Index: AlarmDefs.h =================================================================== diff -u -r81ceae0e4d2c112902748fa454ca8761493188af -rc8c433400ed045ceebadec7e7b1b121a1a857734 --- AlarmDefs.h (.../AlarmDefs.h) (revision 81ceae0e4d2c112902748fa454ca8761493188af) +++ AlarmDefs.h (.../AlarmDefs.h) (revision c8c433400ed045ceebadec7e7b1b121a1a857734) @@ -796,7 +796,7 @@ { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_USAGE_RECORD_CRC }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_CPU_RAM_ERROR }, - { ALARM_PRIORITY_MEDIUM, 839, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_DG_TURN_OFF_INLET_WATER_VALVES }, + { ALARM_PRIORITY_MEDIUM, 500, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_DG_TURN_OFF_INLET_WATER_VALVES }, { ALARM_PRIORITY_HIGH, 100, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPI_COND_SENSOR_INVALID_CHAR }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPO_COND_SENSOR_INVALID_CHAR }, @@ -1068,6 +1068,7 @@ { 441, ALARM_ID_DG_EXCESSIVE_TILT }, { 442, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM }, { 442, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM }, + { 500, ALARM_ID_DG_TURN_OFF_INLET_WATER_VALVES }, { 590, ALARM_ID_HD_PRIME_COMPLETED_MEDIUM }, { 600, ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE }, { 600, ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE }, @@ -1117,7 +1118,6 @@ { 751, ALARM_ID_DG_CHEM_DISINFECT_FLUSH_REMOVE_ACID }, { 752, ALARM_ID_DG_CHEM_DISINFECT_FLUSH_FLUSH_SAMPLE }, { 753, ALARM_ID_DG_RO_PERMEATE_SAMPLE_REMOVE_DIA_CAP }, - { 839, ALARM_ID_DG_TURN_OFF_INLET_WATER_VALVES }, { 840, ALARM_ID_HD_DISINFECT_FLUSH }, { 841, ALARM_ID_HD_DISINFECT_HEAT }, { 842, ALARM_ID_HD_DISINFECT_CHEM }, @@ -1470,7 +1470,7 @@ { 303 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Hall Commutation | DG drain pump direction FPGA fault | If the FPGA indicates an error for a certain period of time. | "}, { 304 , ALARM_ID_HD_INVALID_USAGE_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Usage Rec NVRAM CRC | HD invalid usage reocord | If the CRC of the HD usage record does not match the calculated CRC. | "}, { 305 , ALARM_ID_HD_CPU_RAM_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Processor RAM error | HD_CPU RAM error | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, - { 306 , ALARM_ID_DG_TURN_OFF_INLET_WATER_VALVES , "DG | Turn Off Inlet Water Valves | Heat disinfection is in progress.\n- Turn off inlet water valves. \n | Cleaning: Turn Off Inlet Water Valves | Selection of nocturnal heat disinfect | If the user has selected the nocturnal heat disinfect, the DG Software prompts the user to turn off the inlet water valves once the reservoirs are filled. | "}, + { 306 , ALARM_ID_DG_TURN_OFF_INLET_WATER_VALVES , "DG | Turn Off Inlet Water Valves | Heat disinfection is in progress.\n- Turn off inlet water valves. | Cleaning: Turn Off Inlet Water Valves | Selection of nocturnal heat disinfect | If the user has selected the nocturnal heat disinfect, the DG Software prompts the user to turn off the inlet water valves once the reservoirs are filled. | "}, { 307 , ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT , "HD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. | "}, { 308 , ALARM_ID_DG_CPI_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CPi Comm | DG CPi conductivity sensor transmits invalid characters | If the CPi conductivity sensor transmits invalid characters for a certain period of time. | "}, { 309 , ALARM_ID_DG_CPO_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CPo Comm | DG CPo conductivity sensor transmits invalid characters | If the CPo conductivity sensor transmits invalid characters for a certain period of time. | "},