Index: AlarmDefs.h =================================================================== diff -u -r0f4a0359252bd7bb978fcbd45f4d91f06f83fb29 -re530bf98f1da91d645912c005e61dca4659903c4 --- AlarmDefs.h (.../AlarmDefs.h) (revision 0f4a0359252bd7bb978fcbd45f4d91f06f83fb29) +++ AlarmDefs.h (.../AlarmDefs.h) (revision e530bf98f1da91d645912c005e61dca4659903c4) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) Sean Nash -* @date (last) 20-Sep-2022 +* @author (last) Dara Navaei +* @date (last) 06-Sep-2022 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -78,7 +78,7 @@ ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK = 40, ///< Dialysis outlet pump failed motor off check. Measured speed while commanded off ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK = 41, ///< Dialysis outlet pump failed motor direction check. Measured vs. commanded ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK = 42, ///< Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds - ALARM_ID__AVAILABLE_5 = 43, ///< Alarm ID available + ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT = 43, ///< DG main primary heater FPGA fault ALARM_ID_UI_COMM_TIMEOUT = 44, ///< User interface communication timeout ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS = 45, ///< HD too many bad communications CRC ALARM_ID_CAN_MESSAGE_NOT_ACKED = 46, ///< System message that required acknowledgment was not acknowledged @@ -98,8 +98,8 @@ ALARM_ID_HD_VALVE_NOT_FUNCTIONAL = 60, ///< HD valve not functional ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE = 61, ///< HD valve current out of range ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE = 62, ///< HD valve position out of target - ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT = 63, ///< FPGA reports failure reading arterial pressure sensor - ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT = 64, ///< FPGA reports failure reading venous pressure sensor + ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT = 63, ///< HD blood leak FPGA fault + ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE = 64, ///< DG barometric pressure sensor out of range ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT = 65, ///< HD requests DG command with invalid parameter fault ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE = 66, ///< HD blood leak sensor set point set failure ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE = 67, ///< HD blood pump occlusion self-test failure alarm @@ -119,13 +119,13 @@ ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR = 81, ///< HD syringe pump over travel error ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR = 82, ///< HD syringe pump DAC write failure ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR = 83, ///< HD syringe pump is running while the BP is off - ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR = 84, ///< HD venous pressure sensor not being read + ALARM_ID_DG_SET_RTC_YEAR_INVALID = 84, ///< DG set RTC year is invalid ALARM_ID_PUMP_TRACK_LATCH_OPENED = 85, ///< HD pump track latch opened alarm - ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR = 86, ///< HD BP occlusion sensor not being read + ALARM_ID_HD_SET_RTC_YEAR_INVALID = 86, ///< HD set RTC year is invalid ALARM_ID_DG_HEATING_INVALID_CAL_RECORD = 87, ///< DG heating invalid calibration record ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE = 88, ///< DG concentrate pumps hall sensor out of range - ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR = 89, ///< HD BP occlusion sensor error - ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR = 90, ///< DG load cells FPGA read error. + ALARM_ID__AVAILABLE_15 = 89, ///< Alarm ID available + ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT = 90, ///< DG load cells A1/B1 FPGA fault ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE = 91, ///< HD No dialysate flow data receive in the last 3 seconds ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS = 92, ///< DG conductivity sensor bad status. ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE = 93, ///< DG outlet primary conductivity out of range @@ -143,12 +143,12 @@ ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID = 105, ///< DG invalid calibration CRC ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD = 106, ///< DG dialysate flow sensor invalid calibration record ALARM_ID_DG_RESTARTED_FAULT = 107, ///< HD reports DG restarted fault - ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR = 108, ///< HD syringe pump ADC error + ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT = 108, ///< HD syringe pump ADC FPGA fault ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR = 109, ///< HD syringe pump volume check error ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR = 110, ///< HD syringe pump speed check error ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR = 111, ///< HD syringe pump not stopped in off state error - ALARM_ID__AVAILABLE_7 = 112, ///< Alarm ID available - ALARM_ID__AVAILABLE_8 = 113, ///< Alarm ID available + ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT = 112, ///< DG load cells A1/B1 FPGA fault + ALARM_ID__AVAILABLE_8 = 113, ///< Alarm