Index: AlarmDefs.h =================================================================== diff -u -r3a11a756ea9070be7c8435dff6b485d10f0c2a58 -r2d8b6de8a61f4337a40ab69b184ce1826a7b7ac3 --- AlarmDefs.h (.../AlarmDefs.h) (revision 3a11a756ea9070be7c8435dff6b485d10f0c2a58) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 2d8b6de8a61f4337a40ab69b184ce1826a7b7ac3) @@ -1284,7 +1284,7 @@ { 123 , ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Trimmer Heater Control | DG trimmer heater voltage out of range | If the trimmer voltage is out of range. | "}, { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, - { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press OK to restart blood pump. Dialysate will\n remain in bypass.\n\nBlood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, + { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press OK to proceed to the next step. Dialysate will\n remain in bypass.\n\nBlood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- If there has been an access disconnection or needle\n dislodgement, follow facility policy to address and\nterminate treatment without rinseback.\n- If access connections are properly secured, \npress RESUME to continue treatment.\n\nBlood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, { 129 , ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "HD | Blood Leak | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to RESUME or END TREATMENT.\n\nBlood Stop: Dialyzer Blood Leak Bypass | HD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "},