Index: AlarmDefs.h =================================================================== diff -u -r0b8a815b0e6c7c0ec76fe0f3ef32facd0bf48e93 -rd6b694fb6d35c0f7b70244fa8b27f059ebad023f --- AlarmDefs.h (.../AlarmDefs.h) (revision 0b8a815b0e6c7c0ec76fe0f3ef32facd0bf48e93) +++ AlarmDefs.h (.../AlarmDefs.h) (revision d6b694fb6d35c0f7b70244fa8b27f059ebad023f) @@ -1378,7 +1378,7 @@ { 211 , ALARM_ID_HD_BLOOD_SITTING_WARNING , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. | "}, { 212 , ALARM_ID_HD_END_OF_TREATMENT_ALARM , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment end paused state, if the blood is in the line for more than 4 minutes. | "}, { 213 , ALARM_ID_HD_PRIME_COMPLETED_MEDIUM , "HD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre- Tx Prime Complete | HD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) | "}, - { 214 , ALARM_ID_DG_CONCENTRATE_PUMPS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conc Pump Cal | DG concentrate pump calibration record | If the non-volatile data is valid and if not raise the alarm | "}, + { 214 , ALARM_ID_DG_CONCENTRATE_PUMPS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conc Pump Cal | DG concentrate pump calibration record | If the non-volatile data is invalid raise the alarm. | "}, { 215 , ALARM_ID_AVAILABLE_31 , " | | | | | | "}, { 216 , ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM , "HD | Treatment Complete | Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. | Process: End of Dialysis Tx | HD end treatment sub-mode timeout alarm | If the treatment ended successfully. | "}, { 217 , ALARM_ID_AVAILABLE_32 , " | | | | | | "},