ID available for use ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 114, ///< HD venous air bubble detector self-test failure ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE = 115, ///< DG temperature sensor out of range ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT = 116, ///< DG temperature sensor ADC fault @@ -204,7 +204,7 @@ ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM = 166, ///< HD in treatment re-circ sub-mode for too long ALARM_ID_CARTRIDGE_DOOR_OPENED = 167, ///< HD cartridge door opened alarm ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT = 168, ///< HD active reservoir depletion timeout. - ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE = 169, ///< Dialysate flow rate out of range + ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE = 169, ///< DG Dialysate flow rate out of maximum range ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY = 170, ///< HD syringe empty alarm ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION = 171, ///< HD syringe pump occlusion alarm ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM = 172, ///< HD syringe pump not enough Heparin alarm @@ -215,7 +215,7 @@ ALARM_ID_DG_SOFTWARE_FAULT = 177, ///< DG software fault. Software found itself in an unexpected state ALARM_ID_HD_COMM_TIMEOUT = 178, ///< HD communication timeout ALARM_ID_DG_FPGA_COMM_TIMEOUT = 179, ///< DG FPGA communication down for too long - ALARM_ID_DG_LOAD_CELL_ADC_ERROR = 180, ///< DG load cell ADC error + ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE = 180, ///< DG RO flow out of maximum range ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE = 181, ///< DG load cells weight out of range for tare ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD = 182, ///< DG load cells invalid calibration ALARM_ID_DG_INVALID_LOAD_CELL_VALUE = 183, ///< DG invalid load cell value @@ -231,8 +231,8 @@ ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE = 193, ///< HD dialysate temperature out of low safety range. ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE = 194, ///< DG RO pump duty cycle out of range ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE = 195, ///< DG RO pump pressure out of range - ALARM_ID__AVAILABLE_1 = 196, ///< Alarm ID available for use - ALARM_ID__AVAILABLE_11 = 197, ///< Alarm ID available for use + ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT = 196, ///< DG CPi/CPo sensors FPGA fault + ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT = 197, ///< DG CD1/CD2 sensors FPGA fault ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON = 198, ///< RO flow too low while primary heater is on ALARM_ID_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON = 199, ///< Dialysate flow too low while trimmer heater is on ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE = 200, ///< DG thermistors/sensors temperature out of range @@ -257,8 +257,8 @@ ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED = 219, ///< HD syringe pump syringe removed alarm ALARM_ID_DG_THD_SENSORS_FPGA_FAULT = 220, ///< DG THd sensors FPGA fault ALARM_ID_EMPTY_SALINE_BAG = 221, ///< HD Saline bag is empty - ALARM_ID__AVAILABLE_9 = 222, ///< Alarm ID available - ALARM_ID__AVAILABLE_10 = 223, ///< Alarm ID available + ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT = 222, ///< HD occlusion sensor FPGA fault + ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT = 223, ///< HD arterial sensor FPGA fault ALARM_ID_TREATMENT_STOPPED_BY_USER = 224, ///< Treatment stopped by user action - pressed stop button ALARM_ID_END_OF_TREATMENT_WARNING = 225, ///< HD end of treatment warning ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY = 226, ///< HD prime completed low priority alarm @@ -315,7 +315,7 @@ ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE = 277, ///< HD BP occlusion sensor is reading out of range ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE = 278, ///< HD active reservoir weight out of range. ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT = 279, ///< DG dialysate drain time out - ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR = 280, ///< HD arterial pressure sensor read timeout error + ALARM_ID__AVAILABLE_10 = 280, ///< Alarm ID available for use ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME = 281, ///< Acid concentration bottle low volume alarm ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME = 282, ///< Bicarbonate concentration bottle low volume alarm ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE = 283, ///< DG load cells weight out of range @@ -325,7 +325,7 @@ ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE = 287, ///< HD processor clock speed checks against FPGA clock failure ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE = 288, ///< HD load cells primary/back up drift out of range ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION = 289, ///< DG dialysate or concentrate caps not closed - ALARM_ID__AVAILABLE_12 = 290, ///< Alarm ID available + ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT = 290, ///< HD venous pressure sensor FPGA fault ALARM_ID_HD_DISINFECT_FLUSH = 291, ///< HD processor is in Disinfect Flush mode ALARM_ID_HD_DISINFECT_HEAT = 292, ///< HD processor is in Disinfect Heat mode ALARM_ID_HD_DISINFECT_CHEM = 293, ///< HD processor is in Disinfect Chemical mode @@ -490,7 +490,7 @@ { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETERS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SOFTWARE_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK, 0, ALARM_ID_NO_ALARM }, @@ -505,7 +505,7 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_5, 0, ALARM_ID_NO_ALARM },//add DG version + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS, 0, ALARM_ID_NO_ALARM },//add DG version // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To @@ -526,8 +526,8 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_NOT_FUNCTIONAL, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, @@ -548,13 +548,13 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_SET_RTC_YEAR_INVALID, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PUMP_TRACK_LATCH_OPENED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SET_RTC_YEAR_INVALID, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_15, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -573,11 +573,11 @@ { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESTARTED_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_7, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_8, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -609,7 +609,7 @@ { ALARM_PRIORITY_HIGH, 19, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DG_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 20, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 21, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_OCCLUSION_BLOOD_PUMP, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_RO_FLOW_SENSOR_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE, 0, ALARM_ID_NO_ALARM }, @@ -636,7 +636,7 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM, 0, ALARM_ID_NO_ALARM }, @@ -648,7 +648,7 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_SOFTWARE_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_LOAD_CELL_VALUE, 0, ALARM_ID_NO_ALARM }, @@ -664,8 +664,8 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_1, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_11, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON,0, ALARM_ID_NO_ALARM }, @@ -689,10 +689,10 @@ { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH , 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH , 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_THD_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_THD_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 900, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_EMPTY_SALINE_BAG, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 901, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID__AVAILABLE_9, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, - { ALARM_PRIORITY_LOW, 902, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID__AVAILABLE_10, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 903, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_BY_USER, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_LOW, 904, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_WARNING, 0, ALARM_ID_NO_ALARM }, @@ -752,7 +752,7 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_10, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -762,19 +762,19 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_12, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DISINFECT_FLUSH, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DISINFECT_HEAT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DISINFECT_CHEM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_DISINFECT_CHEM_INSTRUCTION1, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_UI_POST_FAILURE_CLOUDSYNC, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_SERIAL_NUMBER, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_SERIAL_NUMBER, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPU_RAM_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_SERIAL_NUMBER, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_SERIAL_NUMBER, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPU_RAM_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_4, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_USAGE_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_CPU_RAM_ERROR, 0, ALARM_ID_NO_ALARM }, @@ -826,6 +826,8 @@ { 1, ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD }, { 1, ALARM_ID_DG_RO_FLOW_SENSOR_INVALID_CAL_RECORD }, { 1, ALARM_ID_HD_INVALID_USAGE_RECORD_CRC }, + { 1, ALARM_ID_DG_SET_RTC_YEAR_INVALID }, + { 1, ALARM_ID_HD_SET_RTC_YEAR_INVALID }, { 3, ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC }, { 3, ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC }, { 3, ALARM_ID_HD_SOFTWARE_FAULT }, @@ -841,7 +843,7 @@ { 3, ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK }, { 3, ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK }, { 3, ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK }, - { 3, ALARM_ID__AVAILABLE_5 }, + { 3, ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT }, { 3, ALARM_ID_UI_COMM_TIMEOUT }, { 3, ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS }, { 3, ALARM_ID_CAN_MESSAGE_NOT_ACKED }, @@ -861,8 +863,8 @@ { 3, ALARM_ID_HD_VALVE_NOT_FUNCTIONAL }, { 3, ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE }, { 3, ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE }, - { 3, ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT }, - { 3, ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT }, + { 3, ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT }, + { 3, ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE }, { 3, ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT }, { 3, ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, { 3, ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE }, @@ -881,12 +883,10 @@ { 3, ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR }, - { 3, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR }, - { 3, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR }, { 3, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD }, { 3, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE }, - { 3, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR }, - { 3, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR }, + { 3, ALARM_ID__AVAILABLE_15 }, + { 3, ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT }, { 3, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE }, { 3, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS }, { 3, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE }, @@ -900,11 +900,11 @@ { 3, ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID }, { 3, ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS }, { 3, ALARM_ID_DG_RESTARTED_FAULT }, - { 3, ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR }, + { 3, ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT }, { 3, ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR }, - { 3, ALARM_ID__AVAILABLE_7 }, + { 3, ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT }, { 3, ALARM_ID__AVAILABLE_8 }, { 3, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE }, @@ -927,7 +927,7 @@ { 3, ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE }, { 3, ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE }, { 3, ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT }, - { 3, ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR }, + { 3, ALARM_ID__AVAILABLE_10 }, { 3, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE }, { 3, ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE }, { 3, ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE }, @@ -948,15 +948,15 @@ { 3, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE }, { 3, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME }, { 3, ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, - { 3, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE }, + { 3, ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE }, { 3, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE }, { 3, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE }, { 3, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, { 3, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, { 3, ALARM_ID_UV_REACTOR_NOT_HEALTHY }, - { 3, ALARM_ID__AVAILABLE_11 }, - { 3, ALARM_ID__AVAILABLE_1 }, - { 3, ALARM_ID_DG_LOAD_CELL_ADC_ERROR }, + { 3, ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT }, + { 3, ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT }, + { 3, ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE }, { 3, ALARM_ID_DG_FPGA_COMM_TIMEOUT }, { 3, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT }, { 3, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT }, @@ -968,7 +968,7 @@ { 3, ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR }, { 3, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH }, { 3, ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK }, - { 3, ALARM_ID__AVAILABLE_11 }, + { 3, ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT }, { 3, ALARM_ID_HD_RTC_CONFIG_ERROR }, { 3, ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT }, { 3, ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT }, @@ -991,11 +991,13 @@ { 3, ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT }, { 3, ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT }, { 3, ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE }, + { 3, ALARM_ID_DG_INVALID_SERIAL_NUMBER }, + { 3, ALARM_ID_HD_INVALID_SERIAL_NUMBER }, { 3, ALARM_ID_DG_CPU_RAM_ERROR }, { 3, ALARM_ID_HD_CPU_RAM_ERROR }, { 3, ALARM_ID_HD_AIR_TRAP_OVERFILL }, - { 3, ALARM_ID_DG_INVALID_SERIAL_NUMBER }, - { 3, ALARM_ID_HD_INVALID_SERIAL_NUMBER }, + { 3, ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT }, + { 3, ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT }, { 4, ALARM_ID_END_OF_TREATMENT_HIGH }, { 5, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, { 6, ALARM_ID_HD_BLOOD_LEAK_DETECTED }, @@ -1060,8 +1062,6 @@ { 899, ALARM_ID_HD_DISINFECT_CHEM_INSTRUCTION1 }, { 899, ALARM_ID__AVAILABLE_4 }, { 900, ALARM_ID_EMPTY_SALINE_BAG }, - { 901, ALARM_ID__AVAILABLE_9 }, - { 902, ALARM_ID__AVAILABLE_10 }, { 903, ALARM_ID_TREATMENT_STOPPED_BY_USER }, { 904, ALARM_ID_END_OF_TREATMENT_WARNING }, { 905, ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY },