Index: AlarmDefs.h =================================================================== diff -u -r51e66e36e09186658fbfc4102074e2e1c56ca072 -rfb082120d623e65487679937c53156bc32169217 --- AlarmDefs.h (.../AlarmDefs.h) (revision 51e66e36e09186658fbfc4102074e2e1c56ca072) +++ AlarmDefs.h (.../AlarmDefs.h) (revision fb082120d623e65487679937c53156bc32169217) @@ -238,7 +238,7 @@ ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE = 200, ///< DG thermistors/sensors temperature out of range ALARM_ID_PRE_TREATMENT_WET_FLOW_TEST_FAILURE = 201, ///< HD pre-treatment mode wet self-test lc vs FMD failure ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE = 202, ///< HD pre-treatment mode dry pressure normal self-test failure - ALARM_ID_PRESSURE_SENSOR_FAULT = 203, ///< Pressure sensor fault + ALARM_ID___AVAILABLE_3 = 203, ///< Available for use ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE = 204, ///< RO rejection ratio out of range ALARM_ID_CONDUCTIVITY_SENSOR_FAULT = 205, ///< Conductivity sensor fault ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME = 206, ///< DG dialysate fill runs out of time @@ -607,7 +607,7 @@ { ALARM_PRIORITY_HIGH, 4, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_HIGH, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 5, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 6, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_HD_BLOOD_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 6, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_HD_BLOOD_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 8, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_VENOUS_PRESSURE_LOW, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 9, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 10, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK, 0, ALARM_ID_NO_ALARM }, @@ -687,7 +687,7 @@ { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRE_TREATMENT_WET_FLOW_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID___AVAILABLE_3, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME, 0, ALARM_ID_NO_ALARM }, @@ -848,7 +848,7 @@ { 1, ALARM_ID_UI_POST_FAILURE_SDCARD }, { 1, ALARM_ID_UI_POST_FAILURE_RTC }, { 1, ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD }, - { 1, ALARM_ID_PRESSURE_SENSOR_FAULT }, + { 1, ALARM_ID___AVAILABLE_3 }, { 1, ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD }, { 1, ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID }, { 1, ALARM_ID_DG_INVALID_SYSTEM_RECORD_CRC }, @@ -1360,7 +1360,7 @@ { 200 , ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE , "DG | Thermistors Error | Call service to schedule a repair. | DG thermistor or temperature sensor out of range | If thermistor > 80.0 C or < 0.0 C. | "}, { 201 , ALARM_ID_PRE_TREATMENT_WET_FLOW_TEST_FAILURE , "HD | Pre-Treatment Error | Call service to schedule a repair. | HD pre-treatment wet flow test failure | If the pre-treatment wet flow test has failed. | "}, { 202 , ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE , "HD | Pre-Treatment Error | Call service to schedule a repair. | HD pre-treatment dry pressure normal test failure | If the pre-treatment dry pressure normla test failed. | "}, - { 203 , ALARM_ID_PRESSURE_SENSOR_FAULT , "DG | Pressure Sensors Error | Call service to schedule a repair. | DG pressure sensors fault | If pressure sensor ADC reading = 0 or inlet pressure ADC reading > 4095 (12-bit full scale ). | "}, + { 203 , ALARM_ID___AVAILABLE_3 , "NA | Not used | Call service to schedule a repair. | DG pressure sensors fault | If pressure sensor ADC reading = 0 or inlet pressure ADC reading > 4095 (12-bit full scale ). | "}, { 204 , ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE , "DG | RO Rejection Error | Call service to schedule a repair. | RO rejection ratio out of range | If RO rejection ratio > 0.1 and outlet water conductivity > 15.0 uS/cm. | "}, { 205 , ALARM_ID_CONDUCTIVITY_SENSOR_FAULT , "DG | Conductivity Sensors Error | Call service to schedule a repair. | Conductivity sensor fault | If conductivity sensor error count reported by the FPGA > 0 for more than 5 times. | "}, { 206 , ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME , "DG | Dialysate Production Error | Call service to schedule a repair. | DG dialysate fill runs out of time | If DG dialysate fill > 5 minutes. | "},