Index: AlarmDefs.h =================================================================== diff -u -r3e21e2633bccca3a4575fd26169b0855cfba0e2d -r8fc441b150d7a6aedf415c329ada5c29da99bf27 --- AlarmDefs.h (.../AlarmDefs.h) (revision 3e21e2633bccca3a4575fd26169b0855cfba0e2d) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 8fc441b150d7a6aedf415c329ada5c29da99bf27) @@ -1445,8 +1445,8 @@ { 278 , ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "HD | Dialysate Reservoir Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. | Dial. Gen: Reservoir Overfill | HD active reservoir weight out of range. | If the active reservoir weight is > 1900.0 g. | "}, { 279 , ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Drain | DG dialysate drain time out | If the DG dialysate drain exceeded 120 seconds. | "}, { 280 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Direction Fault | DG drain pump invalid direction | If the drain pump's direction is not 1 when draining. | "}, - { 281 , ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME , "DG | Acid Concentrate Low | The acid concentrate is low.\n- Replace acid concentrate container. | Dial. Gen: Acid Conc. Low | DG Acid concentration bottle low volume alarm | If the acid concentration bottle volume > 3240 (3340 - 100). | "}, - { 282 , ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME , "DG | Bicarb Concentrate Low | Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. | Dial. Gen: Bicarb Conc. Low | DG Bicarbonate concentration bottle low volume alarm | If the bicarbonate concentration bottle volume > 3680 (3780 - 100). | "}, + { 281 , ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME , "DG | Acid Concentrate Low | The acid concentrate is low.\n- Replace acid concentrate container. | Dial. Gen: Acid Conc. Low | DG Acid concentration bottle low volume alarm | If the acid concentration bottle volume <= 500 mL. | "}, + { 282 , ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME , "DG | Bicarb Concentrate Low | Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. | Dial. Gen: Bicarb Conc. Low | DG Bicarbonate concentration bottle low volume alarm | If the bicarbonate concentration bottle volume <= 500 mL. | "}, { 283 , ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC Range | DG load cells weight out of range | If the weight of the DG load cells > 4500.0 g or < 0.0 g. | "}, { 284 , ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Redundant LC Check | DG load cells primary/back up drift out of range | If redundant load cells deviate by more than 60g after tare for 5 seconds. | "}, { 285 , ALARM_ID_HD_RO_PERMEATE_SAMPLE , "HD | RO Permeate Sample in Progress | RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: RO Permeate Sample In Progress | DG RO permeate sample started | If DG RO permeate sample successfully started from UI. | "},