarmID::ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE : { result = QObject::tr("DG heat disinfect temperature gradient out of range." ); break; } /* 102*/ /*0103*/case GuiAlarmID::ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID : { result = QObject::tr("HD invalid calibration CRC." ); break; } /* 103*/ @@ -307,7 +307,7 @@ /*0271*/case GuiAlarmID::ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED : { result = QObject::tr("DG POST Safety Shutdown failure." ); break; } /* 271*/ /*0272*/case GuiAlarmID::ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE : { result = QObject::tr("HD Fan RPM out of range." ); break; } /* 272*/ /*0273*/case GuiAlarmID::ALARM_ID_AVAILABLE_10 : { result = QObject::tr("Available for use." ); break; } /* 273*/ -/*0274*/case GuiAlarmID::ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE : { result = QObject::tr("DG inactive reservoir weight out of range." ); break; } /* 274*/ +/*0274*/case GuiAlarmID::ALARM_ID_AVAILABLE_11 : { result = QObject::tr("Available for use." ); break; } /* 274*/ /*0275*/case GuiAlarmID::ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE : { result = QObject::tr("HD arterial pressure sensor is reading out of range." ); break; } /* 275*/ /*0276*/case GuiAlarmID::ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE : { result = QObject::tr("HD venous pressure sensor is reading out of range." ); break; } /* 276*/ /*0277*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE : { result = QObject::tr("HD BP occlusion sensor is reading out of range." ); break; } /* 277*/