Index: AlarmDefs.h =================================================================== diff -u -r070cd5c49b3bc89b466f8114063b619298fccf8c -re915f11bc03507b9ec3eed29337818896b3ab0e4 --- AlarmDefs.h (.../AlarmDefs.h) (revision 070cd5c49b3bc89b466f8114063b619298fccf8c) +++ AlarmDefs.h (.../AlarmDefs.h) (revision e915f11bc03507b9ec3eed29337818896b3ab0e4) @@ -1291,7 +1291,7 @@ { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. | Blood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, - { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access\n issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access issues, \nif applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, { 129 , ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "HD | Blood Leak: Bypass | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. | Blood Stop: Dialyzer Blood Leak Bypass | HD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "}, { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "},