Index: sources/model/hd/alarm/MAlarmMapping.cpp =================================================================== diff -u -re961fc7bcdeba8ffab9c96da20244b07405f8f99 -r7992f1a517435b1801dbee84e5ffc38b8452d2dd --- sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision e961fc7bcdeba8ffab9c96da20244b07405f8f99) +++ sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 7992f1a517435b1801dbee84e5ffc38b8452d2dd) @@ -1,17 +1,23 @@ -/*! - * - * Copyright (c) 2021-2022 Diality Inc. - All Rights Reserved. - * \copyright - * THIS CODE MAY NOT BE COPIED OR REPRODUCED IN ANY FORM, IN PART OR IN - * WHOLE, WITHOUT THE EXPLICIT PERMISSION OF THE COPYRIGHT OWNER. - * - * \file MAlarmMapping.cpp - * \author (last) Behrouz NematiPour - * \date (last) 16-Oct-2022 - * \author (original) Behrouz NematiPour - * \date (original) 04-May-2021 - * - */ +/*! + * + * Copyright (c) 2019-2020 Diality Inc. - All Rights Reserved. + * \copyright + * THIS CODE MAY NOT BE COPIED OR REPRODUCED IN ANY FORM, IN PART OR IN + * WHOLE, WITHOUT THE EXPLICIT PERMISSION OF THE COPYRIGHT OWNER. + * + * \file MAlarmMapping.cpp + * \author (last) Behrouz NematiPour + * \date (last) 24-Oct-2022 + * \author (original) Behrouz NematiPour + * \date (original) 03-May-2021 + * + ********************************************************************************* + * ANY MODIFICATION IN THIS FILE WILL BE LOST. * + * THE CONTENT OF THIS FILE IS AUTOMATICALLY GENERATED ON EACH BUILD ON SERVER. * + * ANY REQUIRED UPDATE NEEDS TO BE DONE IN THE AlarmDefs.h IN common REPOSITORY. * + ******************************************************************************* * + * + */ #include "MAlarmStatusData.h" using namespace Model; @@ -70,7 +76,7 @@ /*0040*/case GuiAlarmID::ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK : { result = QObject::tr("Dialysis outlet pump failed motor off check.\nMeasured speed while commanded off." ); break; } /* 40*/ /*0041*/case GuiAlarmID::ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK : { result = QObject::tr("Dialysis outlet pump failed motor direction check.\nMeasured vs commanded." ); break; } /* 41*/ /*0042*/case GuiAlarmID::ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK : { result = QObject::tr("Dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds." ); break; } /* 42*/ -/*0043*/case GuiAlarmID::ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAULT : { result = QObject::tr("HD blood leak self test fault." ); break; } /* 43*/ +/*0043*/case GuiAlarmID::ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT : { result = QObject::tr("DG main primary heater FPGA fault." ); break; } /* 43*/ /*0044*/case GuiAlarmID::ALARM_ID_UI_COMM_TIMEOUT : { result = QObject::tr("User interface communication timeout." ); break; } /* 44*/ /*0045*/case GuiAlarmID::ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS : { result = QObject::tr("HD too many bad communications CRC." ); break; } /* 45*/ /*0046*/case GuiAlarmID::ALARM_ID_CAN_MESSAGE_NOT_ACKED : { result = QObject::tr("System message that required acknowledgment was not acknowledged." ); break; } /* 46*/ @@ -90,8 +96,8 @@ /*0060*/case GuiAlarmID::ALARM_ID_HD_VALVE_NOT_FUNCTIONAL : { result = QObject::tr("HD valve not functional." ); break; } /* 60*/ /*0061*/case GuiAlarmID::ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE : { result = QObject::tr("HD valve current out of range." ); break; } /* 61*/ /*0062*/case GuiAlarmID::ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE : { result = QObject::tr("HD valve position out of target." ); break; } /* 62*/ -/*0063*/case GuiAlarmID::ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT : { result = QObject::tr("FPGA reports failure reading arterial pressure sensor." ); break; } /* 63*/ -/*0064*/case GuiAlarmID::ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT : { result = QObject::tr("FPGA reports failure reading venous pressure sensor." ); break; } /* 64*/ +/*0063*/case GuiAlarmID::ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT : { result = QObject::tr("HD blood leak FPGA fault." ); break; } /* 63*/ +/*0064*/case GuiAlarmID::ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE : { result = QObject::tr("DG barometric pressure sensor out of range." ); break; } /* 64*/ /*0065*/case GuiAlarmID::ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT : { result = QObject::tr("HD requests DG command with invalid parameter fault." ); break; } /* 65*/ /*0066*/case GuiAlarmID::ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE : { result = QObject::tr("HD blood leak sensor set point set failure." ); break; } /* 66*/ /*0067*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE : { result = QObject::tr("HD blood pump occlusion self-test failure alarm." ); break; } /* 67*/ @@ -101,7 +107,7 @@ /*0071*/case GuiAlarmID::ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE : { result = QObject::tr("HD venous pressure self-test failure alarm." ); break; } /* 71*/ /*0072*/case GuiAlarmID::ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE : { result = QObject::tr("HD No load cell data message receive at least once every 2 seconds." ); break; } /* 72*/ /*0073*/case GuiAlarmID::ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE: { result = QObject::tr("HD No dialysate temperature data message receive at least once every 2 seconds." ); break; } /* 73*/ -/*0074*/case GuiAlarmID::ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_SEND_TIMEOUT : { result = QObject::tr("HD blood leak detector self-test transmit timeout." ); break; } /* 74*/ +/*0074*/case GuiAlarmID::ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW : { result = QObject::tr("DG inlet UV reactor is on with no flow." ); break; } /* 74*/ /*0075*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE : { result = QObject::tr("HD syringe pump self-test failure alarm." ); break; } /* 75*/ /*0076*/case GuiAlarmID::ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE : { result = QObject::tr("HD monitored voltage is out of range." ); break; } /* 76*/ /*0077*/case GuiAlarmID::ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE : { result = QObject::tr("DG monitored voltage is out of range." ); break; } /* 77*/ @@ -111,13 +117,13 @@ /*0081*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR : { result = QObject::tr("HD syringe pump over travel error." ); break; } /* 81*/ /*0082*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR : { result = QObject::tr("HD syringe pump DAC write failure." ); break; } /* 82*/ /*0083*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR : { result = QObject::tr("HD syringe pump is running while the BP is off." ); break; } /* 83*/ -/*0084*/case GuiAlarmID::ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR : { result = QObject::tr("HD venous pressure sensor not being read." ); break; } /* 84*/ +/*0084*/case GuiAlarmID::ALARM_ID_DG_SET_RTC_YEAR_INVALID : { result = QObject::tr("DG set RTC year is invalid." ); break; } /* 84*/ /*0085*/case GuiAlarmID::ALARM_ID_PUMP_TRACK_LATCH_OPENED : { result = QObject::tr("HD pump track latch opened alarm." ); break; } /* 85*/ -/*0086*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR : { result = QObject::tr("HD BP occlusion sensor not being read." ); break; } /* 86*/ +/*0086*/case GuiAlarmID::ALARM_ID_HD_SET_RTC_YEAR_INVALID : { result = QObject::tr("HD set RTC year is invalid." ); break; } /* 86*/ /*0087*/case GuiAlarmID::ALARM_ID_DG_HEATING_INVALID_CAL_RECORD : { result = QObject::tr("DG heating invalid calibration record." ); break; } /* 87*/ /*0088*/case GuiAlarmID::ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE : { result = QObject::tr("DG concentrate pumps hall sensor out of range." ); break; } /* 88*/ -/*0089*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR : { result = QObject::tr("HD BP occlusion sensor error." ); break; } /* 89*/ -/*0090*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR : { result = QObject::tr("DG load cells FPGA read error." ); break; } /* 90*/ +/*0089*/case GuiAlarmID::ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW : { result = QObject::tr("DG outlet UV reactor on with no flow." ); break; } /* 89*/ +/*0090*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT : { result = QObject::tr("DG load cells A1/B1 FPGA fault." ); break; } /* 90*/ /*0091*/case GuiAlarmID::ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE : { result = QObject::tr("HD No dialysate flow data receive in the last 3 seconds." ); break; } /* 91*/ /*0092*/case GuiAlarmID::ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS : { result = QObject::tr("DG conductivity sensor bad status." ); break; } /* 92*/ /*0093*/case GuiAlarmID::ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE : { result = QObject::tr("DG outlet primary conductivity out of range." ); break; } /* 93*/ @@ -133,14 +139,14 @@ /*0103*/case GuiAlarmID::ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID : { result = QObject::tr("HD invalid calibration CRC." ); break; } /* 103*/ /*0104*/case