Index: AlarmDefs.h =================================================================== diff -u -r70b8d98c30e59cc605b89434da237601d0cc17ed -raabdbd3396f04c7398c72495dd39889f03c6d6e6 --- AlarmDefs.h (.../AlarmDefs.h) (revision 70b8d98c30e59cc605b89434da237601d0cc17ed) +++ AlarmDefs.h (.../AlarmDefs.h) (revision aabdbd3396f04c7398c72495dd39889f03c6d6e6) @@ -1379,7 +1379,7 @@ { 212 , ALARM_ID_HD_END_OF_TREATMENT_ALARM , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment end paused state, if the blood is in the line for more than 4 minutes. | "}, { 213 , ALARM_ID_HD_PRIME_COMPLETED_MEDIUM , "HD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre- Tx Prime Complete | HD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) | "}, { 214 , ALARM_ID_DG_CONCENTRATE_PUMPS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conc Pump Cal | DG concentrate pump calibration record | If the non-volatile data is invalid raise the alarm. | "}, - { 215 , ALARM_ID_HD_AIR_TRAP_UPPER_LEVEL_FLUID_TIME_OUT , "HD | Air Trap Overfill | Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots, address as clinically indicated per facility policy. \n- Press Resume/OK to continue treatment. | Air Trap: Air Trap Upper Fluid | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 60 seconds. | "}, + { 215 , ALARM_ID_HD_AIR_TRAP_UPPER_LEVEL_FLUID_TIME_OUT , "HD | Air Trap Overfill | Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment.| Air Trap: Air Trap Upper Fluid | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 60 seconds. | "}, { 216 , ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM , "HD | Treatment Complete | Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. | Process: End of Dialysis Tx | HD end treatment sub-mode timeout alarm | If the treatment ended successfully. | "}, { 217 , ALARM_ID_AVAILABLE_32 , " | | | | | | "}, { 218 , ALARM_ID_HD_SYRINGE_DETECTED , "HD | Syringe Detected | Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. | Process: Remove Syringe | HD syringe detected alarm | If the syringe detection switch > 2.0 volts. | "},