Index: AlarmDefs.h =================================================================== diff -u -r09ea26a9c0658e248682d35c8fde41a8483948d7 -r380028d3278f3e7dae04e930151403fc391652e1 --- AlarmDefs.h (.../AlarmDefs.h) (revision 09ea26a9c0658e248682d35c8fde41a8483948d7) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 380028d3278f3e7dae04e930151403fc391652e1) @@ -203,7 +203,7 @@ ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM = 165, ///< HD in treatment rinseback operation timeout ALARM_ID___AVAILABLE_5 = 166, ///< Available for use ALARM_ID_CARTRIDGE_DOOR_OPENED = 167, ///< HD cartridge door opened alarm - ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT = 168, ///< HD active reservoir depletion timeout. + ALARM_ID__AVAILABLE_70 = 168, ///< Available for use ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE = 169, ///< DG Dialysate flow rate out of maximum range ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY = 170, ///< HD syringe empty alarm ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION = 171, ///< HD syringe pump occlusion alarm @@ -648,7 +648,7 @@ { ALARM_PRIORITY_HIGH, 11, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID___AVAILABLE_5, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 209, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID__AVAILABLE_70, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, 0, ALARM_ID_NO_ALARM }, @@ -1081,7 +1081,7 @@ { 799, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM }, { 799, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM }, { 799, ALARM_ID___AVAILABLE_5 }, - { 799, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT }, + { 799, ALARM_ID__AVAILABLE_70 }, { 799, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION }, { 799, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM }, { 799, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED }, @@ -1323,7 +1323,7 @@ { 165 , ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM , "HD | Rinseback Timeout | Maximum Rinseback time exceeded. Rinseback must end. | HD in treatment rinseback operation timeout | If the maximum time inrinseback operation has been exceeded. | "}, { 166 , ALARM_ID___AVAILABLE_5 , "NA | Not used | Call service to schedule a repair. | | | "}, { 167 , ALARM_ID_CARTRIDGE_DOOR_OPENED , "HD | Door Open | Hemodialysis Delivery device door open.\n - Close open HD door\n - Press RESUME button | HD cartridge door opened alarm | If the cartridge door is opened as reported by the FPGA. | "}, - { 168 , ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT , "DG | Dialysate Recirculation Error | Dialysate recirculation error. | HD active reservoir depletion timeout. | If the active reservoir's depletion time has elapsed. | "}, + { 168 , ALARM_ID__AVAILABLE_70 , "DG | Dialysate Recirculation Error | Dialysate recirculation error. | HD active reservoir depletion timeout. | If the active reservoir's depletion time has elapsed. | "}, { 169 , ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE , "DG | Dialysate Generation Error | Call service to schedule a repair. | Dialysate flow rate out of range | If the dialysate flow rate > 2.00 L/min. | "}, { 170 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY , "HD | Syringe Empty | Heparin syringe empty.\n- Option to resume treatment without heparin. | HD syringe empty alarm | If the syringe pump position is near empty position. | "}, { 171 , ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION , "HD | Heparin Occlusion | Heparin pump occlusion detected.\nCheck heparin line is unclamped and\nthere are no kinks in the heparin line. | HD syringe pump occlusion alarm | If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. | "},