GuiAlarmID::ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS : { result = QObject::tr("Air trap level sensors reporting illegal combination of air/fluid." ); break; } /* 104*/ /*0105*/case GuiAlarmID::ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID : { result = QObject::tr("DG invalid calibration CRC." ); break; } /* 105*/ -/*0106*/case GuiAlarmID::ALARM_ID_DG_DIALYSATE_FLOW_SENSOR_INVALID_CAL_RECORD : { result = QObject::tr("DG dialysate flow sensor invalid calibration record." ); break; } /* 106*/ +/*0106*/case GuiAlarmID::ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD : { result = QObject::tr("DG dialysate flow sensor invalid calibration record." ); break; } /* 106*/ /*0107*/case GuiAlarmID::ALARM_ID_DG_RESTARTED_FAULT : { result = QObject::tr("HD reports DG restarted fault." ); break; } /* 107*/ -/*0108*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR : { result = QObject::tr("HD syringe pump ADC error." ); break; } /* 108*/ +/*0108*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT : { result = QObject::tr("HD syringe pump ADC FPGA fault." ); break; } /* 108*/ /*0109*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR : { result = QObject::tr("HD syringe pump volume check error." ); break; } /* 109*/ /*0110*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR : { result = QObject::tr("HD syringe pump speed check error." ); break; } /* 110*/ /*0111*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR : { result = QObject::tr("HD syringe pump not stopped in off state error." ); break; } /* 111*/ -/*0112*/case GuiAlarmID::ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_SEND_TIMEOUT : { result = QObject::tr("HD blood leak zero command transmit timeout." ); break; } /* 112*/ -/*0113*/case GuiAlarmID::ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_FAULT : { result = QObject::tr("HD blood leak zero command fault." ); break; } /* 113*/ +/*0112*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT : { result = QObject::tr("DG load cells A1/B1 FPGA fault." ); break; } /* 112*/ +/*0113*/case GuiAlarmID::ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE : { result = QObject::tr("DG Drain pump current out of range." ); break; } /* 113*/ /*0114*/case GuiAlarmID::ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE : { result = QObject::tr("HD venous air bubble detector self-test failure." ); break; } /* 114*/ /*0115*/case GuiAlarmID::ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE : { result = QObject::tr("DG temperature sensor out of range." ); break; } /* 115*/ /*0116*/case GuiAlarmID::ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT : { result = QObject::tr("DG temperature sensor ADC fault." ); break; } /* 116*/ @@ -196,7 +202,7 @@ /*0166*/case GuiAlarmID::ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM : { result = QObject::tr("HD in treatment re-circ sub-mode for too long." ); break; } /* 166*/ /*0167*/case GuiAlarmID::ALARM_ID_CARTRIDGE_DOOR_OPENED : { result = QObject::tr("HD cartridge door opened alarm." ); break; } /* 167*/ /*0168*/case GuiAlarmID::ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT : { result = QObject::tr("HD active reservoir depletion timeout." ); break; } /* 168*/ -/*0169*/case GuiAlarmID::ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE : { result = QObject::tr("Dialysate flow rate out of range." ); break; } /* 169*/ +/*0169*/case GuiAlarmID::ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE : { result = QObject::tr("DG Dialysate flow rate out of maximum range." ); break; } /* 169*/ /*0170*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY : { result = QObject::tr("HD syringe empty alarm." ); break; } /* 170*/ /*0171*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION : { result = QObject::tr("HD syringe pump occlusion alarm." ); break; } /* 171*/ /*0172*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM : { result = QObject::tr("HD syringe pump not enough Heparin alarm." ); break; } /* 172*/ @@ -207,7 +213,7 @@ /*0177*/case GuiAlarmID::ALARM_ID_DG_SOFTWARE_FAULT : { result = QObject::tr("DG software fault.\nSoftware found itself in an unexpected state." ); break; } /* 177*/ /*0178*/case GuiAlarmID::ALARM_ID_HD_COMM_TIMEOUT : { result = QObject::tr("HD communication timeout." ); break; } /* 178*/ /*0179*/case GuiAlarmID::ALARM_ID_DG_FPGA_COMM_TIMEOUT : { result = QObject::tr("DG FPGA communication down for too long." ); break; } /* 179*/ -/*0180*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_ADC_ERROR : { result = QObject::tr("DG load cell ADC error." ); break; } /* 180*/ +/*0180*/case GuiAlarmID::ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE : { result = QObject::tr("DG RO flow out of maximum range." ); break; } /* 180*/ /*0181*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE : { result = QObject::tr("DG load cells weight out of range for tare." ); break; } /* 181*/ /*0182*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD : { result = QObject::tr("DG load cells invalid calibration." ); break; } /* 182*/ /*0183*/case GuiAlarmID::ALARM_ID_DG_INVALID_LOAD_CELL_VALUE : { result = QObject::tr("DG invalid load cell value." ); break; } /* 183*/ @@ -223,8 +229,8 @@ /*0193*/case GuiAlarmID::ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE : { result = QObject::tr("HD dialysate temperature out of low safety range." ); break; } /* 193*/ /*0194*/case GuiAlarmID::ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE : { result = QObject::tr("DG RO pump duty cycle out of range." ); break; } /* 194*/ /*0195*/case GuiAlarmID::ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE : { result = QObject::tr("DG RO pump pressure out of range." ); break; } /* 195*/ -/*0196*/case GuiAlarmID::ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT : { result = QObject::tr("DG temperature sensor error bit fault." ); break; } /* 196*/ -/*0197*/case GuiAlarmID::ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT : { result = QObject::tr("DG temperature sensors FPGA fault." ); break; } /* 197*/ +/*0196*/case GuiAlarmID::ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT : { result = QObject::tr("DG CPi/CPo sensors FPGA fault." ); break; } /* 196*/ +/*0197*/case GuiAlarmID::ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT : { result = QObject::tr("DG CD1/CD2 sensors FPGA fault." ); break; } /* 197*/ /*0198*/case GuiAlarmID::ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON : { result = QObject::tr("RO flow too low while primary heater is on." ); break; } /* 198*/ /*0199*/case GuiAlarmID::ALARM_ID_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON : { result = QObject::tr("Dialysate flow too low while trimmer heater is on." ); break; } /* 199*/ /*0200*/case GuiAlarmID::ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE : { result = QObject::tr("DG thermistors/sensors temperature out of range." ); break; } /* 200*/ @@ -249,8 +255,8 @@ /*0219*/case GuiAlarmID::ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED : { result = QObject::tr("HD syringe pump syringe removed alarm." ); break; } /* 219*/ /*0220*/case GuiAlarmID::ALARM_ID_DG_THD_SENSORS_FPGA_FAULT : { result = QObject::tr("DG THd sensors FPGA fault." ); break; } /* 220*/ /*0221*/case GuiAlarmID::ALARM_ID_EMPTY_SALINE_BAG : { result = QObject::tr("HD Saline bag is empty." ); break; } /* 221*/ -/*0222*/case GuiAlarmID::ALARM_ID_DIALYSATE_TEMPERATURE_HIGH : { result = QObject::tr("HD dialysate temperature too high alarm." ); break; } /* 222*/ -/*0223*/case GuiAlarmID::ALARM_ID_DIALYSATE_TEMPERATURE_LOW : { result = QObject::tr("HD dialysate temperature too low alarm." ); break; } /* 223*/ +/*0222*/case GuiAlarmID::ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT : { result = QObject::tr("HD occlusion sensor FPGA fault." ); break; } /* 222*/ +/*0223*/case GuiAlarmID::ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT : { result = QObject::tr("HD arterial sensor FPGA fault." ); break; } /* 223*/ /*0224*/case GuiAlarmID::ALARM_ID_TREATMENT_STOPPED_BY_USER : { result = QObject::tr("Treatment stopped by user action." ); break; } /* 224*/ /*0225*/case GuiAlarmID::ALARM_ID_END_OF_TREATMENT_WARNING : { result = QObject::tr("HD end of treatment warning." ); break; } /* 225*/ /*0226*/case GuiAlarmID::ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY : { result = QObject::tr("HD prime completed low priority alarm." ); break; } /* 226*/ @@ -307,7 +313,7 @@ /*0277*/case GuiAlarmID::ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE : { result = QObject::tr("HD BP occlusion sensor is reading out of range." ); break; } /* 277*/ /*0278*/case GuiAlarmID::ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE : { result = QObject::tr("HD active reservoir weight out of range." ); break; } /* 278*/ /*0279*/case GuiAlarmID::ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT : { result = QObject::tr("DG dialysate drain time out." ); break; } /* 279*/ -/*0280*/case GuiAlarmID::ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR : { result = QObject::tr("HD arterial pressure sensor read timeout error." ); break; } /* 280*/ +/*0280*/case GuiAlarmID::ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID : { result = QObject::tr("DG drain pump direction invalid." ); break; } /* 280*/ /*0281*/case GuiAlarmID::ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME : { result = QObject::tr("Acid concentration bottle low volume alarm." ); break; } /* 281*/ /*0282*/case GuiAlarmID::ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME : { result = QObject::tr("Bicarbonate concentration bottle low volume alarm." ); break; } /* 282*/ /*0283*/case GuiAlarmID::ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE : { result = QObject::tr("DG load cells weight out of range." ); break; } /* 283*/ @@ -317,7 +323,7 @@ /*0287*/case GuiAlarmID::ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE : { result = QObject::tr("HD processor clock speed checks against FPGA clock failure." ); break; } /* 287*/ /*0288*/case GuiAlarmID::ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE : { result = QObject::tr("HD load cells primary/back up drift out of range." ); break; } /* 288*/ /*0289*/case GuiAlarmID::ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION : { result = QObject::tr("DG dialysate or concentrate caps not closed." ); break; } /* 289*/ -/*0290*/case GuiAlarmID::ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON : { result = QObject::tr("DG flow too low while heater is on." ); break; } /* 290*/ +/*0290*/case GuiAlarmID::ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT : { result = QObject::tr("HD venous pressure sensor FPGA fault." ); break; } /* 290*/ /*0291*/case GuiAlarmID::ALARM_ID_HD_DISINFECT_FLUSH : { result = QObject::tr("HD processor is in Disinfect Flush mode." ); break; } /* 291*/ /*0292*/case GuiAlarmID::ALARM_ID_HD_DISINFECT_HEAT : { result = QObject::tr("HD processor is in Disinfect Heat mode." ); break; } /* 292*/ /*0293*/case GuiAlarmID::ALARM_ID_HD_DISINFECT_CHEM : { result = QObject::tr("HD processor is in Disinfect Chemical mode." ); break; } /* 293*/ @@ -326,14 +332,16 @@ /*0296*/case GuiAlarmID::ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT : { result = QObject::tr("DG TDi sensors FPGA fault." ); break; } /* 296*/ /*0297*/case GuiAlarmID::ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT : { result = QObject::tr("DG TRo sensors FPGA fault." ); break; } /* 297*/ /*0298*/case GuiAlarmID::ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT : { result = QObject::tr("DG baro sensor FPGA fault." ); break; } /* 298*/ -/*0299*/case GuiAlarmID::ALARM_ID_DG_DISINFECT_HAS_BEEN_EXPIRED : { result = QObject::tr("DG disinfect has been expired." ); break; } /* 299*/ -/*0300*/case GuiAlarmID::ALARM_ID_DG_SERIVCE_TIME_INTERVAL_HAS_ELAPSED : { result = QObject::tr("DG service time interval has elapsed." ); break; } /* 300*/ +/*0299*/case GuiAlarmID::ALARM_ID_DG_INVALID_SERIAL_NUMBER : { result = QObject::tr("DG invalid serial number." ); break; } /* 299*/ +/*0300*/case GuiAlarmID::ALARM_ID_HD_INVALID_SERIAL_NUMBER : { result = QObject::tr("HD invalid serial number." ); break; } /* 300*/ /*0301*/case GuiAlarmID::ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE : { result = QObject::tr("DG primary heater power voltage out of range." ); break; } /* 301*/ /*0302*/case GuiAlarmID::ALARM_ID_DG_CPU_RAM_ERROR : { result = QObject::tr("DG processor RAM error." ); break; } /* 302*/ -/*0303*/case GuiAlarmID::ALARM_ID_HD_SERVICE_TIME_INTERVAL_HAS_ELAPSED : { result = QObject::tr("HD service time interval has elapsed." ); break; } /* 303*/ +/*0303*/case GuiAlarmID::ALARM_ID__AVAILABLE_4 : { result = QObject::tr("Alarm ID available for use." ); break; } /* 303*/ /*0304*/case GuiAlarmID::ALARM_ID_HD_INVALID_USAGE_RECORD_CRC : { result = QObject::tr("HD invalid usage record." ); break; } /* 304*/ /*0305*/case GuiAlarmID::ALARM_ID_HD_CPU_RAM_ERROR : { result = QObject::tr("HD processor RAM error." ); break; } /* 305*/ -/*0306*/case GuiAlarmID::NUM_OF_ALARM_IDS : { result = QObject::tr("Total number of alarms." ); break; } /* 306*/ +/*0306*/case GuiAlarmID::ALARM_ID_HD_AIR_TRAP_OVERFILL : { result = QObject::tr("HD air trap over-fill alarm." ); break; } /* 306*/ +/*0307*/case GuiAlarmID::ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT : { result = QObject::tr("HD AC power lost second alarm." ); break; } /* 307*/ +/*0308*/case GuiAlarmID::NUM_OF_ALARM_IDS : { result = QObject::tr("Total number of alarms." ); break; } /* 308*/ default: result = QObject::tr("ALARM_ID_UNDEFINED [%1]").arg(int(vAlarmID));